Brand:

Beamex 9010180 PGPH Pneumatic Calibration Pump 140 bar

Description

Beamex 9010180 PGPH pneumatic calibration pump: overview

The Beamex 9010180 is the PGPH, a lever-operated pneumatic calibration pump that produces gauge pressure up to 140 bar (2000 psi) or vacuum down to -0.95 bar (-14 psi) from ordinary clean air. It sits at the heavy end of the Beamex hand-pump range: where the PGC stops at 35 bar and the PGM at 20 bar, the 9010180 carries on to 140 bar without needing a hydraulic liquid, a reservoir or any cleaning of oil residues afterwards. Technicians use it to drive pressure gauges, transmitters, transducers and switches through their span while a calibrator reads the reference value.

Because the 9010180 is a bench-style pump with a long lever, a shut-off valve, a release valve and a fine adjust handwheel, it behaves more like workshop equipment than a pocket tool. The body is built to be set on a table; the pump weighs 7.1 kg, so it stays put while you pump and trim.

Features of the 9010180 that matter at the bench

  • Pressure and vacuum from one pump: a pull-out/push-in selector chooses pressure (pulled out) or vacuum (pushed in), and a selector lock prevents accidental changes. The manual warns never to change the selector while the system is pressurized.
  • Shut-off valve that isolates the pump from the measurement circuit once you have reached the approximate point, so the fine adjust works on a closed system.
  • Fine adjust handwheel: turning clockwise raises pressure, which is how you land on exact calibration points after the lever has done the coarse work.
  • Built-in relief valve, with a stated safety pressure of 180 bar (less than 2600 psi).
  • Two equivalent G 1/4 inch female pressure connectors, one for the device under test and one for the calibrator or pressure module, both hand tightened.
  • A plugged G 1/4 inch clean-out connection at the bottom lets you open the pump and flush out impurities that jam the pipeline.

9010180 technical specifications

Parameter Value
Order code 9010180
Pressure range -0.95 bar (-14 psi) to 140 bar (2000 psi)
Safety pressure 180 bar, under 2600 psi, relief valve built in
Pressure medium Clean air
Pressure resolution 0.1 mbar / 0.002 psi
Operating temperature 0 to 50 degrees C
Humidity below 85 %RH
Height 178 mm lever down, 506 mm lever up
Length 540 mm lever down, 450 mm lever up
Width 270 mm
Weight 7.1 kg (15.7 lb)
Output connectors Two G 1/4 inch female, plus plugged G 1/4 inch female clean-out
Standard accessories G 1/4 inch male plug, NBR70 seal set 6 x 2, guarantee leaflet, instruction manual

Vacuum depth depends on local atmosphere: with 1 bar ambient it reaches -0.95 bar, and at any other ambient pressure P the achievable vacuum is -0.95 multiplied by P.

Where a 140 bar air pump earns its place in Oman and the GCC

Most refinery, gas-processing and power-plant instrument shops in the region keep a pneumatic route for pressures up to about 140 bar because it avoids contaminating transmitters with oil. Pressure transmitters on compressor discharge lines, regulators on instrument-air headers, and high-range switches on gas pipelines all fall inside the pump range. In water and desalination plants the same unit checks pump discharge gauges, while manufacturing sites use it for pneumatic test benches. The pump is air only, so it is not suited to liquid-filled systems or explosive atmospheres; the manual notes that compressed air may cause problems in explosive or corrosive environments.

Selecting the 9010180 against the other Beamex pump models

Pick the 9010180 pump when your highest test point is above 35 bar but you want to stay on air. Under 35 bar, the PGC (-0.95 to 35 bar) is far lighter at 820 g; for liquid media above 140 bar the hydraulic PGHH (to 700 bar) or PGHS (to 1000 bar) are the alternatives. The PGPH has no integral hose, so pair it with a hose rated for the test pressure. The manual lists a 400 bar hose and connector set 8003380 (PGPH end G 1/4 inch male, instrument end G 1/4 inch NPT male) and a second set 8003365 for connecting a calibrator high-pressure module (PGPH end G 1/4 inch male, module end G 1/8 inch male). An external EXT module does not need the second set.

Connecting and operating the 9010180

Preparation follows a fixed sequence: select pressure mode, open the shut-off and release valves, move the fine adjust to the middle, then close the release valve and connect the device under test to one G 1/4 inch port and the reference to the other. Pump until near the point, close the shut-off valve, and trim with the fine adjust. After reaching the top of the range, crack the release valve to step down in controlled stages. Expect a small pressure drift right after a change because the hose stretches and gas temperature changes; re-trim with the handwheel rather than pumping again. Always keep a reliable pressure indicator attached, and stop immediately if the lever feels hard without a pressure rise.

Maintenance and leak testing of the 9010180

The main piston seals wear with use and are replaced from the optional service seal kit 8003180, which carries its own instructions. A tube of grease, code 0005180, is listed for the fine adjust stem. Hard-to-turn valves usually mean they were closed too tightly last time, or that the new pump needs running in. For a leak check, connect an accurate reference such as an EXT250 module, pump to 140 bar and note the reading at 5 and 10 minutes; a loss ratio under 1 percent means the system is tight. For vacuum, drop to -0.95 bar and the loss over the same five minutes should stay under 1 kPa. Never over-tighten the connectors.

How 9010180 differs from other PGPH and Beamex pump variants

Within the pneumatic family, the 9010180 is the only PGPH order code; its neighbours are separate product lines, not options. Against the PGC (9010050 and 9010060) it offers four times the pressure ceiling but weighs 7.1 kg instead of 820 g. Against the PGL (9010090) it covers 140 bar instead of plus or minus 400 mbar. The PGPH relies on clean air whereas the PGHH uses mineral hydraulic oil or distilled water and a 200 ml reservoir. Its ports are G 1/4 inch female, while the PGC side port is G 1/8 inch with a Bx 40 bar hose fitting.

Documents and quotation

The instruction manual covering the 9010180 describes connections, specifications, hose data and the warning list, and is the reference we use for this page. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer. To request pricing and lead time, send an RFQ stating Beamex 9010180, the quantity and any accessory codes such as 8003180 or 8003380 you need alongside it.

Datasheets & Documents
  • PDF Datasheet Beamex 9010180 PGPH Pneumatic Calibration Pump 140 bar Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Beamex is a Finnish maker of calibration equipment and software for process instruments. Four calibration technicians founded the company in 1975 because they could not find the equipment they wanted; their first product was a voltage and current simulator, and the first factory opened in 1977. Beamex is headquartered in Pietarsaari (Jakobstad), Finland, and supplies process industries in more than 100 countries.

Seven Star LLC is an independent supplier listing Beamex documenting calibrators, temperature dry blocks, pressure controllers, calibration pumps and pressure modules for buyers in Oman and the GCC. Equipment is quoted against the model and the pressure or temperature ranges required.

Beamex company profile

Origin Founded in 1975 in Finland by four calibration technicians; first factory 1977
Headquarters Pietarsaari (Jakobstad), Finland; ISO 9001 certified since 1992
Product families MC series multifunction calibrators, temperature dry blocks, automatic and hand pressure generators, external pressure modules, CMX and LOGiCAL calibration software
Typical duty Field and workshop calibration of pressure, temperature and electrical signal instruments, including HART and fieldbus transmitters

Milestones

  • 1975: Four calibration technicians found the company in Finland.
  • 1977: The first factory opens.
  • 1992: ISO 9001 certification.

Beamex calibrators and accessories

Model Type Typical use
MC6 Advanced field calibrator and communicator Pressure, temperature and electrical calibration with a built-in HART, FOUNDATION Fieldbus and PROFIBUS PA communicator, documented results
MC6-Ex Intrinsically safe field calibrator Calibration inside Zone 0 and Zone 1 hazardous areas without a hot work permit
MC6-T Temperature calibrator with integrated dry block Temperature sensors and transmitters, with multifunction electrical and pressure measurement
MC4, MC2 Documenting and hand-held process calibrators Routine loop, transmitter and switch checks in the field
FB, MB Field and metrology temperature dry blocks FB for field calibration; MB for higher-accuracy workshop and laboratory work
ePG, POC8 Electric pressure pump and automatic pressure controller Automated pressure calibration; POC8 for workshop calibration benches
PGC, PGM, PGL, PGPH, PGHH Hand pressure pumps Pressure and vacuum generation in the field, from low pressure (PGL, ±400 mbar) to hydraulic 700 bar (PGHH)
EXT modules External pressure modules Extending MC calibrators to additional ranges, 70 kPa to 100 MPa

Where Beamex equipment is used

Instrument workshops and maintenance teams in oil and gas, refining, petrochemicals, power, water and pharmaceuticals use Beamex calibrators to verify transmitters, switches, gauges and temperature sensors. Documenting calibrators send results to calibration software, which keeps the audit trail required by quality systems and custody transfer metering procedures. Intrinsically safe models are used for in-situ calibration in classified areas.

Ordering Beamex equipment: what we need

State the model and the options you need: internal and external pressure ranges, temperature range, communication protocols (HART, FOUNDATION Fieldbus, PROFIBUS PA), Ex approval, and whether an accredited calibration certificate is required. For spare modules and pumps send the serial number of the calibrator they will be used with, so compatibility can be confirmed.

Facts checked against Beamex company history and product information, October 2026.

FAQ

Questions about the Beamex 9010180 PGPH Pneumatic Calibration Pump 140

Answers from our engineering team on installation, compatibility and integration.

01What pressure range does the Beamex 9010180 cover?
The 9010180 PGPH generates gauge pressure up to 140 bar (2000 psi) and vacuum to -0.95 bar (-14 psi), using clean air as the medium.
02Does the 9010180 include a relief valve?
Yes. A built-in relief valve is part of the pump and the stated safety pressure is 180 bar, below 2600 psi.
03Which hose can I use with the 9010180?
Use Beamex hoses rated for the test pressure; the manual lists 400 bar hose sets 8003380 and 8003365 with G 1/4 inch male ends on the pump side.
04How is the 9010180 serviced?
The piston seals are replaced from service seal kit 8003180, and the pump has a plugged clean-out hole at the bottom to flush impurities.
05Can I use the 9010180 for hydraulic calibration?
No. It is an air-only pump, so liquid-filled systems need a hydraulic model such as the PGHH or PGHS.