Brand: Beamex
Beamex 8003124 PGC Piston and Handle Assembly
Description
Beamex 8003124: PGC piston and handle assembly
Beamex 8003124 is the preassembled bottom part of the Beamex PGC hand pump. It contains the two handles and the piston with its piston seals, and it is designed to be exchanged as a unit when the bottom half of a PGC is broken. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.
The PGC itself is a hand-operated pump that makes both pressure and vacuum from air, up to 35 bar and down to minus 0.95 bar, with a selector for switching between the two and a fine adjust for trimming the setpoint.
Serial number rule when ordering
One detail makes 8003124 different from most spares: when this item is installed the pump receives a new serial number. For that reason the maker’s shop limits purchase to one unit and asks for the pump serial number when ordering. Please include the existing serial number in the RFQ so the correct assembly can be confirmed and the replacement record kept straight.
Specification table for the PGC fitted with 8003124
| Item | Detail |
|---|---|
| Part number | 8003124 |
| Description | PGC piston and handle assembly |
| Parts included | Handles, piston and piston seals, preassembled |
| Pump pressure range | Minus 0.95 to 35 bar / minus 13.7 to 510 psi |
| Pressure medium | Air |
| Side connection | G1/8 inch female with pre-installed Bx G1/8 male fitting, for 40 bar hoses |
| Top connection | Plugged G3/8 inch B female port |
| Pump weight | 820 g / 1.81 lb |
| Pump height x width x depth | 220 x 120 x 65 mm |
| Materials | Aluminum, brass, ABS, NBR |
| Operating temperature | 0 to 50 degrees C |
| Storage temperature | Minus 20 to 60 degrees C |
What wears in the lower assembly
Every pump stroke loads the piston seal, and the manual lists the seal positions that may need changing after enough use: the piston seal as a square ring, four G1/8 bonded seals, three 3.17 x 1.78 O-rings, an 8 x 2.5 O-ring, a gasket, a brass washer and the cylinder mounting screws. When the troubleshooting steps fail, such as secure connections, plugged ports, a closed release valve and a selector fully in position, worn seals are the next suspect.
Instead of replacing individual seals, 8003124 lets a workshop exchange the complete moving half, including handles, which also cures cracked or loose handles.
Typical wear indications are a pump that needs more strokes than before to reach a point, pressure that sags when the handles are released, or a piston that feels rough. In shops where the pump is used for hundreds of strokes per day, such wear arrives within a couple of years, and having the whole lower half on the shelf makes the repair a ten-minute swap instead of a seal-by-seal rebuild.
How 8003124 differs from other PGC spares
The cylinder assembly 8003126 is the top part of the pump, with the pressure and vacuum selector, nameplate sticker and mounting screws. The fine adjust with release valve is 8003122. 8003124 is the bottom section: handles plus piston. The PGC spare parts set 8003120 covers seals only. Choose 8003124 for broken handles or a piston assembly that has been damaged beyond seal replacement; choose the cylinder assembly if the selector or cylinder bore is at fault.
Fitting notes
Depressurize the pump and keep it free of any pressure source; the manual forbids connecting it to external pressure sources. Remove the old lower assembly, fit the new one, and check that the handle spread works. To adjust the handle spread, loosen the lower nut, turn the upper nut to set spring tension, then tighten the lower nut against the upper one. The lower nut normally sits about 1 centimeter from the lowest position.
Never use PTFE tape on any part of the pump, and tighten hose connections by hand only.
After the swap, set the handle spread. The overview drawing of the PGC lists a stroke length setting, and the spread is set with the two nuts as described above. The manual says the handles may be adjusted even during the calibration run.
Leak test after the repair
The manual gives numbers. For pressure, generate 25 bar with the handles, increase to 28 bar with the fine adjust and hold for one minute; the pressure must not fall by more than 0.09 bar. For vacuum, pull to minus 0.85 bar and hold for one minute; the fall must not exceed 0.02 bar. If the first stroke after long storage feels sluggish, the manual says this usually disappears in use.
Applications and cautions in the field
A compact air pump of this kind serves instrument technicians in oil and gas, power, water treatment and building services across Oman and the GCC, mostly for low and medium pressure transmitters, switches and gauges. Use only hoses marked Max. 40 bar, never use one hose for both gases and liquids, and replace hoses at once if they show damage. The maker also recommends replacing all hoses every five years.
Because the aluminum body can be damaged by excessive force, the manual advises against using the plugged top connection for the pressure circuit.
Process instruments can hold moisture and dirt that may enter the pump and cause serious internal corrosion, so vent and flush external systems before connecting them to the PGC. The maker’s vent valve for 40 bar hoses can be used for venting.
Documents and quotation
The PGC user manual is the datasheet for Beamex 8003124. To obtain a quotation, send an RFQ to Seven Star LLC with part number 8003124, quantity and the PGC pump serial number.
Ordering Checklist for the Beamex 8003124
Before you ask for a quotation for the Beamex 8003124, check the full order code on the nameplate of the unit you are replacing, the supply voltage, and the process or signal ranges your system expects. If this is an upgrade rather than a like-for-like swap, tell us the existing make and model so that mounting, wiring and connection details can be checked against the process, pressure & temperature calibrators already installed. For multi-unit orders, list the quantity per site and the delivery location in Oman or the wider GCC, so packing, paperwork and lead time can be confirmed in a single quotation.
Other Models in the Pgc Spares Range
Seven Star LLC supplies models from the Pgc Spares range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.
| Model | Category | |
|---|---|---|
| Beamex 8003124 PGC Piston and Handle Assembly | Process, Pressure & Temperature Calibrators | You are viewing this model |
| Beamex 8003126 PGC Cylinder Assembly with Selector | Process, Pressure & Temperature Calibrators | View details |
Datasheets & Documents
- Datasheet Beamex 8003124 PGC Piston and Handle Assembly Datasheet Download
Reviews (0)
How would you rate this product? Cancel reply
About brand
Beamex is a Finnish maker of calibration equipment and software for process instruments. Four calibration technicians founded the company in 1975 because they could not find the equipment they wanted; their first product was a voltage and current simulator, and the first factory opened in 1977. Beamex is headquartered in Pietarsaari (Jakobstad), Finland, and supplies process industries in more than 100 countries.
Seven Star LLC is an independent supplier listing Beamex documenting calibrators, temperature dry blocks, pressure controllers, calibration pumps and pressure modules for buyers in Oman and the GCC. Equipment is quoted against the model and the pressure or temperature ranges required.
Beamex company profile
| Origin | Founded in 1975 in Finland by four calibration technicians; first factory 1977 |
|---|---|
| Headquarters | Pietarsaari (Jakobstad), Finland; ISO 9001 certified since 1992 |
| Product families | MC series multifunction calibrators, temperature dry blocks, automatic and hand pressure generators, external pressure modules, CMX and LOGiCAL calibration software |
| Typical duty | Field and workshop calibration of pressure, temperature and electrical signal instruments, including HART and fieldbus transmitters |
Milestones
- 1975: Four calibration technicians found the company in Finland.
- 1977: The first factory opens.
- 1992: ISO 9001 certification.
Beamex calibrators and accessories
| Model | Type | Typical use |
|---|---|---|
| MC6 | Advanced field calibrator and communicator | Pressure, temperature and electrical calibration with a built-in HART, FOUNDATION Fieldbus and PROFIBUS PA communicator, documented results |
| MC6-Ex | Intrinsically safe field calibrator | Calibration inside Zone 0 and Zone 1 hazardous areas without a hot work permit |
| MC6-T | Temperature calibrator with integrated dry block | Temperature sensors and transmitters, with multifunction electrical and pressure measurement |
| MC4, MC2 | Documenting and hand-held process calibrators | Routine loop, transmitter and switch checks in the field |
| FB, MB | Field and metrology temperature dry blocks | FB for field calibration; MB for higher-accuracy workshop and laboratory work |
| ePG, POC8 | Electric pressure pump and automatic pressure controller | Automated pressure calibration; POC8 for workshop calibration benches |
| PGC, PGM, PGL, PGPH, PGHH | Hand pressure pumps | Pressure and vacuum generation in the field, from low pressure (PGL, ±400 mbar) to hydraulic 700 bar (PGHH) |
| EXT modules | External pressure modules | Extending MC calibrators to additional ranges, 70 kPa to 100 MPa |
Where Beamex equipment is used
Instrument workshops and maintenance teams in oil and gas, refining, petrochemicals, power, water and pharmaceuticals use Beamex calibrators to verify transmitters, switches, gauges and temperature sensors. Documenting calibrators send results to calibration software, which keeps the audit trail required by quality systems and custody transfer metering procedures. Intrinsically safe models are used for in-situ calibration in classified areas.
Ordering Beamex equipment: what we need
State the model and the options you need: internal and external pressure ranges, temperature range, communication protocols (HART, FOUNDATION Fieldbus, PROFIBUS PA), Ex approval, and whether an accredited calibration certificate is required. For spare modules and pumps send the serial number of the calibrator they will be used with, so compatibility can be confirmed.
Facts checked against Beamex company history and product information, October 2026.
Browse Beamex products by category: Portable Test & Measurement, Temperature Instrumentation.
FAQ
Questions about the Beamex 8003124 PGC Piston and Handle Assembly
Answers from our engineering team on installation, compatibility and integration.
01Why does Beamex 8003124 need the pump serial number?
Installing it gives the pump a new serial number, so the maker's shop asks for the existing number when ordering and limits the purchase to one unit.
02What is included in the assembly?
The preassembled bottom part of the PGC with handles and a piston carrying its piston seals.
03What pressure range does the PGC cover?
Minus 0.95 to 35 bar, using air as the medium, with a selector to switch between pressure and vacuum.
04How do I check the pump after fitting?
Hold 28 bar for one minute with a fall of no more than 0.09 bar, then check vacuum at minus 0.85 bar with a fall of no more than 0.02 bar.
05Does 8003124 include the pressure and vacuum selector?
No. The selector belongs to the cylinder assembly 8003126, the top part of the pump.