Brand:

Additel ADT762 Automated Hydraulic Pressure Calibrator

Description

Hydraulic calibration to 15,000 psi without a hand pump

The Additel ADT762 is a fully automated hydraulic calibrator. You set a pressure from the 7 inch touchscreen and the built-in electric pump, controller and reservoir bring the system to the set point and hold it to 0.005% FS control stability, from 100 to 15,000 psi for more than five minutes. It targets the high-pressure work that normally needs a screw press, a reference gauge and a lot of patience: transmitters, switches, dial and digital gauges and sensors, including HART and PROFIBUS devices.

Because the pump, filter, reservoir, documentation and HART tools are all inside one 12.8 kg case, one technician can run an automated test on a pressure transmitter, save the as-found and as-left data and print a report afterwards.

Notable functions

  • Dual-range accuracy: the instrument switches between two stored calibrations as the pressure rises, in automatic or manual mode.
  • Built-in auto-purge routine that removes air from the hydraulic circuit at the touch of a button.
  • Documented task feature for up to 1,000 tasks, with automatic error analysis and test reports.
  • Dual-mode HART communication, automatic calibration mode or manual mode, plus a full HART communicator.
  • Wi-Fi, LAN, Bluetooth, USB and Ethernet, with ACal, Additel Land and the Link mobile app.

Technical specification

Parameter Specification
Pressure range 15 to 10,000 psi (1 to 700 bar) for GP10K; 15 to 15,000 psi (1 to 1,035 bar) for GP15K
Range selection GP10K: 3K and 10K psi; GP15K: 6K and 15K psi; manual or automatic
Accuracy 0.01% FS (7 digits) or 0.02% FS (6 digits)
Control stability (30 s) 0.005% FS or better from 100 to 15,000 psi; stability duration over 5 minutes
Maximum external load 80 ml at 60 MPa; 60 ml at 100 MPa; 50 ml recommended
Reservoir 350 ml maximum, built-in filter
Media Sebacate oil
Pressure module One internal dual-range module; external control modules ADT161-01/02-GPxx from 1,000 to 15,000 psi
External measure modules All ADT161 modules, two inputs
mA and voltage measure +/-25 and +/-50 mA (0.008% RD + 1.0 / 2.0 uA); +/-300 mV to +/-30 V (0.008% RD + 6 uV to 600 uV)
Sources mA 0 to 25 mA; V 0 to 16 V; loop 24 V at 50 mA; power source 16 to 30 V at 70 mA
Display and interface 7 in 800 x 480 TFT touchscreen and keypad; IP31 enclosure
Communication USB, LAN, Bluetooth, Wi-Fi, Ethernet
Battery Li-ion, typically 8 h, recharge under 4 h
Size and weight 300 x 220 x 192 mm (11.81 x 8.66 x 7.56 in); 12.8 kg (28 lb) without media
Storage, logging and tasks Over 8 GB; 1,000,000 readings; 1,000 tasks
Environment Operating 0 to 50 C; storage -20 to 70 C; under 90% RH
Pump life Over 1,000,000 cycles
Certificate and warranty ISO 17025 accredited, NIST-traceable data; CE; 18 months

Ordering the correct variant

The code is ADT762-01-GP15K-PSI-N. The accuracy field is 01 (0.01% FS) or 02 (0.02% FS), the range is GP10K or GP15K, the range type is PSI or BAR and the female pressure port is N (1/4 NPT), N2 (1/2 NPT), B (1/4 BSP), B2 (1/2 BSP) or M (M20 x 1.5). Choose GP10K if nothing above 10,000 psi is tested, because its two calibrations (3,000 and 10,000 psi) give better low-end accuracy than the 6,000 / 15,000 psi pair. Large external volumes degrade performance, so keep the connected load at the recommended 50 ml or less.

Setup, purging and care

The manual fills the reservoir through a removable cover: fill the working medium between the Min and Max marks, refit the cover and select the In Use position. Refilling through the injection port with the small metal funnel can only be done before power-on or after shutdown, and the level must not exceed the Max mark. Seal the pressure outlet before running auto-tune. A 500 ml bottle of sebacate oil is included and a one litre bottle 9201 is optional; run the built-in auto-purge routine to remove air. Hoses and filters should be replaced after about 10 years or at the first sign of wear.

Typical customers in Oman and the GCC

High-pressure calibration is needed for wellhead and pipeline transmitters, hydraulic test benches, valve actuator testing, subsea and drilling instrumentation. A portable hydraulic calibrator saves the trip to a central lab: the unit can be taken to a compressor station or workshop and run as a stand-alone reference with NIST-traceable data.

Selection and installation notes

Calibration equipment should be set up before it is relied on. Check the range and connections against the devices you will test, allow the instrument to warm up and stabilise, keep the fittings and tubing clean and use the correct adapters and seals. Keep the calibration certificate on file, plan the re-calibration interval according to your quality system, charge and store the battery as the manual describes and protect the case from impact when it is carried to site.

Hydraulic systems need a suitable clean fluid and careful venting, so check that the test fluid is compatible with the devices under test and with the wetted parts of the calibrator.

Information needed for a quotation

Send the full model code, the quantity, the delivery location in Oman or the GCC, the intended use and any accessories needed, so that the complete package can be quoted together.

Documents and supply information

The manufacturer data sheet is attached to this page and should be read together with the project specification and the site risk assessment before a purchase is confirmed. Ratings, standards and dimensions are quoted as printed on that sheet. Lead time, stock position and certificates available with the shipment are confirmed in the quotation, and we supply to Oman and the wider GCC region for industrial, utility and building services projects.

Accessories and related items

  • Included: test leads 9022, USB cable, 9725 Li-ion battery, 9818 power adapter, funnel, high-pressure hose, small accessory case, transport cap, carrying case 9907-762, sebacate oil, O-rings and an ISO 17025 certificate.
  • Optional: ADT107-X-KIT dual-port manifold for 15,000 psi, ADT161 modules, replacement accumulator 9241 and sebacate oil 9201.

Datasheet and quotation

The datasheet is attached. Please send the pressure range, accuracy and thread you need and our team will prepare a quotation; related tools are under portable test and measurement. Confirm the model code before ordering.

Datasheets & Documents
  • PDF Datasheet Additel 762 automated hydraulic pressure calibrator datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Additel is an American manufacturer of pressure and temperature calibration equipment, headquartered in Brea, California, where it also runs an NVLAP-accredited calibration laboratory. Its range covers automated pressure calibrators and controllers, documenting process calibrators, digital pressure gauges, test pumps and dry well calibrators. Additel has offices in Utah, Beijing and Odense, Denmark, and its products are used in more than 100 countries.

Seven Star LLC is an independent supplier in Muscat. We list Additel instruments with their official datasheets for instrument and calibration teams in oil and gas, power, water and manufacturing in Oman and the GCC.

Additel company profile

Headquarters Brea, California, USA
Calibration laboratory NVLAP-accredited, Brea
Other offices Pleasant Grove (Utah), Beijing and Odense (Denmark)
Reach Products used in more than 100 countries
Product families ADT760A and ADT761A automatic handheld pressure calibrators, ADT762 hydraulic calibrator, ADT226 documenting process calibrator, ADT681A and ADT601Ex digital pressure gauges, ADT875 dry well calibrators, pressure test pumps

Additel product series

Pressure calibration equipment includes the ADT760A and 760S automatic handheld pressure calibrators, the ADT761A portable automatic calibrator, the ADT762 automated hydraulic calibrator and the ADT927 hydraulic test pump for 700 bar. Digital pressure gauges include the ADT681A wireless gauge with accuracy up to 0.01% of full scale and the ADT601Ex intrinsically safe gauge.

The ADT226 is a multifunction documenting process calibrator, and the ADT295 communicates with HART, Foundation Fieldbus and Profibus PA devices. Temperature calibration is covered by the ADT875 dry well calibrators and the ADT878 reference dry well calibrators.

Where Additel instruments are used

Field calibration of pressure and temperature transmitters, gauges, switches and RTDs, loop checks and documented calibration for quality audits, and calibration labs in plants and service companies.

Additel products we list

Family Examples
Portable Test & Measurement ADT226 Multifunction Documenting Process…, ADT760A / 760S Automatic Handheld Pressu…, ADT761A Portable Automatic Pressure Cali…, ADT762 Automated Hydraulic Pressure Cali…
Pressure Instrumentation ADT601Ex Intrinsically Safe Digital Pres…, ADT681A Wireless Digital Pressure Gauge
Temperature Instrumentation ADT875 Series Dry Well Temperature Calib…

Ordering Additel instruments: what we need

Send the model and pressure range, the medium (gas, oil or water), accuracy class and whether results must be documented. For dry wells give the temperature range, stability and insert size. We confirm the exact model code before ordering.

Facts checked against Additel's own catalogue and calibration services brochure, October 2026.

Frequently Asked Questions

What pressure does the ADT762 generate?
It generates and controls hydraulic pressure from 15 psi to 10,000 psi (1 to 700 bar) in the GP10K model and from 15 psi to 15,000 psi (1 to 1,035 bar) in the GP15K model.
What does dual-range accuracy mean?
The built-in module has two calibrations: 0 to 3,000 and 0 to 10,000 psi on the GP10K, or 0 to 6,000 and 0 to 15,000 psi on the GP15K. The unit switches automatically or manually, giving 0.01% FS or 0.02% FS accuracy on each range.
What fluid does the ADT762 use?
The datasheet lists sebacate oil as the media, with a 350 ml reservoir that has a built-in filter. A 500 ml bottle of sebacate oil is included.
Can it calibrate HART transmitters?
Yes. It has a dual-mode HART function, automatic and manual, plus a full HART communicator and HART and PROFIBUS PA support, with DD files updated periodically.
How heavy is the ADT762?
28 lb (12.8 kg) without media, measuring 300 x 220 x 192 mm, so it is portable but not a handheld tool.