Brand:

Normalab NAE 440 Automatic Twin Aniline Point Tester

Description

Twin-station aniline point tester using a dry bath

The Normalab NAE 440 is an automatic aniline point tester with two independent, self-contained stations. Normalab describes it as the unique twin tester on the market using the dry bath concept, so no liquid heating bath is needed, and the heating element is protected if a tube breaks. Each station detects the aniline point automatically in clear, coloured and dark products, and visual checking remains possible.

Technical specifications

Parameter Specification
Manufacturer / model Normalab NAE 440
Order number 40500
Stations 2 independent
Test temperature range -10 to 150 °C (brochure); an external cooler or cryostat is needed below ambient
Heating Dry bath
Product detection Clear, coloured and dark products up to 8 on the ASTM colour scale
Data storage 200 results
Self-checks Autodiagnostic
Output RS232 for printer or LIMS
Power 230 V, 50/60 Hz, 1.5 A
Dimensions and weight 520 x 490 x 540 mm; 30 kg

Standards and what the test tells you

The brochure lists ASTM D611, ISO 2977 method 5, ISO 21493:2019 and IP 2 and related methods. They cover the aniline point of petroleum products and hydrocarbon solvents, and the mixed aniline point for products whose aniline point lies below the temperature at which aniline crystallizes from the mixture. The aniline point is the lowest temperature at which equal volumes of aniline and sample are fully miscible; it indicates the aromatic content of a hydrocarbon, with a low value meaning more aromatics. It is used for diesel ignition quality estimates (aniline gravity product) and for solvent and rubber oil selection.

Two independent stations let a laboratory run two samples at different times without waiting for each other, which suits mixed routine work.

Scope of delivery and options

Delivered ready to use with two U-tubes (P/N 40546) and two probes (40536). Options on the product page: a cryostat from -15 to 60 °C with a 5 L volume and lid, an insulated tubing set for the cryostat, a Dispensette S Organic digital dispenser 1 to 10 mL with 0.05 mL resolution and a valve, a thermal printer, a printer cable and a 10-spool pack of 440-range printer paper.

Compared with a conventional manual aniline point apparatus, an automatic tester removes the need to judge the cloud point by eye, which is difficult with dark or coloured oils. The sample and aniline are mixed in the U-tube, heated until the mixture becomes clear and then cooled until the turbidity returns, and the instrument records the temperature at which this occurs. The memory keeps 200 results, and the RS232 output lets a laboratory print each run or send it to a LIMS, which supports routine solvent and diesel testing.

Typical uses are the aromatic content check of solvents and kerosene, diesel quality screening and the control of rubber process oils. Keep the aniline in a closed, labelled container, use a dispenser for repeatable volumes and follow the hazard instructions of your site for handling and waste. Clean the U-tubes thoroughly between samples to avoid carry-over.

Who benefits from an automatic twin tester

The NAE 440 suits petroleum and petrochemical laboratories that run aniline point tests as part of routine product control, for example refinery quality control rooms, solvent producers, blending plants and independent testing laboratories. Where the test is performed several times per shift, two independent stations reduce waiting time and allow one sample to be prepared while the other is running. Laboratories that handle a mix of clear distillates and darker products gain most from automatic detection, because the visual judgement of the cloud point becomes less reliable as the sample colour deepens.

Selecting the configuration for your laboratory

The first decision is the temperature range you actually need. Products with aniline points above ambient can be tested with the dry bath alone, while samples whose aniline point lies below ambient need the external cooler or cryostat listed among the options. Review the products in your test schedule and the methods you follow, including whether the mixed aniline point method is required, before deciding which options to include. The dispenser option is worth considering when operators need repeatable volumes of aniline and sample without handling the reagent with open glassware.

Installation and laboratory set-up considerations

Place the instrument on a stable bench with adequate space around it, close to a suitable power outlet and, if a cryostat is used, with room for the cooling unit and its tubing. Because aniline is a hazardous reagent, the workstation should be located in an area with appropriate ventilation and close to the facilities your site requires for chemical handling. Plan where the reagent, samples and waste will be stored so that the workflow from sample preparation to cleaning is short and safe. During commissioning, run reference samples on both stations to confirm that results agree with your previous method and with each other.

Data handling and quality records

Stored results and the RS232 output support traceable records. Decide early whether results will be transferred to a LIMS, so that the connection and data format can be arranged with your IT or laboratory systems team. Combine the instrument records with your own sample identification so that each result can be traced to its batch.

Routine operation and cleaning

Consistent results depend as much on routine as on the instrument. Use the same sample and aniline volumes for every run, keep the U-tubes and probes clean and dry before each test, and allow each station to complete its cycle before removing the tube. Cleaning thoroughly between samples prevents carry-over, which is especially important when switching from a dark product to a clear one. Record any deviations from the normal procedure so that unusual results can be investigated later.

How to request a quotation

Quote the model NAE 440 and order number 40500, the quantity, and list the options you require from those shown above. Mention the standards you work to and the typical products you test, so that the configuration can be checked before the offer is prepared.

Documents and quotation

The brochure is linked here. Seven Star LLC, an independent supplier in Muscat, quotes the tester within its liquid analytical systems range. Aniline is a hazardous reagent, so confirm local handling and supply, the cryostat and the dispenser before ordering.

Datasheets & Documents
  • PDF Brochure Normalab NAE 440 brochure (EN, PDF) Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Normalab designs and manufactures laboratory test instruments and laboratory glassware for petroleum products at its site in Valliquerville, France. Seven Star LLC, an independent supplier in Oman, lists Normalab spare parts: the 60606 circulation pump for the coolant circuit of the NDI 450 automatic distillation analyzer, and the 60448 thermal detection thermocouple. We source Normalab parts by their five-digit reference number.

Normalab instrument families

The manufacturer arranges its instruments by the property tested: flammability, volatility, viscosity, cold flow properties, cleanliness and purity, lubricants, and bitumen, wax and grease. Its glassware line supports the same tests. The methods referenced are ASTM, ISO and IP. Refinery and fuel quality laboratories in the GCC run these instruments daily, which is why the usual request is for a replacement part rather than a complete instrument.

Ordering Normalab parts: what we need

Send the Normalab reference number of the part, the instrument model and its serial number, and a photo of the instrument nameplate. Where the part is a sensor or pump, a photo of the old item with its connector helps the manufacturer confirm the version fitted to your serial number. If you need a complete instrument, name the test method and the products you test.

Frequently Asked Questions

What does the NAE 440 measure?
The aniline point of petroleum products and hydrocarbon solvents, including the mixed aniline point.
How many stations are there?
Two independent, self-contained stations.
What temperature range is supported?
-10 to 150 °C; an external cooling source is needed for temperatures below ambient.
Can it test dark samples?
Yes. It detects clear, coloured and dark products up to 8 on the ASTM colour scale, and visual checking is possible.
Which standards are listed?
ASTM D611, ISO 2977 method 5, ISO 21493:2019 and IP 2.