Michell XTC601-EX1 Explosion-Proof Binary Gas Analyzer
Description
Michell XTC601-EX1 explosion-proof thermal conductivity analyzer
The Michell XTC601-EX1 is the Ex d (flameproof) member of the XTC601 family, built for binary gas measurement in hazardous areas where hydrogen, methane or other flammable gases are involved. Its cast enclosure carries ATEX and UKCA marking II 2 G D Ex db IIB +H2 T6 Gb with Ex tb IIIC T85 °C Db IP66, IECEx equivalents, and cQPSus Class I Division 1 Groups B, C and D T6. A typical duty is monitoring hydrogen purity in turbine generator cooling, hydrogen from electrolysis, or CO2 in biomethane. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.
EX1-specific design points
The sample path carries flame arrestors and the gas ports are 1/8 inch NPT female, with sample flow set to 270 to 330 ml/min instead of the 100 to 500 ml/min of the general-purpose models. The thermal conductivity cell uses a matched thermistor pair in a Wheatstone bridge, one in the sample and one in a sealed reference chamber, with a temperature-controlled isothermal housing. Touch-sensitive HMI through the glass means calibration and adjustment need no hot-works permit. A SIL 2 capable option meeting IEC 61508 can be ordered with a supplementary manual.
XTC601-EX1 technical specifications
| Part number | XTC601-EX1 |
| Hazardous area marking | II 2 G D Ex db IIB +H2 T6 Gb; Ex tb IIIC T85 °C Db IP66 |
| Ambient temperature | +5 to +60 °C (cQPSus +5 to +50 °C) |
| Ta by O-ring | silicone -40 to +60 °C; Viton -15 to +60 °C; Ekraz -10 to +60 °C |
| Gas connections | 1/8 inch NPT female with flame arrestors |
| Sample flow (Ex version) | 270 to 330 ml/min |
| Measured gases | air, argon, methane, CO2, hydrogen, helium or nitrogen (order codes add NMHC and CO) |
| Measurement range | selectable from 0 to 1 percent up to 0 to 100 percent, or 50 to 100 up to 98 to 100 percent |
| Accuracy | better than ±1 percent of range or ±0.05 percent H2 or He, whichever is greater |
| Response T90 | under 50 seconds; under 20 seconds for H2 or He |
| Repeatability / linearity | ±0.2 percent / ±1 percent of range |
| Zero and range stability | ±0.5 percent of range per month each |
| Sample pressure | fixed 0.75 to 1.5 bar A; maximum safe 1 barg |
| Outputs | 2 x 4 to 20 mA, Modbus RTU RS485, 2 changeover alarm relays (250 V, 5 A) |
| Supply | 24 V DC, 1.0 A maximum |
| Wetted materials | 316L stainless steel, borosilicate glass, platinum, plus O-ring |
| Size and weight | 234 x 234 x 172 mm; 9.7 kg; IP66 / NEMA 4 |
How XTC601-EX1 differs from XTC601-GP1 and XTC601-GP2
The EX1 code adds the flameproof enclosure and flame arrestors that GP1 lacks. GP1 has 1/4 inch NPT ports and a 100 to 500 ml/min flow window; EX1 has 1/8 inch ports, a tight 270 to 330 ml/min window and an O-ring-dependent hazardous temperature range. The order sheet’s own example, XTC601-EX1-BABCCBB, describes an Ekraz cell O-ring, hydrogen target in an oxygen background, 0 to 5 percent range, 60 °C cell, SIL 2 capable and cleaned for oxygen service. EX1 is the model to specify whenever H2 or CH4 is the target gas.
Applications of XTC601-EX1 in GCC power and gas plants
The datasheet lists hydrogen coolant in electricity turbines, hydrogen generation by electrolysis, air-separation product quality, syngas production, helium recovery, fuel cell research, and air in argon. For hydrogen-cooled generator duty in regional power stations, the fast T90 of under 20 seconds for hydrogen and the 0 to 100 percent or 90 to 100 percent ranges are directly relevant to purity-alarm settings.
Why a thermal conductivity cell suits XTC601-EX1 duties
In hydrogen-cooled generators and electrolysers, the dangerous failure is a drop in hydrogen purity, so the analyzer must be linear, stable and quick near 100 percent. XTC601-EX1 offers ranges as narrow as 98 to 100 percent and 90 to 100 percent, linearity of ±1 percent of range, and zero and range stability of ±0.5 percent of range per month. The sensor design largely eliminates the drift associated with comparable devices, and the cell has no moving parts, so vibration at a turbine hall or offshore module does not upset it. XTC601-EX1 can mount close to the sample point, giving faster response and less sample line.
Configuring XTC601-EX1
Choose cell O-ring (Viton, Ekraz), target gas (hydrogen, helium, methane, argon, CO2, nitrogen, NMHC or CO), background gas (nitrogen, oxygen, air, argon, CO2), range, cell temperature 50 or 60 °C, SIL option and oxygen cleaning. Cleaning for oxygen service is standard for oxygen applications. Spares listed include blanking plugs, M20 conduit entries, M20 compression and barrier glands, a lid O-ring and tube adaptors.
Installation and maintenance of XTC601-EX1
Plan the installation around three practical limits that the datasheet imposes: a non-condensing sample with particles under 3 µm, a fixed pressure window with no pulsation, and a constant gas temperature. A small filter and pressure regulator upstream of the flame arrestors normally cover all three, and a flowmeter with a low-flow alarm in the sample line protects against a blocked line that would otherwise leave a stale reading on the display. Keep sample temperature constant (+5 to +45 °C, or +55 °C where allowed), pressure fixed between 0.75 and 1.5 bar A, and calibrate at that same pressure. The maximum safe pressure is 1 barg. Observe the ambient temperature limit for the chosen O-ring. Power from 24 V DC, 1.0 A maximum, and connect the Modbus RTU line for remote configuration. Warm-up is under 25 minutes and stabilisation 5 minutes. Recalibration and extended warranty of 24 or 36 months are available as options.
Documents and quotation
The XTC series datasheet and order codes are provided as PDF. For a quotation on XTC601-EX1, send an RFQ stating the part number, the complete option string, quantity and certificate requirements such as a BS EN 10204 3.1 material certificate.
Other Models in the Michell Xtc601 Range
Seven Star LLC supplies models from the Michell Xtc601 range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.
| Model | Category | |
|---|---|---|
| Michell XTC601-EX1 Explosion-Proof Binary Gas Analyzer | Gas Analyzer Systems | You are viewing this model |
| Michell XTC601-GP1 Binary Gas Analyzer, General Purpose | Gas Analyzer Systems | View details |
Datasheets & Documents
- Datasheet Michell XTC601-EX1 Explosion-Proof Binary Gas Analyzer Datasheet Download
Reviews (0)
How would you rate this product? Cancel reply
About brand
Michell Instruments is a British manufacturer of humidity and dew point instruments, founded in 1974 in Cambridge, England, by Andrew Michell and incorporated on 12 September 1974. It worked with Cambridge University scientists on an aluminium oxide dew point sensor and later developed chilled mirror hygrometers and hydrocarbon dew point analysers. Michell is based in Ely, Cambridgeshire, and is part of Process Sensing Technologies (PST).
Seven Star LLC lists Michell dew-point transmitters, the Condumax II analyzer and its replacement sensor parts, and supplies them to gas plants and compressed air users in Oman by exact order code.
MICHELL Instruments company profile
| Founded | 1974, Cambridge, England, by Andrew Michell |
|---|---|
| Headquarters | Ely, Cambridgeshire, United Kingdom, since 2007 |
| Ownership | Process Sensing Technologies (PST) |
| Technology | Ceramic and aluminium oxide moisture sensors, chilled mirror dew point, Condumax hydrocarbon dew point (licensed from Shell) |
| Product families | Easidew and Promet dew point transmitters, S8000 and Optidew chilled mirror hygrometers, Condumax II hydrocarbon dew point analysers, MDM300 portable dew point meters, oxygen analysers |
Milestones
- 1974: Andrew Michell founds Michell Instruments in Cambridge.
- 2007: The company moves to Ely, Cambridgeshire.
- 2024: Michell marks 50 years, now as part of PST.
Michell product families on this page
The Condumax II is an automatic hydrocarbon dew-point analyzer for natural gas. Several of its parts are listed separately: the CD2-SCR-002 hydrocarbon dew-point sensor cell, the CD2-WDC-002 water dew-point sensor assembly and the CD2-SX-WDC-002, which is the exchange version of that water dew-point sensor. The Easidew line is represented by the Easidew EA2 dew point transmitter and by the Easidew PRO I.S., an intrinsically safe moisture transmitter for hazardous areas. A ceramic impedance moisture sensor module is also listed, the sensing element behind these transmitters.
Where Michell analyzers are used
Hydrocarbon and water dew point are contract quality parameters for natural gas, so the Condumax II is found at gas processing plants, pipeline entry points and custody transfer stations. Easidew transmitters monitor dryers for instrument air and industrial gases, glove boxes, and dry gas feeds to processes that cannot tolerate water.
MICHELL Instruments products we list
Ordering Michell parts: what we need
For Condumax II parts send the analyzer serial number and the part code, such as CD2-SCR-002, and state if a new assembly or the exchange item is wanted. For Easidew transmitters we need the measuring span in dew point or ppm, the process connection thread, the output signal, the operating pressure, the gas and the area certification. A photo of the label on the old transmitter usually answers most of these in one go.
Facts checked against UK Companies House records and PST's Michell 50-year material, October 2026.
Browse MICHELL Instruments products by category: Gas Analyzer Accessories, Gas Analyzer Systems.
FAQ
Questions about the Michell XTC601-EX1 Explosion-Proof Binary Gas
Answers from our engineering team on installation, compatibility and integration.
01Which approvals does XTC601-EX1 carry?
The datasheet lists ATEX/UKCA II 2 G D Ex db IIB +H2 T6 Gb, Ex tb IIIC T85 °C Db IP66, IECEx equivalents and cQPSus Class I Division 1 Groups B, C, D T6.
02Can XTC601-EX1 measure hydrogen?
Yes. Hydrogen and methane target gases require an explosion-proof analyzer or one with flame arrestors, and XTC601-EX1 is the flameproof version with flame arrestors.
03What flow does XTC601-EX1 need?
The Ex version needs 270 to 330 ml/min through the cell, with a fixed pressure of 0.75 to 1.5 bar A and maximum safe pressure of 1 barg.
04Does ambient limit depend on the O-ring in XTC601-EX1?
Yes. The hazardous-area ambient limits are -40 to +60 °C with silicone, -15 to +60 °C with Viton and -10 to +60 °C with Ekraz O-rings.
05Is a SIL option available for XTC601-EX1?
Yes. The XTC601 can be ordered as SIL 2 capable under IEC 61508, with a supplementary SIL manual supplied for the functional safety system.


















Reviews
There are no reviews yet.