Michell TDL600-SYS OptiPEAK Natural Gas Moisture Analyzer
Description
Michell OptiPEAK TDL600-SYS moisture in natural gas analyzer
The Michell TDL600-SYS is the parent order code of the OptiPEAK TDL600, a tunable diode laser absorption spectroscopy (TDLAS) analyzer that measures water vapour in natural gas and biomethane from 1 to 1000 ppmV. It is supplied as a complete system: an explosion-proof aluminium analyzer head with a 4.3 inch colour touchscreen, plus an integrated sample-conditioning system on an indoor panel or in an outdoor stainless enclosure. Because the measurement is optical and non-contact, the datasheet says maintenance is minimal even with sour gas or changing methane content. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.
How the TDL600-SYS measures
The analyzer uses a dual-pass cell, avoiding the high top-end range limit of Herriott multi-pass cells that can saturate quickly at higher moisture. Continuous laser optimisation keeps the laser locked to the correct water absorption peak, a multi-stage control holds laser temperature within tight limits, and a built-in self-verification checks itself against calibration data on every update cycle. The D-MET system adds dynamic background-gas compensation, so readings are almost independent of changes in methane level without manual correction factors; shale gas and biomethane are named as intended streams. For gas with more than 2.5 percent CO2 or H2S, a special calibration can be applied.
TDL600-SYS specification table
| Part number | TDL600-SYS |
| Technology | tunable diode laser spectroscopy (TDLAS), dual-pass cell |
| Measurement range | 1 to 1000 ppmV |
| Accuracy | ±1 ppmV or ±3 percent of reading, whichever is greater |
| Repeatability | ±0.5 ppmV or ±1.5 percent of reading, whichever is greater |
| Limit of detection | 1 ppmV |
| Units | ppmV, lb/MMSCF, mg/Nm3, dew point (ISO 18453 or IGT#8) |
| Display update | 2 to 3 seconds |
| Factory calibration | 3-point, traceable to NPL and NIST |
| Maximum inlet pressure | 100 barg (1450 psig) |
| Sample flow | 0.5 Nl/min through the cell; 1 to 5 Nl/min filter by-pass |
| Power | 110 V AC or 230 V AC, 50/60 Hz; 180 W indoor, 180 to 380 W outdoor |
| Outputs | 3 x 4 to 20 mA, 3 alarms (250 V AC, 3 A), RS485 Modbus RTU |
| Hazardous area | ATEX/UKEX II 2 G Ex db ib op is IIC T5 Gb; IECEx; cQPSus Class I Division 1 |
| Weight | 40 to 75 kg depending on configuration |
How TDL600-SYS differs from other OptiPEAK and Michell moisture analyzers
Several options are chosen through letters in the order code. Feature B selects ATEX/IECEx (B1) or North American Class 1 Division 2 at 110 V AC (B2). Feature C selects calibration: C1 is the standard 1 to 1000 ppmV range with points at 5, 50 and 100 ppmV and less than 2.5 percent CO2, C91 suits 2.6 to 5 percent CO2 and C92 suits 5.1 to 7 percent CO2. Sampling can be indoor or outdoor, with or without armoured flowmeters rated 130 barg; outdoor systems add heaters (20 °C or 40 °C fixed, 110 or 240 V), optional enclosure cooling for 45 °C ambient, a 3 m heat-traced sample line, a validation gas inlet with a three-way valve, and a 100 or 160 barg line-pressure transmitter. Compared with quartz-microbalance or ceramic-sensor analyzers from the same maker, the TDL600 is a non-contact method with a 1 ppmV detection limit.
Applications of TDL600-SYS for Oman and Gulf gas networks
Natural gas glycol dehydration, transmission monitoring, custody transfer, storage, offshore export pipelines, LNG receiving and regasification, vaporised NGL and biomethane are all named in the datasheet, and it can handle natural gas that contains up to 20 percent hydrogen without modification. Regional gas operators use this class of analyzer to prove that export gas meets water-content limits and that glycol contactors are performing. Additional approvals include TRCU 012, PESO, NEPSI, Ukraine Ex, Kazakhstan GOST-K and ECAS (UAE).
Sample conditioning supplied with TDL600-SYS
A well designed sampling system decides whether a trace-moisture reading can be trusted, so each TDL600-SYS is delivered with a high-quality, in-house designed conditioning system using state-of-the-art filtration and multi-stage pressure reduction to present a clean, wholly gaseous sample. Adsorption and desorption of water on tube walls matters at single- or double-digit ppm levels, and the manufacturer stresses that sample-system design and finish decide the performance of the whole measuring system. The enclosure is 304L or 316L stainless steel, the filter by-pass accelerates sample transport, and the validation gas inlet lets operators challenge TDL600-SYS with a known gas without breaking the sample line.
Planning installation of TDL600-SYS
Indoor versions operate from +10 to +45 °C, outdoor from -20 to +45 °C and outdoor with cooling from -20 to +55 °C, and an outdoor analyzer must be shaded from direct sunlight. Cell vent pressure must be 0.7 to 1.4 bara and the filter by-pass at 3 barg maximum. The sample panel measures 970 x 750 mm for indoors, and the outdoor enclosure is 1045 x 800 mm. Electrical entries are three M20 glands, and gas connections are 1/4 inch NPT female.
Operation and service of TDL600-SYS
In day-to-day use TDL600-SYS needs little more than housekeeping. A colour capacitive-touch HMI allows field interrogation without a hot-work permit and without a stylus, and remote application software is supplied as standard for configuration. Data logging captures all process variables at a user-selected period of 10 seconds to one day. Regular field calibration is not required under normal operating conditions, though user or quality-system rules may call for periodic checks using the optional validation gas inlet.
Documents and quotation
The OptiPEAK TDL600 datasheet and order-code sheet are available as PDF references. For a quotation on TDL600-SYS, send an RFQ giving the part number, the full option string (B, C, sampling type, heater, cooling, heat trace, validation gas, pressure transmitter, material certification), quantity and site location.
Typical Questions From Maintenance Teams
Maintenance engineers replacing a MICHELL Instruments TDL600-SYS usually ask whether the new unit is a direct fit, whether settings can be copied from the old one, and whether any accessories must be ordered separately. The answer depends on the exact order code and the generation of the installed analyzer systems, so the nameplate details are the best starting point. Send them with your request and we will check the match, point out any differences in connections or options, and list the parts needed so that the job can be finished in one visit.
Datasheets & Documents
- Datasheet Michell TDL600-SYS OptiPEAK Natural Gas Moisture Analyzer Datasheet Download
Reviews (0)
How would you rate this product? Cancel reply
About brand
Michell Instruments is a British manufacturer of humidity and dew point instruments, founded in 1974 in Cambridge, England, by Andrew Michell and incorporated on 12 September 1974. It worked with Cambridge University scientists on an aluminium oxide dew point sensor and later developed chilled mirror hygrometers and hydrocarbon dew point analysers. Michell is based in Ely, Cambridgeshire, and is part of Process Sensing Technologies (PST).
Seven Star LLC lists Michell dew-point transmitters, the Condumax II analyzer and its replacement sensor parts, and supplies them to gas plants and compressed air users in Oman by exact order code.
MICHELL Instruments company profile
| Founded | 1974, Cambridge, England, by Andrew Michell |
|---|---|
| Headquarters | Ely, Cambridgeshire, United Kingdom, since 2007 |
| Ownership | Process Sensing Technologies (PST) |
| Technology | Ceramic and aluminium oxide moisture sensors, chilled mirror dew point, Condumax hydrocarbon dew point (licensed from Shell) |
| Product families | Easidew and Promet dew point transmitters, S8000 and Optidew chilled mirror hygrometers, Condumax II hydrocarbon dew point analysers, MDM300 portable dew point meters, oxygen analysers |
Milestones
- 1974: Andrew Michell founds Michell Instruments in Cambridge.
- 2007: The company moves to Ely, Cambridgeshire.
- 2024: Michell marks 50 years, now as part of PST.
Michell product families on this page
The Condumax II is an automatic hydrocarbon dew-point analyzer for natural gas. Several of its parts are listed separately: the CD2-SCR-002 hydrocarbon dew-point sensor cell, the CD2-WDC-002 water dew-point sensor assembly and the CD2-SX-WDC-002, which is the exchange version of that water dew-point sensor. The Easidew line is represented by the Easidew EA2 dew point transmitter and by the Easidew PRO I.S., an intrinsically safe moisture transmitter for hazardous areas. A ceramic impedance moisture sensor module is also listed, the sensing element behind these transmitters.
Where Michell analyzers are used
Hydrocarbon and water dew point are contract quality parameters for natural gas, so the Condumax II is found at gas processing plants, pipeline entry points and custody transfer stations. Easidew transmitters monitor dryers for instrument air and industrial gases, glove boxes, and dry gas feeds to processes that cannot tolerate water.
MICHELL Instruments products we list
Ordering Michell parts: what we need
For Condumax II parts send the analyzer serial number and the part code, such as CD2-SCR-002, and state if a new assembly or the exchange item is wanted. For Easidew transmitters we need the measuring span in dew point or ppm, the process connection thread, the output signal, the operating pressure, the gas and the area certification. A photo of the label on the old transmitter usually answers most of these in one go.
Facts checked against UK Companies House records and PST's Michell 50-year material, October 2026.
Browse MICHELL Instruments products by category: Gas Analyzer Accessories, Gas Analyzer Systems.
FAQ
Questions about the Michell TDL600-SYS OptiPEAK Natural Gas Moisture
Answers from our engineering team on installation, compatibility and integration.
01What does TDL600-SYS measure?
TDL600-SYS is the OptiPEAK TDL600 system that measures moisture in natural gas and biomethane from 1 to 1000 ppmV using tunable diode laser absorption spectroscopy.
02What accuracy does TDL600-SYS achieve?
The datasheet gives ±1 ppmV or ±3 percent of reading, whichever is greater, with repeatability of ±0.5 ppmV or ±1.5 percent and a 1 ppmV detection limit.
03Is TDL600-SYS certified for hazardous areas?
Yes. The marking is ATEX/UKEX II 2 G Ex db ib op is IIC T5 Gb, with IECEx and cQPSus Class I Division 1 Groups A to D, plus several national approvals.
04What inlet pressure can TDL600-SYS handle?
Maximum inlet pressure is 100 barg (1450 psig). Armoured flowmeters rated 130 barg and 100 or 160 barg pressure transmitters are among the order options.
05Does TDL600-SYS need field calibration?
The datasheet says no regular field calibration is required under normal operating conditions; a validation gas inlet with a three-way valve is available as an option.


















Reviews