Brand:

Beamex 8003194 PGHH Reservoir Replacement Kit

Description

Beamex 8003194 at a glance

Beamex 8003194 is the PGHH reservoir replacement kit. It contains the acrylic reservoir that holds the pressure medium of the Beamex PGHH hand pump, two 67 x 3 mm O-rings that seal it, and a product label to stick on the new reservoir. The reservoir volume is 200 ml / 12.2 cubic inches. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.

The reservoir is the only transparent part of the pump, and the manual stresses that it is not to be pressurized. A cracked or clouded reservoir therefore has to be exchanged rather than repaired.

Typical symptoms that lead to an order are fluid seeping around the reservoir base, visible hairline cracks, or a body so clouded that the oil level can no longer be judged against the two-thirds and three-quarters marks of correct filling.

Specification summary

Item Detail
Part number 8003194
Kit name PGHH reservoir replacement kit
Reservoir material Acrylic (glass-like)
Reservoir volume 200 ml / 12.2 cubic inch
Sealing 2 x O-ring 67 x 3 mm
Label Product label for the reservoir included
Maximum output pressure printed on label 700 bar
Allowed fluids Mineral oil RENOLIN MR 5 VG 22 or distilled water
Filling level At least two thirds and at most three quarters
Fit Beamex PGHH pump

Compared with a generic acrylic tube, the kit reservoir is made to match the pump: it carries the printed label stating 0 to 700 bar / 10000 psi hydraulic calibration pump, model PGHH, with the warning not to pressurize the glass reservoir. Reusing the label on the new part keeps that warning visible to every user of the pump.

Why reservoirs fail

The manual gives three reasons. Impurities from the device under test can flow back and cloud or attack the acrylic; an unsuitable fluid such as synthetic hydraulic oil can damage the reservoir and seals; and a blocked pressure compensation screw can stop replacement air from entering as the fluid level falls, which may break the reservoir. Any of these can lead to a leak or a crack.

Chemical cleaners are also excluded. The maker allows only a wet, lint-free cloth, or a flush with distilled water or the approved low viscosity oil.

Fitting 8003194 step by step

  1. Depressurize the pump and empty the reservoir through the filling plug and drain screw.
  2. Open the pump following the maker’s video for this kit, which is meant for skilled personnel.
  3. Remove the old reservoir and O-rings, clean the seats and lightly coat the new O-rings with silicone grease.
  4. Seat the new reservoir and tighten connections to the manual torque of 25 Nm, keeping the drain screw to a maximum of 15 Nm.
  5. Fix the label, refill to the correct level and run the leak test.

After fitting 8003194, keep the pump upright for the whole session so air can move through the compensation screw.

How 8003194 differs from related PGHH parts

Unlike the 8003192 pressure cylinder, this kit has no check valves or suction hose; it is a smaller job that touches only the reservoir seats. Unlike the 8003199 stroke selector kit, it is about fluid containment, not handle travel. Its O-rings are 67 x 3 mm, whereas the vent valve 8003187 uses a bonded seal and a 6 mm steel ball. Choose 8003194 when the cylinder body is cracked, cloudy beyond cleaning, or leaking at its base.

Use in calibration workshops

High-pressure hydraulic pumps are common in instrument shops serving refineries, gas plants and power stations, where pressure transmitters, gauges and EXT modules are tested far above the reach of pneumatic pumps. Being able to renew the reservoir in the workshop keeps the pump on the bench.

Distilled water users in particular should expect to inspect the reservoir often, since the manual requires the pump to be emptied and dried after each use with water.

Care and handling after replacement

Fill with at least two thirds and no more than three quarters of the reservoir volume. Do not top up while pressure is rising, because the reservoir may overflow. If clouds, particles or dirt appear in the fluid, change it. Never use the pump in direct sunlight, near hot objects, or in explosive atmospheres, as listed in the manual.

Reservoir care: ordinary checks that prevent failure

Before each session look at the fluid in the reservoir. Clear fluid is normal; the first sign of cloudiness, particles or dirt is the cue to change it. After a session on distilled water, unscrew the reservoir filling plug and the drain screw at the bottom of the reservoir, empty it and let the pump dry.

The pump is shipped after testing with WACKER AK 10 silicone fluid, so small residues may remain; flush with the medium you intend to use before connecting instruments. The manual also asks users to flush the device under test first if possible, because process residues carried into the pump can attack the seals, O-rings and reservoir.

Never leave the pump pressurized when unattended, and never drop it: the manual warns that an accidentally dropped pump may be damaged, and an acrylic reservoir is the part most exposed to impact.

Ordering information

Quote the pump serial number and the symptom, such as a hairline crack, a leaking base seal or a clouded body that will not clear after flushing. If the pump will go to a service centre instead, empty the reservoir first, since the manual warns that additional waste handling fees may be charged for fluid left inside.

Documents and quotation

The PGHH user manual is included as the datasheet for Beamex 8003194. To ask for a quotation, send an RFQ to Seven Star LLC with part number 8003194, quantity and pump serial number so the kit can be confirmed.

Other Models in the Pghh Spares Range

Seven Star LLC supplies models from the Pghh Spares range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Beamex 8003187 PGHH Vent Valve Spare AssemblyProcess, Pressure & Temperature CalibratorsView details
Beamex 8003192 PGHH Pressure Cylinder Service KitProcess, Pressure & Temperature CalibratorsView details
Beamex 8003194 PGHH Reservoir Replacement KitProcess, Pressure & Temperature CalibratorsYou are viewing this model
Beamex 8003199 PGHH Stroke Selector Service KitProcess, Pressure & Temperature CalibratorsView details
Datasheets & Documents
  • PDF Datasheet Beamex 8003194 PGHH Reservoir Replacement Kit Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Beamex is a Finnish maker of calibration equipment and software for process instruments. Four calibration technicians founded the company in 1975 because they could not find the equipment they wanted; their first product was a voltage and current simulator, and the first factory opened in 1977. Beamex is headquartered in Pietarsaari (Jakobstad), Finland, and supplies process industries in more than 100 countries.

Seven Star LLC is an independent supplier listing Beamex documenting calibrators, temperature dry blocks, pressure controllers, calibration pumps and pressure modules for buyers in Oman and the GCC. Equipment is quoted against the model and the pressure or temperature ranges required.

Beamex company profile

Origin Founded in 1975 in Finland by four calibration technicians; first factory 1977
Headquarters Pietarsaari (Jakobstad), Finland; ISO 9001 certified since 1992
Product families MC series multifunction calibrators, temperature dry blocks, automatic and hand pressure generators, external pressure modules, CMX and LOGiCAL calibration software
Typical duty Field and workshop calibration of pressure, temperature and electrical signal instruments, including HART and fieldbus transmitters

Milestones

  • 1975: Four calibration technicians found the company in Finland.
  • 1977: The first factory opens.
  • 1992: ISO 9001 certification.

Beamex calibrators and accessories

Model Type Typical use
MC6 Advanced field calibrator and communicator Pressure, temperature and electrical calibration with a built-in HART, FOUNDATION Fieldbus and PROFIBUS PA communicator, documented results
MC6-Ex Intrinsically safe field calibrator Calibration inside Zone 0 and Zone 1 hazardous areas without a hot work permit
MC6-T Temperature calibrator with integrated dry block Temperature sensors and transmitters, with multifunction electrical and pressure measurement
MC4, MC2 Documenting and hand-held process calibrators Routine loop, transmitter and switch checks in the field
FB, MB Field and metrology temperature dry blocks FB for field calibration; MB for higher-accuracy workshop and laboratory work
ePG, POC8 Electric pressure pump and automatic pressure controller Automated pressure calibration; POC8 for workshop calibration benches
PGC, PGM, PGL, PGPH, PGHH Hand pressure pumps Pressure and vacuum generation in the field, from low pressure (PGL, ±400 mbar) to hydraulic 700 bar (PGHH)
EXT modules External pressure modules Extending MC calibrators to additional ranges, 70 kPa to 100 MPa

Where Beamex equipment is used

Instrument workshops and maintenance teams in oil and gas, refining, petrochemicals, power, water and pharmaceuticals use Beamex calibrators to verify transmitters, switches, gauges and temperature sensors. Documenting calibrators send results to calibration software, which keeps the audit trail required by quality systems and custody transfer metering procedures. Intrinsically safe models are used for in-situ calibration in classified areas.

Ordering Beamex equipment: what we need

State the model and the options you need: internal and external pressure ranges, temperature range, communication protocols (HART, FOUNDATION Fieldbus, PROFIBUS PA), Ex approval, and whether an accredited calibration certificate is required. For spare modules and pumps send the serial number of the calibrator they will be used with, so compatibility can be confirmed.

Facts checked against Beamex company history and product information, October 2026.

FAQ

Questions about the Beamex 8003194 PGHH Reservoir Replacement Kit

Answers from our engineering team on installation, compatibility and integration.

01What is in the Beamex 8003194 kit?
One acrylic 200 ml reservoir, two 67 x 3 mm O-rings and a product label to place on the new reservoir.
02Is the PGHH reservoir pressurized?
No. The manual states the 200 ml reservoir is not to be pressurized, and the label reads do not pressurize glass reservoir.
03Which fluids can I put in the new reservoir?
Low viscosity mineral hydraulic oil such as RENOLIN MR 5 VG 22 or distilled water; other fluids can damage seals and the reservoir.
04What reservoir fill level does the manual recommend?
Fill to at least two thirds and at most three quarters of the reservoir volume.
05Does 8003194 include check valves?
No. Check valves are part of the 8003192 pressure cylinder kit; this kit only renews the reservoir, O-rings and label.