Brand: Palmer Wahl
Palmer Wahl C150 Documenting Multifunction Calibrator
Description
A documenting multifunction calibrator for temperature, process and pressure
The Palmer Wahl C150 is the top instrument of the C series of calibrators. The datasheet describes it as an advanced documenting multifunction calibrator that works as a simulator (in and out) and as a dual channel thermometer (in and in) for comparison calibration. It measures and generates temperature, resistance, process, pressure and frequency signals over two isolated channels, calibrates thermistors, and has a HART communicator integrated into the ACL500 modem for HART transmitter calibration.
The C150 stores audit trails and the datasheet states that it complies with both 21 CFR Part 11 and the NADCAP Heat Treatment standard AMS 2750. Those statements are quoted from the datasheet as printed. Ten calibration procedures can be stored on the instrument, results are saved on site and uploaded later with DATACAL software to issue calibration certificates.
Features
- Temperature accuracy up to 0.005 percent of reading; resistance up to 0.006 percent of reading with a 50 kohm range; current up to 0.007 percent of reading and 100 mA range with a 24 V loop supply; voltage up to 0.005 percent of reading and 50 V range; frequency up to 0.01 percent of reading and 100 kHz range.
- Thermocouples (14 types) and RTDs (12 types), plus thermistors measured with the Steinhart-Hart equation on the 50 kohm range.
- Pressure measurement and simulation with the external digital pressure module ACL433, in absolute and relative ranges to 1000 bar absolute, with accuracy of 0.05 percent full scale from 10 to 40 C.
- Easy connect terminals accept wires to 3 mm or 10 AWG, 4 mm banana plugs and compensated thermocouple connectors.
- Memory of 10,000 data points, recalled on screen as a curve or a list; scaling, ramps, steps, synthesizer, square root and statistical functions.
- IP54 case in an anti-shock rubber holster with a wide backlit display; can drive baths and dry-blocks with the ACL600 cable.
C150 specifications
| Parameter | Specification |
|---|---|
| Channels | Two isolated channels; simulator (in/out) and dual channel thermometer (in/in) |
| Thermocouple types | 14 types; K, T, J, R, S, B, U and N tabulated with PlatineL, Mo, NiMo/NiCo, G, D, L and C in the manual |
| Thermocouple K accuracy | 0.005 percent of reading as the top figure on the datasheet comparison |
| Thermistor input | 50 kohm range with Steinhart-Hart equation |
| Current measure range | 0 to 24 mA; 4-20 mA from 3 to 24 mA; 100 mA; loop supply 24 V plus or minus 10 percent |
| HART | Integrated communicator via ACL500 modem; input impedance 280 ohm |
| Pressure module | ACL433 digital module on RS485, 0-1 to 0-1000 bar absolute, 0-1 to 0-30 bar relative as tabulated |
| Memory | 10,000 data points |
| Software | DATACAL calibration software, USB cable supplied |
| Protection | IP54 in anti-shock rubber holster |
| Standards stated | 21 CFR Part 11 and NADCAP heat treatment AMS 2750 |
| Supplied with | Quick start manual, battery charger, six test leads, carrying strap, factory test report, certificate of conformance |
| Packing information | 210 x 110 x 50 mm; 900 g (as packed) |
| Datasheet date | 05-08-2022 |
Resistive probe accuracy
The datasheet gives these measurement specifications at 23 C plus or minus 5 C with 4-wire probes; accuracy is in percent of reading plus a fixed value, and resolution is 0.01 C except for Cu10 at 0.10 C.
| Sensor | Range | Accuracy per 1 year |
|---|---|---|
| Pt50 (3851) | -220 to 850 C | 0.006% rdg + 0.04 C |
| Pt100 (3851) | -220 to 850 C | 0.006% rdg + 0.03 C |
| Pt100 (3916) | -200 to 510 C | 0.006% rdg + 0.03 C |
| Pt100 (3926) | -210 to 850 C | 0.006% rdg + 0.03 C |
| Pt200 (3851) | -220 to 850 C | 0.006% rdg + 0.04 C |
| Pt500 (3851) | -220 to 850 C | 0.006% rdg + 0.03 C |
| Pt1000 (3851) | -220 to 740 C | 0.006% rdg + 0.03 C |
| Ni100 (618) | -60 to 180 C | 0.006% rdg + 0.05 C |
| Ni120 (672) | -40 to 205 C | 0.006% rdg + 0.05 C |
| Ni1000 (618) | -60 to 180 C | 0.006% rdg + 0.05 C |
| Cu10 (427) | -50 to 150 C | 0.006% rdg + 0.18 C |
| Cu50 (428) | -50 to 200 C | 0.006% rdg + 0.05 C |
Process ranges
| Function | Ranges | Accuracy per 1 year |
|---|---|---|
| DC current measure | 0-20 mA, 4-20 mA (3 to 24 mA), 100 mA | 0.007% rdg + 0.8 uA (100 mA range 0.009% rdg + 2 uA) |
| DC current source | 24 mA, 4-20 mA, 0-20 mA | 0.007% rdg + 0.8 uA |
| DC voltage measure | 100 mV, 1 V, 10 V, 50 V | 0.005% rdg + 2 uV to 0.007% rdg + 0.5 mV |
| DC voltage source | 100 mV, 1 V, 10 V, 50 V | 0.005% rdg + 2 uV to 0.007% rdg + 0.5 mV |
| Frequency measure | 10 kHz, 100 kHz | 0.01% rdg |
| Frequency source | 1000 Hz, 100 kHz | 0.01% rdg |
| Resistance measure | 400 ohm, 3600 ohm, 50 kohm | 0.006% rdg + 8 mohm; 0.006% rdg + 50 mohm; 0.008% rdg + 1 ohm |
| Resistance source | 400 ohm, 3600 ohm | 0.006% rdg + 20 mohm; 0.006% rdg + 100 mohm |
Pre-programmed steps include 4-20 mA and 0-20 mA linear and quadratic and a valve test sequence, which suit transmitter and valve positioner checks.
C series comparison and selection
- The datasheet compares C50, C75, C100 and C150; the C150 is the only member with the 100 mA range, the 50 kohm range and the 100 kHz frequency range printed in its column.
- Choose it when you need documenting calibration with audit trails, HART transmitter calibration and dual channel comparison, and a smaller C-series model if you only need simulation.
- Order the ACL433 module separately for pressure work, and the ACL600 cable if you drive baths or dry-blocks; call for the full accessory list.
- Use the DATACAL software to configure the unit and to manage and export results.
Applications in Oman and the GCC
Instrument technicians in refineries, power stations and heat treatment shops use one portable calibrator to check RTD and thermocouple transmitters, 4-20 mA loops, pressure transmitters and process indicators. The AMS 2750 statement matters for aerospace heat treatment work, so check the requirement with your auditor before relying on it.
Selection and installation notes
A multifunction calibrator is selected from the signals and sensors you need to measure and simulate. Check the ranges and sockets, the pressure modules, the leads and the battery before use, let the instrument warm up, and enter the device settings correctly before documenting a result. Keep the calibration certificate on file and plan the re-calibration interval according to your quality system.
Information needed for a quotation
Send the full model code, the quantity, the delivery location in Oman or the GCC, the intended use and any accessories needed, so that the complete package can be quoted together.
Documents and supply information
The manufacturer data sheet is attached to this page and should be read together with the project specification and the site risk assessment before a purchase is confirmed. Ratings, standards and dimensions are quoted as printed on that sheet. Lead time, stock position and certificates available with the shipment are confirmed in the quotation, and we supply to Oman and the wider GCC region for industrial, utility and building services projects.
Related items and documents
See more portable test and measurement instruments and calibration items in our range. The datasheet is attached; Seven Star LLC will quote the C150 with the accessories you need.
Datasheets & Documents
- Datasheet Palmer Wahl C150 advanced documenting multifunction calibrator datasheet (05-08-2022) Download
Reviews (0)
How would you rate this product? Cancel reply
About brand
Palmer Wahl Instrumentation Group is an American maker of thermometers, temperature and pressure calibrators and infrared instruments, based in Asheville, North Carolina, since 1987. It traces its roots to 1836, when James Foster, Jr. & Co. began making thermometers in Cincinnati. Charles Palmer introduced his red-reading mercury thermometer in 1929, and Palmer acquired Wahl Instruments in 1997.
Seven Star LLC is an independent supplier in Muscat. We list Palmer Wahl thermometers, calibrators and test instruments for maintenance and calibration teams in Oman and the GCC.
Palmer Wahl company profile
| Origins | 1836, Cincinnati, Ohio, as James Foster, Jr. & Co. |
|---|---|
| Headquarters | Asheville, North Carolina (since 1987) |
| Key product | Red-reading mercury thermometer, Charles Palmer, 1929 |
| Merger | Acquired Wahl Instruments, 1997 |
| Product families | Bimetal and industrial dial thermometers, Digi-Stem RTD thermometers, documenting multifunction calibrators, pocket RTD and thermocouple calibrators, portable black body sources, infrared thermometers |
Milestones
- 1836: Thermometer making begins in Cincinnati.
- 1929: Charles Palmer introduces the red-reading thermometer.
- 1987: The company moves to Asheville.
- 1997: Palmer acquires Wahl Instruments.
Palmer Wahl product series
Thermometers include the 1050 direct-drive 5-inch dial thermometer, the All Star bimetal thermometers in 3 and 5 inch sizes, the 9AL and 9FL economy industrial thermometers and the DST600 and DST640 Digi-Stem RTD thermometers. The DHS520L Heat Spy is a high-temperature infrared thermometer.
Calibration and test equipment includes the TC622A pocket RTD calibrator, the C150 documenting multifunction calibrator, the P32 handheld process signal calibrator, the HSICBB-P portable black body for checking infrared thermometers and the DSPG digital pressure gauge data logger.
Where Palmer Wahl instruments are used
Temperature indication on tanks, pipes and process equipment, field calibration of loops and temperature transmitters, infrared temperature checks on furnaces and kilns, and pressure logging during tests.
Palmer Wahl products we list
Ordering Palmer Wahl instruments: what we need
Send the instrument family, range, stem or probe length, process connection, case style and any calibration certificate or mercury-free requirement. The datasheet ordering guides allow the full code to be confirmed before an offer is issued.
Facts checked against Palmer Wahl's own company history (palmerwahl.com), October 2026.
Browse Palmer Wahl products by category: Portable Test & Measurement, Temperature Instrumentation.
Related products
Frequently Asked Questions
Temperature, resistance, process, pressure and frequency signals over two isolated channels, plus thermistors and HART transmitters.
Yes. The datasheet says a HART communicator is integrated into the ACL500 modem.
Through the external ACL433 digital pressure module, which connects by RS485 to the digital input.
The memory holds 10,000 data points, shown on screen as a curve or a list.
No. Seven Star LLC is an independent supplier in Muscat that sources genuine products on request; we do not claim any distributor status.