The Michell ES20-M is the metric (6 mm tube and fittings) base model of the ES20 compact sampling system, a modular stainless-steel panel that prepares a gas sample for a dew-point transmitter or hygrometer at inlet pressures up to 20 barg. It brings filtration, pressure control, flow control and an optional display together for compressed air, instrument air, pneumatics, plastic moulding, hydrogen coolant, breathing air, inert gas and heat-treatment gas lines. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.

What is on the ES20-M panel

A 316 stainless steel base plate carries inlet and outlet fittings in 6 mm tube, a needle-type metering valve, a 316 stainless filter housing with a borosilicate-microfibre cartridge, a dual-scale bar and psi pressure gauge, a 5/8 inch 316 stainless sensor sample block, and a borosilicate glass flowmeter with Viton seals reading 2 to 12 Nl/min. Flow through the sensor is kept at 1 to 5 litres per minute. A terminal rail with 13 terminals and a 1/8 DIN bracket for a process indicator are fitted when a monitor or pump is included.

ES20-M specification table

Part number ES20-M (metric 6 mm tube and fittings)
Inlet pressure up to 20 barg (290 psig)
Base plate 316 stainless steel
Filter housing 316 stainless steel, 340 barg maximum, Viton gasket
Filter media borosilicate glass microfibre, 99.5+ percent removal of 0.1 µm particles
Metering valve needle valve type
Pressure gauge dual scale bar and psi
Sensor sample block 5/8 inch, 316 stainless steel
Flowmeter borosilicate glass, 2 to 12 Nl/min, Viton seals
Optimum sensor flow 1 to 5 litres per minute
Process tube 6 mm or 1/4 inch fittings in 316 stainless steel

Configuring ES20-M with order feature codes

After ES20-M, choose the filter (B1 particulate or B2 coalescing with drain, both removing 99.5 percent of 0.1 µm solids, with B2 also removing aerosols), the enclosure (C1 none and to be mounted inside a customer enclosure to comply with EN 61010-1:2010+A1:2019; C2 GRP with window IP66/NEMA 4X; C3 316 stainless with window IP66), the pressure mode (D1 process or atmospheric pressure with at least 0.2 barg inlet; D2 or D3 sub-atmospheric with a 230 V or 110 V AC pump, only with B1), the plate (E1 stainless plate or E2 with a monitor bracket), instrument options (F1 transmitter and monitor, F2 transmitter only) and material certification (M1 none or M2 to BS EN 10204 type 3.1). The datasheet’s example is ES20-M + B1 + C2 + D1 + E2 + F1 + M1.

How ES20-M differs from ES20-I and from ES70 systems

ES20-I is identical in function but supplied with 1/4 inch tube and fittings; ES20-M is for 6 mm tube. ES20-M accepts transmitters on a 5/8 inch UNF-style sensor block, whereas a PRO XP configuration of the ES20 uses a 3/4 inch block. The ES20 handles up to 20 barg, while the ES70 series is rated to 138 barg for natural gas and refinery duties, so ES20-M is the economical choice for compressed air and similar utilities.

Compatible instruments for ES20-M

The datasheet lists Easidew transmitter, Easidew I.S. transmitter, Easidew PRO I.S., Easidew PRO XP, Easidew Online hygrometer and Cermet II hygrometer. A sample from a pressurised line can be measured at pressure or let down to atmospheric, and the vacuum-pump option allows sub-atmospheric sampling.

Installation and use

Mount the panel or enclosure on a wall near the sample point, pipe the 6 mm inlet from a tapping fitted with an isolation valve, vent the outlet to a safe place, and set the needle valve so the flowmeter shows a steady flow within range. Without an enclosure (C1), the panel must be built into a protective cabinet. Each ES20 is supplied with a CD-ROM containing manual, datasheet, flow diagrams, EC declaration of conformity and optional material certificates.

Maintenance

Replace the cartridge when differential or contamination increases. The cartridges are recyclable and the housing is accessible for service. Keep tube runs short and metallic where possible, since water absorbs on plastic. If the panel is used outdoors in the GCC summer, shade the enclosure from direct sun, because a heated sample line or an overheated GRP window can change the sample temperature at the sensor block. For plants with several dew-point points, ES20-M panels can be repeated on each dryer, each with its own gauge and flowmeter, so that a technician reading one point does not disturb another.

Why conditioning matters for dew-point measurement with ES20-M

A transmitter can only report what reaches it, so the sampling system sets the quality of the reading. Pressure changes dew point directly, which is why ES20-M includes a gauge and a metering valve that can be arranged for atmospheric or line-pressure measurement. Flow matters because too little flow lengthens response, so the glass flowmeter shows 2 to 12 Nl/min while the sensor needs 1 to 5 l/min. Contamination, oil or liquid water would foul a sensor, so the filter housing in 316 stainless steel is rated 340 barg and uses recyclable cartridges. The base-plate layout keeps the gas path short, and materials in the wetted path were chosen by the manufacturer for optimum response to moisture changes. ES20-M is therefore a cost-effective way to turn a bare transmitter into a repeatable measuring point.

Documents and quotation

The ES20 datasheet and order-code sheet are provided as PDF references. For a quotation on ES20-M, send an RFQ with the part number, the feature string (filter, enclosure, pressure mode, plate, instrument option, certificate), quantity and the transmitter to be fitted.

Other Models in the Michell Sampling Systems Range

Seven Star LLC supplies models from the Michell Sampling Systems range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Michell ES20-M Compact Dew-Point Sampling System, MetricGas Analyzer AccessoriesYou are viewing this model
Michell ES70G-M Sampling System for Moisture AnalyzersGas Analyzer AccessoriesView details
Datasheets & Documents
  • PDF Datasheet Michell ES20-M Compact Dew-Point Sampling System, Metric Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Michell Instruments is a British manufacturer of humidity and dew point instruments, founded in 1974 in Cambridge, England, by Andrew Michell and incorporated on 12 September 1974. It worked with Cambridge University scientists on an aluminium oxide dew point sensor and later developed chilled mirror hygrometers and hydrocarbon dew point analysers. Michell is based in Ely, Cambridgeshire, and is part of Process Sensing Technologies (PST).

Seven Star LLC lists Michell dew-point transmitters, the Condumax II analyzer and its replacement sensor parts, and supplies them to gas plants and compressed air users in Oman by exact order code.

MICHELL Instruments company profile

Founded 1974, Cambridge, England, by Andrew Michell
Headquarters Ely, Cambridgeshire, United Kingdom, since 2007
Ownership Process Sensing Technologies (PST)
Technology Ceramic and aluminium oxide moisture sensors, chilled mirror dew point, Condumax hydrocarbon dew point (licensed from Shell)
Product families Easidew and Promet dew point transmitters, S8000 and Optidew chilled mirror hygrometers, Condumax II hydrocarbon dew point analysers, MDM300 portable dew point meters, oxygen analysers

Milestones

  • 1974: Andrew Michell founds Michell Instruments in Cambridge.
  • 2007: The company moves to Ely, Cambridgeshire.
  • 2024: Michell marks 50 years, now as part of PST.

Michell product families on this page

The Condumax II is an automatic hydrocarbon dew-point analyzer for natural gas. Several of its parts are listed separately: the CD2-SCR-002 hydrocarbon dew-point sensor cell, the CD2-WDC-002 water dew-point sensor assembly and the CD2-SX-WDC-002, which is the exchange version of that water dew-point sensor. The Easidew line is represented by the Easidew EA2 dew point transmitter and by the Easidew PRO I.S., an intrinsically safe moisture transmitter for hazardous areas. A ceramic impedance moisture sensor module is also listed, the sensing element behind these transmitters.

Where Michell analyzers are used

Hydrocarbon and water dew point are contract quality parameters for natural gas, so the Condumax II is found at gas processing plants, pipeline entry points and custody transfer stations. Easidew transmitters monitor dryers for instrument air and industrial gases, glove boxes, and dry gas feeds to processes that cannot tolerate water.

MICHELL Instruments products we list

Family Examples
Gas Analyzer Systems Easidew PRO I.S. Intrinsically Safe Mois…, Michell CD2-WDC-002 Water Dew-Point Sens…, Michell Condumax II Hydrocarbon Dew-Poin…, Easidew EA2 Dew Point Transmitter, Michell OptiPEAK TDL600 Moisture in Natu…, Michell Promet EExd Process Moisture Ana…, Michell S8000 Precision Chilled Mirror H…
Gas Analyzer Accessories MICHELL CD2-SX-WDC-002 Water Dew-Point S…, Michell Ceramic Impedance Moisture Senso…, Michell Condumax II CD2-SCR-002 Hydrocar…

Ordering Michell parts: what we need

For Condumax II parts send the analyzer serial number and the part code, such as CD2-SCR-002, and state if a new assembly or the exchange item is wanted. For Easidew transmitters we need the measuring span in dew point or ppm, the process connection thread, the output signal, the operating pressure, the gas and the area certification. A photo of the label on the old transmitter usually answers most of these in one go.

Facts checked against UK Companies House records and PST's Michell 50-year material, October 2026.

FAQ

Questions about the Michell ES20-M Compact Dew-Point Sampling System,

Answers from our engineering team on installation, compatibility and integration.

01What is ES20-M?
ES20-M is the metric (6 mm) version of the ES20 compact sampling system for dew-point transmitters, rated to 20 barg, built on a 316 stainless steel base plate.
02What filter options does ES20-M offer?
Particulate (B1) or coalescing with adjustable drain (B2) cartridges, both borosilicate microfibre removing 99.5 percent of 0.1 µm particles, with B2 also handling aerosols.
03Which enclosures are available with ES20-M?
Options are none (C1, must be built into a customer enclosure), a windowed GRP enclosure IP66/NEMA 4X (C2) or a windowed 316 stainless enclosure IP66 (C3).
04Can ES20-M sample below atmospheric pressure?
Yes, with the D2 or D3 vacuum pump option (230 V or 110 V AC), which is available only with the particulate filter option B1.
05Which instruments mount on ES20-M?