MICHELL Instruments Liquidew I.S. Moisture in Liquid Analyzer

Description

Overview

Water dissolved in hydrocarbon liquids is hard to catch with a laboratory sample, because the sample changes while it travels. The MICHELL Instruments Liquidew I.S. measures it continuously on line instead. It is a moisture-in-liquid analyzer for non-polar liquids, including flammable ones, reporting from 0.01 ppmW up to saturation through a rack-mounted control unit in a safe area and an intrinsically safe Easidew PRO I.S. sensor in the field. Seven Star LLC is an independent supplier serving Oman and the GCC.

System layout of the MICHELL Instruments Liquidew I.S.

The sensor and optional sampling system sit in the hazardous zone, certified ATEX and UKCA II 1 G Ex ia IIC T4 Ga, IECEx Ex ia IIC T4 Ga, TR CU 0Ex ia IIC T4 Ga and cQPSus Class I Division 1. A standard instrument twisted-pair cable returns the signal to a 19 inch sub-rack control unit, which also houses galvanic-isolation I.S. barriers. This arrangement allows an existing cable run to be reused in a retrofit.

The control unit can hold up to four fully independent channels in a multi-channel unit. Each channel has a 2.8 inch colour touch screen, and channels can mix Liquidew I.S. moisture-in-liquid, Promet I.S. moisture-in-gas and the Minox-i oxygen transmitter. Maintenance on one channel does not disturb the others, and blank channels can be ordered for later expansion.

MICHELL Instruments Liquidew I.S. specifications

Sensor technology Michell ceramic metal-oxide moisture sensor, Easidew PRO I.S. version
Measurement range 0.001 to 1000 ppmW; higher on request, depends on fluid solubility
Calibration range -100 to +20 C dew point
Dew point accuracy Plus or minus 1 C from -59.9 to +20 C; plus or minus 2 C from -60 to -100 C
Display resolution 0.1 C dp, 0.1 F dp, 0.001 ppmW, 0.1 C temperature
Temperature measurement Pt100, -20 to +70 C, accuracy plus or minus 0.2 C
Analysis pressure / temperature Up to 5 MPa (50 barg); 0 to +50 C
Sample flow 0.01 to 10 l/min; 0.1 to 0.3 l/min recommended
Analog outputs Three 4-20 mA, 500 ohm maximum load, user definable
Digital output RS485 Modbus RTU
Alarms Four relays: two Form C and two Form A, 30 V DC, 5 A non-inductive
Power supply 85 to 265 V AC 50/60 Hz or 20 to 28 V DC, 30 VA maximum
Control unit 19 inch sub-rack, 132 x 483 x 375 mm (h x w x d)
Sampling system enclosure 304 stainless steel (316 optional), 800 x 600 x 300 mm, IP66
Optional Sample cooling, enclosure heater or thermostat, oxygen sensor channel

How the measurement works

A ceramic sensor with a hygroscopic layer under one micron thick responds to the partial pressure of water vapour dissolved in the liquid, which relates directly to dew point. Firmware then converts dew point to ppmW using Henry’s Law: C in ppmW equals CS multiplied by e divided by es, where CS is the saturation concentration of the specific fluid at measurement temperature, e is the actual vapour pressure and es is the saturation vapour pressure. Preset CS values cover common hydrocarbons, two user tables take other fluids, and a proportional mixing mode handles two-solute blends such as the propane and butane in LPG.

Applications in Oman and the GCC

The datasheet cites naphtha feed to an isomerization catalyst, hexane solvent in polyethylene processes, benzene in styrene production, LNG and LPG, diesel and aero fuels to avoid water phase separation, BTX, ethylene and propylene feed, and butadiene and styrene. Regional refineries, petrochemical complexes and fuel terminals can use it to confirm dryer performance and to detect water breakthrough before it harms catalysts or causes icing.

How Liquidew I.S. differs from other Michell moisture analyzers

  • Liquidew I.S. reads moisture in liquid as ppmW from 0.001 to 1000 ppmW; the sister Promet I.S. measures moisture in gas to 45 MPa.
  • Liquidew analysis pressure is up to 5 MPa (50 barg) and sample flow 0.01 to 10 l/min; Promet I.S. handles 1 to 5 Nl/min.
  • Liquidew shows ppmW, dew point and temperature; Promet I.S. adds ppmV, natural gas units and pressure.

Installation and maintenance

Plan the cable route from the field sensor to the rack before ordering: the interconnection is ordinary instrument twisted pair, so long retrofit runs are practical, and no special barrier boxes are needed beyond the integrated galvanic isolation.

Install the sensor in the sampling system with the recommended 0.1 to 0.3 l/min flow, locate the control unit indoors in a safe area at 0 to 50 C and below 90 percent rh, and allow 100 mm of rear clearance for cables. Calibration data live in the sensor memory, so the Calibration Exchange Service lets operators swap a certified sensor without programming. Premium sampling systems include single or dual stage particulate filtration.

Sampling matters as much as sensing in oils and solvents. The MICHELL Instruments Liquidew I.S. premium sampling system filters particulates in one or two stages, offers a panel-mounted version for indoor racks, field enclosures with heating options, and a water heat exchanger to cool hot process fluid before it reaches the sensor, so that readings remain stable through plant swings.

Related products and ordering notes

For a quotation on Liquidew I.S., state the number of channels, the liquid, sample temperature and pressure, and whether a field enclosure with heating or cooling is required.

Related items you may also need: Michell SF82-TX SF82 Industrial Dew-Point Transmitter | Michell EPR-IS Easidew PRO I.S. Intrinsically Safe Transmitter | Pura PUR-2W-M12 Ultra-Pure Moisture Transmitter, M12 | All MICHELL Instruments products.

When requesting a quotation for the MICHELL Instruments Liquidew I.S., mention whether you need a single unit or a quantity for several sites and whether alternative equivalent items would be acceptable. Clear instructions help the team respond with accurate information and prevent delays caused by missing quantities or unclear substitution rules.

Details needed to quote the MICHELL Instruments Liquidew I.S.

The MICHELL Instruments Liquidew I.S. is a model family that Michell builds to a configuration, so the orderable unit is fixed by the options below and not by the model name alone.

  • Measuring range and the gas or sample the MICHELL Instruments Liquidew I.S. will analyse.
  • Output signals and communication needed by your control system, for example 4-20 mA, relays or a digital protocol.
  • Supply voltage and the mounting style required on site.
  • Hazardous area classification of the installation location, if any.
  • Sample handling and process connection details, including pressure, temperature and any sampling system or accessories.

Send these details, or the full code from the nameplate of an existing unit, and the quotation will confirm the exact configuration before order.

Documents and quotation

The manufacturer datasheet for the Liquidew I.S. is attached to this page. To request availability, send an RFQ to Seven Star LLC with the part number Liquidew I.S., the quantity and your destination in Oman or the GCC.

Datasheets & Documents
  • PDF Datasheet MICHELL Instruments Liquidew I.S. Moisture in Liquid Analyzer Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Michell Instruments is a British manufacturer of humidity and dew point instruments, founded in 1974 in Cambridge, England, by Andrew Michell and incorporated on 12 September 1974. It worked with Cambridge University scientists on an aluminium oxide dew point sensor and later developed chilled mirror hygrometers and hydrocarbon dew point analysers. Michell is based in Ely, Cambridgeshire, and is part of Process Sensing Technologies (PST).

Seven Star LLC lists Michell dew-point transmitters, the Condumax II analyzer and its replacement sensor parts, and supplies them to gas plants and compressed air users in Oman by exact order code.

MICHELL Instruments company profile

Founded 1974, Cambridge, England, by Andrew Michell
Headquarters Ely, Cambridgeshire, United Kingdom, since 2007
Ownership Process Sensing Technologies (PST)
Technology Ceramic and aluminium oxide moisture sensors, chilled mirror dew point, Condumax hydrocarbon dew point (licensed from Shell)
Product families Easidew and Promet dew point transmitters, S8000 and Optidew chilled mirror hygrometers, Condumax II hydrocarbon dew point analysers, MDM300 portable dew point meters, oxygen analysers

Milestones

  • 1974: Andrew Michell founds Michell Instruments in Cambridge.
  • 2007: The company moves to Ely, Cambridgeshire.
  • 2024: Michell marks 50 years, now as part of PST.

Michell product families on this page

The Condumax II is an automatic hydrocarbon dew-point analyzer for natural gas. Several of its parts are listed separately: the CD2-SCR-002 hydrocarbon dew-point sensor cell, the CD2-WDC-002 water dew-point sensor assembly and the CD2-SX-WDC-002, which is the exchange version of that water dew-point sensor. The Easidew line is represented by the Easidew EA2 dew point transmitter and by the Easidew PRO I.S., an intrinsically safe moisture transmitter for hazardous areas. A ceramic impedance moisture sensor module is also listed, the sensing element behind these transmitters.

Where Michell analyzers are used

Hydrocarbon and water dew point are contract quality parameters for natural gas, so the Condumax II is found at gas processing plants, pipeline entry points and custody transfer stations. Easidew transmitters monitor dryers for instrument air and industrial gases, glove boxes, and dry gas feeds to processes that cannot tolerate water.

MICHELL Instruments products we list

Family Examples
Gas Analyzer Systems Easidew PRO I.S. Intrinsically Safe Mois…, Michell CD2-WDC-002 Water Dew-Point Sens…, Michell Condumax II Hydrocarbon Dew-Poin…, Easidew EA2 Dew Point Transmitter, Michell OptiPEAK TDL600 Moisture in Natu…, Michell Promet EExd Process Moisture Ana…, Michell S8000 Precision Chilled Mirror H…
Gas Analyzer Accessories MICHELL CD2-SX-WDC-002 Water Dew-Point S…, Michell Ceramic Impedance Moisture Senso…, Michell Condumax II CD2-SCR-002 Hydrocar…

Ordering Michell parts: what we need

For Condumax II parts send the analyzer serial number and the part code, such as CD2-SCR-002, and state if a new assembly or the exchange item is wanted. For Easidew transmitters we need the measuring span in dew point or ppm, the process connection thread, the output signal, the operating pressure, the gas and the area certification. A photo of the label on the old transmitter usually answers most of these in one go.

Facts checked against UK Companies House records and PST's Michell 50-year material, October 2026.

FAQ

Questions about the MICHELL Instruments Liquidew I.S. Moisture in

Answers from our engineering team on installation, compatibility and integration.

01What does the Liquidew I.S. measure?
It measures dissolved moisture in non-polar liquids such as hydrocarbons, reporting ppmW from 0.001 to 1000 ppmW, dew point and temperature on a touch screen.
02Is the Liquidew I.S. certified for hazardous areas?
The sensor and sampling system carry ATEX and UKCA II 1 G Ex ia IIC T4 Ga, IECEx Ex ia IIC T4 Ga, TR CU and cQPSus Class I Division 1 certification.
03How many channels can the Liquidew I.S. control unit hold?
The 19 inch sub-rack multi-channel unit holds up to four independent measurement channels, which can mix Liquidew, Promet I.S. and Minox-i oxygen channels.
04What outputs does the Liquidew I.S. provide?
It provides three user definable 4-20 mA outputs, RS485 Modbus RTU and four alarm relays, two Form C and two Form A rated 30 V DC and 5 A.
05How is the Liquidew I.S. sensor recalibrated?
A Calibration Exchange Service supplies replacement sensors with calibration stored in on-board memory, so no user programming is needed after fitting.