Brand: Swan Analytical
Swan AMI Soditrace Trace Sodium Analyzer for Condensate and High-Purity Water
Description
Sodium at parts-per-trillion level for ultra-pure water
The Swan AMI Soditrace measures sodium ions in the trace range in condensate and high-purity water. The manual states that sodium measurement is used for quality control in high-purity water applications and to monitor breakthrough of ion exchangers. The swan.ch page lists power, semiconductor and nuclear applications and a lowest detection limit of 0.001 ppb (1 ppt).
It is a complete system with transmitter, sensors, flow cell, pH-controlled reagent dosing and a device for automatic calibration and sensor regeneration, all on a stainless steel mounting plate. The measurement uses a glass ion-sensitive electrode, and sodium readings below 1 ppb need a special glass formulation for the sensing electrode.
Technical specifications
Values are from the Operator’s Manual A-96.250.551 (section 2.2) and the swan.ch product page.
| Parameter | Specification |
|---|---|
| Measuring range (product page) | 0.001 ppb to 10.000 ppb |
| Accuracy (product page) | +/-0.005 ppb or +/-10% of reading |
| Reproducibility (product page) | +/-0.001 ppb or +/-5% of reading |
| Response time (product page) | 120 s (90%) |
| Electrodes | Sodium glass electrode and calomel reference electrode |
| Sample flow | Minimum 100 ml/min |
| Sample conditions | 5 to 45 C (41 to 113 F), inlet 0.3 to 3 bar (4 to 43 PSI), pressure-free outlet |
| Sample quality | pH 7.0 or higher; ammonium below 10 ppm; acidity below 50 ppm (CaCO3); dissolved solids below 10 ppm; no oil or grease |
| Connections | Sample inlet Serto PVDF 6 mm; outlet G 1/2 in adapter for 20 x 15 mm flexible tube |
| Power supply | 100-240 VAC (+/-10%), 50/60 Hz, or 10-36 VDC; maximum 35 VA |
| Transmitter | Aluminum, IP 66 / NEMA 4X; -10 to +50 C; backlit LCD 75 x 45 mm |
| Panel weight | Stainless steel panel, 13.0 kg (28.7 lb) |
Reagent conditioning and automatic care
A measuring limit of 0.001 ppb needs the sample conditioned to at least pH 12 while keeping sample integrity, and the best results are obtained with diisopropylamine (DIPA). An air pump carries reagent vapor to an air lift pump, and a conductivity sensor measures the amount of reagent added. A PID controller switches a valve to hold the conductivity at 450 uS/cm for low sodium readings, and a fall below the set point raises an error when the reagent is exhausted.
The system regenerates the sodium electrode automatically at a programmed interval, or 12 hours before a calibration, by etching it with a regeneration solution. Calibration is a three-point procedure in which a measured volume of standard, about 500 times smaller than the sample volume, is added to a closed loop and the reading is allowed to stabilize.
What to confirm before ordering
The sample must meet the stated limits: pH of at least 7, ammonium below 10 ppm and clean, oil-free water. Plan for a DIPA reagent bottle, sodium standard solution, regeneration solution and reference electrolyte, which are consumables. The unit is for indoor mounting and the installation steps include a 24 hour run-in period.
Panel dimensions are given as 860 mm in the text but 850 mm in the drawing of the manual, so the drawing should be requested for mounting. State AC or DC supply and communication options in the enquiry.
How the Swan AMI Soditrace works
The AMI Soditrace measures sodium with a glass electrode that responds to sodium ions. A small stream of sample passes through a flow cell, where it meets the sodium electrode and a calomel reference electrode. The sample is first raised to a high pH by adding a reagent vapor, so that hydrogen ions do not disturb the reading. The voltage between the electrodes is converted to a sodium value. A conductivity sensor checks that the right amount of reagent has been added.
Where the Swan AMI Soditrace is used
- Power plant condensate
- Boiler feed water
- Semiconductor high-purity water
- Nuclear plant water
- Ion exchanger breakthrough monitoring
Selection checklist for the Swan AMI Soditrace
- Measuring range from 0.001 to 10 ppb
- Sample flow of at least 100 ml/min
- Sample temperature of 5 to 45 degrees C
- Inlet pressure of 0.3 to 3 bar
- Reagent supply for conditioning
- Power: AC or DC supply
- Panel weight of 13 kg
- Calibration and regeneration needs
Installation, commissioning and maintenance notes for the AMI Soditrace
Mount the panel on a wall in a clean spot. Connect the sample line to the inlet and the drain to the outlet, keeping the outlet free of pressure. Fill the reagent bottle and connect the power. Allow the electrodes to settle, then calibrate. Replace the electrodes and the reagent on the interval in the manual, and watch the alarms.
Requesting a quotation for the Swan AMI Soditrace
Please send the sample type, the range needed, the power supply and the quantity. Seven Star LLC serves industrial customers in Oman and the GCC and replies with a quotation based on the information provided.
AMI Soditrace accuracy
The product page gives accuracy of plus or minus 0.005 ppb or 10 percent of reading, and reproducibility of plus or minus 0.001 ppb or 5 percent.
AMI Soditrace response time
The response time is 120 s to 90 percent of the final value.
AMI Soditrace sample quality
The sample should have pH 7.0 or higher, with no oil or grease, and ammonium below 10 ppm.
AMI Soditrace power and transmitter
Power is 100 to 240 VAC or 10 to 36 VDC, up to 35 VA. The transmitter has an IP66 case and a backlit display.
Analyser records for the AMI Soditrace
Keep calibration results, electrode dates and reagent batches with the plant record.
Choosing the right AMI Soditrace setup
Choose the reagent route from the lowest sodium level you must measure.
Documents and quotation
The attached Swan Operator’s Manual A-96.250.551 (firmware V6.20 and higher) and the swan.ch product page are the sources for this page. Seven Star LLC supplies Swan products as an independent supplier in Muscat. Give the supply type, options and quantity when you enquire. Further analyzers are collected under liquid analytical systems.
Datasheets & Documents
- Manual Swan AMI Soditrace Operator's Manual A-96.250.551 (firmware V6.20 and higher, 2021) Download
Reviews (0)
How would you rate this product? Cancel reply
About brand
Swan Analytical is the brand of Swan Analytische Instrumente AG, a Swiss manufacturer in Hinwil that has built online water analyzers since 1991. Its AMI range monitors water quality continuously in power plant steam and water cycles, ultra-pure water, potable water, pools, pharmaceutical and semiconductor processes. Seven Star LLC is an independent supplier in Muscat and lists selected Swan AMI analyzers for power and water plants.
Swan AMI analyzer families
Swan builds panel-mounted AMI monitors for disinfectants (Codes-II, Trides), dissolved oxygen (Oxytrace), conductivity (Powercon), silica (Silitrace), sodium (Soditrace, Sodium A and P), hydrazine, turbidity (Turbiwell, Turbitrace) and total organic carbon (LineTOC). Some earlier AMI models are shown as discontinued on swan.ch, where a newer AMI-II range is listed separately.
Ordering Swan analyzers: what we need
Send the model, measuring range, AC or DC supply, number of sample streams and options. We confirm the model status and the configuration against the Swan operator manual before quoting. Reagents and sensor consumables are ordered separately.
Related products
Frequently Asked Questions
The swan.ch page lists 0.001 ppb to 10.000 ppb sodium, with a lowest detection limit of 1 ppt.
DIPA (diisopropylamine) is recommended to condition the sample to at least pH 12 through an air lift pump.
It has automatic three-point calibration and automatic electrode regeneration, with a programmed interval.
The manual specifies a sample pH of 7.0 or higher and ammonium below 10 ppm.
The product page gives 120 seconds for 90% of a change.