Brand:

Bürkert 559612 Type 8222 ELEMENT Conductivity Transmitter

Description

Overview

Bürkert 559612 is the Type 8222 ELEMENT standard conductivity meter built around a graphite-electrode cell with a constant of 1.0 per centimetre, a PVDF union nut and three configurable outputs. It targets liquids that carry a noticeable amount of dissolved salts, where an electronic transmitter must hand a clean 4 to 20 mA value and two switching outputs to a PLC without a separate converter box.

Seven Star LLC supplies Bürkert 559612 to plants across Oman, and the explanation below sticks to the figures printed in the maker’s data sheet so an engineer can decide quickly whether this exact article number matches the loop being replaced or extended.

What Bürkert 559612 actually measures

The sensing head combines a two-electrode conductivity cell with a Pt1000 temperature element that sits inside the PVDF probe holder. Raw cell current is converted in the housing electronics into a conductivity reading that can span 0.05 microsiemens per centimetre up to 10 millisiemens per centimetre, while the temperature element reports from minus 20 up to plus 100 degrees Celsius. The stated conductivity deviation is 3 percent of the measured value, the resolution is 1 nanosiemens per centimetre, and the temperature deviation is 1 degree Celsius. The analogue current output adds an uncertainty of 1 percent of the current range.

Because the electrode of the 1.0 cell constant is graphite rather than stainless steel, this article suits mid to high conductivity media such as industrial process water, cooling loops, rinse water and moderately concentrated wastewater. Users who need to watch ultrapure water should look at the 0.01 or 0.1 cell constant siblings instead, which use stainless steel 1.4571 electrodes.

Key features of Bürkert 559612

  • Two-wire loop-powered operation for the current output, supplied with 14 to 36 V DC from a filtered, regulated and limited-energy source
  • Two transistor outputs that can be set to PNP or NPN behaviour for limit switching or pulse duty, up to 700 mA in total and 0.5 A per transistor when both are wired
  • One 4 to 20 mA output, sourcing or sinking in the same mode as the transistors, with a 150 ms response time from 10 to 90 percent
  • Removable backlit 128 by 64 dot-matrix display and configuration module, ordered separately, needed for set-up but not for continuous operation
  • Selectable temperature compensation: none, linear, sodium chloride curve, ultrapure water curve, or a curve defined for your own process
  • User-adjustable factor that converts conductivity into total dissolved solids
  • Reverse-polarity and overvoltage protection on the supply, and short-circuit protection on the transistors

Bürkert 559612 technical specifications

Parameter Value
Article number 559612
Type / variant 8222 ELEMENT standard, 3 outputs (2-wire)
Cell constant C = 1.0 per cm, graphite electrodes
Probe holder PVDF with stainless steel 1.4571 (316Ti)
Union nut PVDF, G 1½ internal thread for S022 adapter
Measuring range 0.05 µS/cm to 10 mS/cm
Temperature range measured -20 to +100 °C, Pt1000
Conductivity deviation 3 % of measured value
Resolution 1 nS/cm
Operating voltage 14 to 36 V DC (35 V in humid or outdoor sites)
Outputs 2 transistors NPN/PNP plus 1 x 4 to 20 mA
Maximum loop impedance 1100 ohm at 36 V, 610 ohm at 24 V, 180 ohm at 14 V
Electrical connection 5-pin M12 male connector (mating female plug ordered separately)
Fluid pressure Maximum PN 16
Fluid temperature -20 to +100 °C, limited by the adapter: PP 0 to 80 °C, PVC 0 to 50 °C, metal -20 to 100 °C
Ambient temperature -10 to +60 °C, humidity up to 85 % without condensation
Housing Stainless steel 1.4404 (316L) with PPS, PC cover
Pipe size DN 32 to DN 110 with S022 fittings, smaller pipes with reducers
Approval of this code CE, PED article 4 paragraph 1; no UL marking on this article

How 559612 differs from other 8222 variants

The Type 8222 family shares one electronic design, so the choice of article number is mostly a choice of cell constant, nut material, output count and UL marking. Article 559612 is the C = 1.0, PVDF, no-UL, three-output version. Its closest relatives behave as follows:

  • 559610 uses the same C = 1.0 cell and three outputs but carries a PVC nut, which limits the fluid to 0 to 50 degrees Celsius.
  • 559622 repeats this PVDF and C = 1.0 combination with UL Recognized marking.
  • 559611 and 559613 keep C = 1.0 but swap to four outputs, adding a second 4 to 20 mA signal and a 12 V minimum supply, using a three-wire hookup.
  • 559614 to 559620 move to the 0.1 and 0.01 cell constants with stainless electrodes for cleaner water.

Typical applications in Oman and the GCC

Gulf utilities run reverse osmosis trains, cooling towers and boiler feed lines where conductivity is the first alarm for scaling or carry-over. A graphite 1.0 cell sits well in the brine reject line and in cooling tower blowdown, where the operator wants a loop-powered reading plus a high-limit transistor that trips a dosing or blowdown valve. Refinery wash-water circuits, desalination post-treatment, food and beverage rinse monitoring and CIP return lines are further places where a PVDF wetted part and a stainless housing tolerate cleaning chemicals and salt-laden air.

Selection and compatibility notes

Before ordering confirm four points: the expected conductivity band against the 1.0 cell, the fluid temperature against the adapter material you plan to use, the number of analogue channels the PLC card provides, and whether a UL marking is required by the project. The sensor must be mounted in a Bürkert S022 insertion adapter or a compatible fitting with a G 1½ external thread. The device can be combined with the eCONTROL universal controller Type 8611 or the multiCELL Type 8619 when a local display or control function is wanted. Order the M12 female cable plug and the display module as separate lines.

Installation and commissioning

Fit the probe into the S022 adapter so the electrodes sit fully in the flow, keep the orientation described in the manual to avoid trapped air, and tighten the union nut by hand. Wire the supply through a limited-energy source, then plug in the display module to enter the measuring range, the unit, the temperature compensation mode and the limit values. Simulation of process values lets you prove the 4 to 20 mA scaling and the transistor switching before the line is filled. Protect the unit from strong electromagnetic fields and from direct sun when it is installed outdoors.

Maintenance

Inspect the electrode surface at intervals matched to fouling in your fluid. Graphite cells tolerate gentle rinsing, and a calibration against a known standard restores accuracy after cleaning. Replace the housing seals when the cover is removed repeatedly in dusty or humid conditions, and keep the M12 connector coupling tight to preserve the ingress rating.

Related products and ordering notes

When you request a quotation for Bürkert 559612, state the quantity, whether the display module and M12 plug are needed, and the adapter material of the pipe fitting. We can then group the S022 adapter, connector and software interface into one reply.

Seven Star LLC is an independent supplier serving Oman and the GCC. Product names and numbers belong to their respective owner.

Related items you may also need: Bürkert Type 8681 Control Head for Hygienic Process Valves | Bürkert 221602 Type 6213 Servo-Assisted 2/2-Way Solenoid Valve | Bürkert 178275 Type 6027 Stainless 2/2-Way Plunger Valve | All Bürkert products.

Documents and quotation

The data sheet for Type 8222 covers the technical data, dimensions, pressure and temperature curves and the ordering chart behind Bürkert 559612. Send your RFQ with the part number 559612 and the quantity you need, and add the cell constant and nut material you expect so that the match can be double checked.

Datasheets & Documents
  • PDF Datasheet Bürkert 559612 Type 8222 ELEMENT Conductivity Transmitter Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Bürkert is a German family-owned maker of fluid control systems, headquartered in Ingelfingen, Baden-Württemberg. Christian Bürkert founded the company in 1946, and it is known for solenoid valves, process and control valves, flow and analytical sensors, and mass flow controllers. Its products serve water treatment, hygienic processing, gas handling and laboratory applications.

Seven Star LLC lists Bürkert valves and valve automation devices and supplies Bürkert solenoid valves and spare parts to plants in Oman as an independent supplier.

Bürkert company profile

Founded 1946, by Christian Bürkert
Headquarters Ingelfingen, Germany
Ownership Family-owned
Product families Solenoid valves, angle seat process valves, positioners, flow meters, mass flow controllers

Bürkert types listed on this page

Two types are listed. Type 6026 is a direct-acting 2/2-way plunger solenoid valve, with a stainless steel bellows that separates the medium from the coil, and as a direct-acting valve it switches without needing a pressure difference. Type 8681 is the Bürkert control head that mounts on top of hygienic process valves to pilot the actuator and report valve position to the control system. Control heads of this kind are common in food, beverage and pharmaceutical plants in the GCC.

Ordering Bürkert valves: what we need

The type number alone is not enough. Send the identification number from the label, or else the type with orifice, port connection, body and seal material, coil voltage and frequency, circuit function and pressure range. For a Type 8681 control head add the communication variant and the make and model of the process valve it sits on.

Facts checked against Bürkert's published company history, October 2026.

FAQ

Questions about the Bürkert 559612 Type 8222 ELEMENT Conductivity

Answers from our engineering team on installation, compatibility and integration.

01What is the cell constant of Bürkert 559612?
Article 559612 uses C = 1.0 per centimetre with graphite electrodes, so it is aimed at mid to high conductivity liquids rather than ultrapure water.
02Does Bürkert 559612 have a UL marking?
No. Within the Type 8222 ordering chart the UL Recognized version of this PVDF and C = 1.0 combination is a different article, 559622.
03How is Bürkert 559612 powered and connected?
It is a two-wire loop device that needs 14 to 36 V DC through a limited-energy source, with a 5-pin M12 male connector. The female cable plug is not supplied.
04Is a display included with 559612?
No. The removable display and configuration module is ordered separately and is required for set-up, though the meter operates without it afterwards.
05What fluid temperature can 559612 handle?
The PVDF nut allows minus 20 to plus 100 degrees Celsius, but the real limit is set by the adapter: PVC 0 to 50, PP 0 to 80, metal minus 20 to 100 degrees Celsius.