Michell EA2-OL overview: panel-mount dew-point hygrometer with display and alarms

Michell EA2-OL is the parent order code of the Michell Easidew Online, a 1/8 DIN panel-mount hygrometer with a digital display, analog and digital outputs and two alarm relays. The order-code sheet describes it as a two-wire system supplied with a sampling block and a 0.8 metre cable assembly.

The datasheet gives a measurement range of minus 100 to plus 20 degrees Celsius dew point, with minus 110 to plus 20 on request, an accuracy of plus or minus 2 degrees Celsius dew point, a 4 to 20 mA or 0 to 20 mA output, Modbus RTU over RS485 and a choice of AC or DC supply.

It is common in compressed air stations because the display and relays give a local indication and a trip signal without extra instrumentation.

The image shows the Easidew Online panel monitor with its transmitter from the manufacturer datasheet.

Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer. We list Michell EA2-OL so that maintenance, instrumentation and procurement teams can quote the exact code that appears in the manufacturer documentation.

Key features of EA2-OL

  • Panel-mount 1/8 DIN monitor, 96 by 48 by 85 mm, IP65 front panel.
  • Dew point or ppmV moisture content output, with dual programmable relay alarms.
  • 4-20 mA or 0-20 mA output and Modbus RTU over RS485.
  • Monitor supply of 85 to 264 V AC or 24 V DC; current consumption 60 mA maximum.
  • Sensor with 5/8 inch UNF thread, MiniDIN connector and cable lengths from 0.8 to 10 metres.
  • Traceable 13-point calibration certificate and optional oxygen cleaning.

EA2-OL technical specifications

The table below lists values published by the manufacturer for the product family of EA2-OL. Where a value depends on the ordered configuration, the table says so and the order-code sheet governs.

Parameter Value
Order code (parent) EA2-OL, Easidew Online hygrometer (2-wire, 1/8 DIN panel mount, 2 alarms, sampling block, 0.8 m cable)
Measurement range Minus 100 to plus 20 degC dp (minus 110 to plus 20 degC dp available)
Accuracy Plus or minus 2 degC dew point
Response time 5 minutes to T95 (dry to wet)
Outputs 4-20 mA or 0-20 mA; Modbus RTU RS485; dual relay alarms
Monitor supply AC 85 to 264 V; DC 24 V
Current consumption 60 mA maximum
Sensor operating temperature Minus 40 to plus 60 degC
Monitor operating temperature 0 to plus 50 degC
Operating pressure 52.5 MPa (525 barg) maximum
Sensor ingress protection IP66; monitor IP65 (NEMA 4X) front panel
Monitor dimensions 96 x 48 x 85 mm (w x h x d)
Sensor process connection 5/8 inch 18 UNF
Sensor weight 150 g

Typical applications of EA2-OL in Oman and the GCC

  • Membrane and adsorption dryers in compressed air systems.
  • Medical and surgical air and breathing air.
  • Additive manufacturing, plastic injection moulding and glove boxes.

In Oman and across the GCC, moisture measurement matters wherever gas has to stay dry: gas gathering and processing plants, LNG and gas-to-power facilities, refineries and petrochemical complexes, desalination and utility compressed-air systems, and industrial gas and hydrogen users. High ambient temperatures and humidity make dryer performance and sensor drift a real operating cost, so a transmitter such as EA2-OL is usually specified against a defined dew-point or ppmV target and a documented calibration.

Selection and compatibility notes

The EA2-OL order string has five letters: range, protection (HDPE or sintered stainless steel), monitor supply (85 to 265 V AC or 24 V DC), sensor cable length of 0.8, 2, 5 or 10 metres, and oxygen cleaning. A typical example given on the sheet is EA2-OL-AAAAA.

A spare monitor is listed separately as EA2-OL-MON.

Confirm four things before ordering EA2-OL: the dew-point or ppmV range you need, the process connection and sample handling arrangement, the hazardous-area certification that your site requires, and the electrical interface of your control system. Because Michell builds many variants from one parent code, quote the complete configured code from the order-code sheet where one applies, and tell us the operating pressure so that ppmV conversions are programmed correctly.

Installation and commissioning of EA2-OL

Mount the monitor in a panel cut-out, fit the sensor in the supplied sampling block and route the sensor cable away from power cables. The accessory list includes an RS485 serial comms cable EA2-OL-MON-29739.

Mount the sensor in a sampling arrangement that delivers a clean, representative gas stream at a controlled flow, keep sample lines short and use stainless steel or other low-adsorption tubing, and avoid dead legs, because trace moisture adsorbs on tubing walls and slows the response. Wire the 4-20 mA loop with screened cable, check loop supply and load against the datasheet limits, and allow the sensor to stabilise on dry gas before judging the first readings. Record the as-found reading and the calibration certificate number at commissioning.

Maintenance, calibration and service

Recalibration and service exchange are available for the sensor, as for other Easidew transmitters, with spares EA2-SC and EA2-SCG for the MiniDIN connector.

Ceramic moisture sensors are stable, but periodic verification keeps a dryer control loop honest. Plan recalibration or sensor exchange at the interval set by your quality system, inspect the sensor guard and seals at each visit, and replace guards that are clogged by particulates or contaminated by oil mist. If a reading drifts, check the sample system for leaks and condensate before suspecting the sensor.

Related products and ordering notes for EA2-OL

  • EA2-TX: stand-alone Easidew transmitter.
  • EA2-AOL: Easidew Advanced Online hygrometer.
  • EA2-TX-M12: Easidew with M12 connector.

When you request EA2-OL, tell us whether the item is for a new installation, a like-for-like replacement or a spares stock, because that changes which accessories and documents are worth adding to the same quotation. Our team can also check related codes from the same order-code sheet so that one RFQ covers the transmitter or analyzer, its consumables and its calibration needs.

Frequently asked questions about EA2-OL

  • What is EA2-OL? EA2-OL is the manufacturer parent code for the Easidew Online dew-point hygrometer with a 1/8 DIN panel display, two alarms and a sampling block.
  • What does EA2-OL suit? It suits local display and alarming of dew point in compressed air and gas systems, such as dryers, breathing air and glove boxes.
  • What outputs does EA2-OL provide? The monitor gives 4-20 mA or 0-20 mA analog output, Modbus RTU over RS485 and two programmable relay alarms.
  • What should I send with an RFQ for EA2-OL? Send the part number EA2-OL, the quantity, the full configuration or host instrument model and serial number, any certificate needs and the delivery location so we can prepare an accurate quotation. The manufacturer datasheet for the product family can accompany the quotation.

Documents and quotation

A manufacturer document is available for the product family of EA2-OL and we can send it with your quotation where the family datasheet applies. To request pricing and lead time, send an RFQ to Seven Star LLC with the part number EA2-OL, the quantity, the full configuration or host-instrument details, the delivery location and any certificate requirements such as material certificates or calibration documents. We will confirm availability and respond with a quotation for EA2-OL in writing.

Other Models in the Michell Easidew Range

Seven Star LLC supplies models from the Michell Easidew range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Michell EA2-TX Easidew 2-Wire Dew-Point Transmitter 5/8 UNFHumidity & Dew Point InstrumentsView details
Michell EA2-TX-M12 Easidew Dew-Point Transmitter M12 ConnectorHumidity & Dew Point InstrumentsView details
Michell EA34-TX Easidew 34 Dew-Point Transmitter 3/4 UNFHumidity & Dew Point InstrumentsView details
Michell EA2-OL Easidew Online Dew-Point HygrometerHumidity & Dew Point InstrumentsYou are viewing this model
Datasheets & Documents
  • PDF Datasheet Michell EA2-OL Easidew Online Dew-Point Hygrometer Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Michell Instruments is a British manufacturer of humidity and dew point instruments, founded in 1974 in Cambridge, England, by Andrew Michell and incorporated on 12 September 1974. It worked with Cambridge University scientists on an aluminium oxide dew point sensor and later developed chilled mirror hygrometers and hydrocarbon dew point analysers. Michell is based in Ely, Cambridgeshire, and is part of Process Sensing Technologies (PST).

Seven Star LLC lists Michell dew-point transmitters, the Condumax II analyzer and its replacement sensor parts, and supplies them to gas plants and compressed air users in Oman by exact order code.

MICHELL Instruments company profile

Founded 1974, Cambridge, England, by Andrew Michell
Headquarters Ely, Cambridgeshire, United Kingdom, since 2007
Ownership Process Sensing Technologies (PST)
Technology Ceramic and aluminium oxide moisture sensors, chilled mirror dew point, Condumax hydrocarbon dew point (licensed from Shell)
Product families Easidew and Promet dew point transmitters, S8000 and Optidew chilled mirror hygrometers, Condumax II hydrocarbon dew point analysers, MDM300 portable dew point meters, oxygen analysers

Milestones

  • 1974: Andrew Michell founds Michell Instruments in Cambridge.
  • 2007: The company moves to Ely, Cambridgeshire.
  • 2024: Michell marks 50 years, now as part of PST.

Michell product families on this page

The Condumax II is an automatic hydrocarbon dew-point analyzer for natural gas. Several of its parts are listed separately: the CD2-SCR-002 hydrocarbon dew-point sensor cell, the CD2-WDC-002 water dew-point sensor assembly and the CD2-SX-WDC-002, which is the exchange version of that water dew-point sensor. The Easidew line is represented by the Easidew EA2 dew point transmitter and by the Easidew PRO I.S., an intrinsically safe moisture transmitter for hazardous areas. A ceramic impedance moisture sensor module is also listed, the sensing element behind these transmitters.

Where Michell analyzers are used

Hydrocarbon and water dew point are contract quality parameters for natural gas, so the Condumax II is found at gas processing plants, pipeline entry points and custody transfer stations. Easidew transmitters monitor dryers for instrument air and industrial gases, glove boxes, and dry gas feeds to processes that cannot tolerate water.

MICHELL Instruments products we list

Family Examples
Gas Analyzer Systems Easidew PRO I.S. Intrinsically Safe Mois…, Michell CD2-WDC-002 Water Dew-Point Sens…, Michell Condumax II Hydrocarbon Dew-Poin…, Easidew EA2 Dew Point Transmitter, Michell OptiPEAK TDL600 Moisture in Natu…, Michell Promet EExd Process Moisture Ana…, Michell S8000 Precision Chilled Mirror H…
Gas Analyzer Accessories MICHELL CD2-SX-WDC-002 Water Dew-Point S…, Michell Ceramic Impedance Moisture Senso…, Michell Condumax II CD2-SCR-002 Hydrocar…

Ordering Michell parts: what we need

For Condumax II parts send the analyzer serial number and the part code, such as CD2-SCR-002, and state if a new assembly or the exchange item is wanted. For Easidew transmitters we need the measuring span in dew point or ppm, the process connection thread, the output signal, the operating pressure, the gas and the area certification. A photo of the label on the old transmitter usually answers most of these in one go.

Facts checked against UK Companies House records and PST's Michell 50-year material, October 2026.

FAQ

Questions about the Michell EA2-OL Easidew Online Dew-Point Hygrometer

Answers from our engineering team on installation, compatibility and integration.

01What is EA2-OL?
EA2-OL is the manufacturer parent code for the Easidew Online dew-point hygrometer with a 1/8 DIN panel display, two alarms and a sampling block.
02What does EA2-OL suit?
It suits local display and alarming of dew point in compressed air and gas systems, such as dryers, breathing air and glove boxes.
03What outputs does EA2-OL provide?
The monitor gives 4-20 mA or 0-20 mA analog output, Modbus RTU over RS485 and two programmable relay alarms.
04Can Seven Star LLC supply EA2-OL in Oman?
Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer. Send an RFQ with the part number, quantity and delivery location and we will confirm availability.
05What should I send with an RFQ for EA2-OL?