Brand:

ifm SM9000 Magnetic-Inductive Flow Meter, G 2 DN50, 5 to 300 l/min

Description

Half-range DN50 metering: the ifm SM9000

Sometimes a DN50 pipe carries less water than its size suggests, and a meter scaled for 600 l/min wastes resolution. The ifm SM9000 covers 5 to 300 l/min in the same DN50, G 2 housing, with measuring dynamics of 1:60. It uses the magnetic-inductive principle for conductive liquids with at least 20 microsiemens/cm, and offers switching, analogue, pulse, frequency and IO-Link signals.

What to note

  • Range 0.3 to 18 cubic metres per hour.
  • Weight 3176 g, 32 g lighter than the SM2000.
  • Six green LEDs identify the displayed unit: l/min, m3/h, l, m3, 10 to the power 3 or degC.
  • Response 0.35 s, repeatability plus/minus 0.2 percent of end value.
  • Temperature accuracy plus/minus 1 K at 25 degC and flows above 15 l/min.

Technical table: SM9000

Product type Magnetic-inductive flow meter
Measuring range 5 to 300 l/min (0.3 to 18 m3/h); dynamics 1:60
Process connection G 2 external thread, DN50, flat seal, connection to pipe by adapter
Media Conductive liquids, water, hydrous media; conductivity at least 20 microsiemens/cm
Medium temperature Minus 10 to 90 degC (a further entry of minus 10 to 70 degC appears in the datasheet)
Pressure rating 16 bar
Operating voltage 18 to 32 V DC (SELV/PELV)
Outputs Switching, analogue 4 to 20 mA, pulse, frequency, IO-Link
Repeatability plus/minus 0.2 percent of measuring range end value
Temperature accuracy plus/minus 1 K (25 degC, flow above 15 l/min)
Response time 0.35 s
Dimensions 200 x 103 x 117 mm
Materials Stainless steel 316L and 316Ti, PEI, FKM, PBT-GF20, TPE-U
Weight / protection 3176 g, IP 65 and IP 67
Display 6 green LEDs (l/min, m3/h, l, m3, 10^3, degC)
MTTF 85 years

Read as running text, the manufacturer datasheet of the SM9000 states the following. Product type: Magnetic-inductive flow meter. Measuring range: 5 to 300 l/min (0.3 to 18 m3/h); dynamics 1:60. Process connection: G 2 external thread, DN50, flat seal, connection to pipe by adapter. Media: Conductive liquids, water, hydrous media; conductivity at least 20 microsiemens/cm. Medium temperature: Minus 10 to 90 degC (a further entry of minus 10 to 70 degC appears in the datasheet). Pressure rating: 16 bar. Operating voltage: 18 to 32 V DC (SELV/PELV). Outputs: Switching, analogue 4 to 20 mA, pulse, frequency, IO-Link. Repeatability: plus/minus 0.2 percent of measuring range end value. Temperature accuracy: plus/minus 1 K (25 degC, flow above 15 l/min). Response time: 0.35 s. Dimensions: 200 x 103 x 117 mm. Materials: Stainless steel 316L and 316Ti, PEI, FKM, PBT-GF20, TPE-U. Weight / protection: 3176 g, IP 65 and IP 67. Display: 6 green LEDs (l/min, m3/h, l, m3, 10^3, degC). MTTF: 85 years.

The datasheet explains the effect of measuring dynamics on usable range: a ratio of 1:60 means the highest flow in range is sixty times the lowest, here 300 l/min against 5 l/min. A lower ratio than the 1:120 of the SM2000 does not mean lower accuracy within the range, since repeatability is stated at plus/minus 0.2 percent of the end value for both. The temperature drift of plus/minus 0.0333 degC per K and temperature accuracy of plus/minus 1 K at 25 degC with flow above 15 l/min are identical to the larger device. A pressure loss diagram in the datasheet helps verify that a DN50 meter does not choke a low-flow branch.

Connection to a controller is flexible: the frequency output suits older counters, the analogue output suits loop indicators and IO-Link gives the controller access to diagnostics. Run shielded cable and keep it away from variable speed drives.

Typical installations for SM9000

Boiler feed lines, closed cooling loops, water dosing to mixing units, washdown supply and wastewater return lines are typical. In Oman, plants fed by treated municipal water or groundwater with conductivity in the usual range meet the 20 microsiemens/cm minimum, while demineralised water does not. A SM9000 on a medium-size branch can feed a SCADA system through IO-Link or the 4 to 20 mA output.

How SM9000 differs from the SM2000

Hydraulically the two are twins: same thread, same body size, same 16 bar rating. The difference is range and dynamics. The SM9000 runs from 5 to 300 l/min with 1:60 dynamics, the SM2000 from 5 to 600 l/min with 1:120. Weight is 3176 g versus 3208 g. If your normal flow is below about 300 l/min, the SM9000 spreads the output signal over a range that matches your flow.

Installation

Orient so the pipe stays full, keep the electrical connector accessible and observe the arrow. Use the flat seals with adapters, support the meter weight and avoid placing pumps or valves directly upstream.

Maintenance

Keep the process in the conductivity limits, inspect for deposit, test the analogue loop and note the second medium temperature entry in the datasheet before use with hot water.

Notes for buyers

Reference SM9000 in your RFQ. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.

Frequently asked questions about the SM9000

  • What flow range does the ifm SM9000 cover? It covers 5 to 300 l/min, which is 0.3 to 18 cubic metres per hour, with measuring dynamics of 1:60.
  • Which liquids suit the SM9000? Conductive liquids such as water and hydrous media with a conductivity of at least 20 microsiemens/cm.
  • What is the pipe connection of the SM9000? A G 2 external thread, DN50, with a flat seal and adapter.
  • How heavy is the SM9000? The datasheet gives 3176 g and dimensions of 200 x 103 x 117 mm, with IP 65 and IP 67.
  • How do I request a quotation for the SM9000? Send the RFQ with part number SM9000, quantity and destination to Seven Star LLC.

Documents and quotation

Use the attached datasheet as the reference for the SM9000. When you are ready, submit the RFQ giving part number SM9000, quantity and delivery point; availability is confirmed before a quotation is returned.

Datasheets & Documents
  • PDF Datasheet ifm SM9000 Magnetic-Inductive Flow Meter, G 2 DN50, 5 to 300 l/min Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

ifm is a German family-owned maker of sensors and automation products, founded in Essen in October 1969 by Robert Buck and Gerd Marhofer. Its first products were inductive proximity sensors sold under the efector name. ifm is still headquartered in Essen, and the founders' sons, Martin Buck and Michael Marhofer, have led it since 2001.

Seven Star LLC lists ifm position, distance and pressure sensors and supplies them by exact article number to plants and machine builders in Oman and the wider Gulf region.

ifm company profile

Founded 1969, Essen, Germany
Founders Robert Buck and Gerd Marhofer
Headquarters Essen, Germany
Ownership Family-owned
Product families Inductive and capacitive proximity sensors, PN pressure sensors, O1D laser distance sensors, flow and level sensors, IO-Link masters

Milestones

  • 1969: ifm is founded in Essen.
  • 2001: The founders' sons take over management.

ifm sensors listed on this page

ifm identifies every device with a short article number, and that number is what we order against. Three are listed here:

  • O1D100: laser distance sensor for position and distance measurement
  • PN7093: pressure sensor with local display
  • NG5021: inductive proximity sensor with NAMUR output

Other ifm inductive sensors, pressure sensors and connection cables can be quoted on request when you send the article number.

ifm products we list

Family Examples
Inductive Proximity Sensors NG5021 Inductive NAMUR Proximity Sensor
Laser Distance, LiDAR & Radar Sensors O1D100 Laser Distance Sensor
Pressure Instrumentation PN7093 Pressure Sensor with Display

Ordering ifm sensors: what we need

Send the six-character article number printed on the housing (two letters and four digits, or similar). If the label is worn, a photo plus the thread size, sensing range, output type and connector style is enough for us to check the part against ifm documentation. For NAMUR sensors, tell us which isolating amplifier is in the loop. Buyers in the GCC who need a successor for a discontinued article can send the old number and we will check the replacement the manufacturer names.

Facts checked against ifm's own 50-year chronicle (ifm.com), October 2026.

FAQ

Questions about the ifm SM9000 Magnetic-Inductive Flow Meter, G 2 DN50,

Answers from our engineering team on installation, compatibility and integration.

01What flow range does the ifm SM9000 cover?
It covers 5 to 300 l/min, which is 0.3 to 18 cubic metres per hour, with measuring dynamics of 1:60.
02Which liquids suit the SM9000?
Conductive liquids such as water and hydrous media with a conductivity of at least 20 microsiemens/cm.
03What is the pipe connection of the SM9000?
A G 2 external thread, DN50, with a flat seal and adapter.
04How heavy is the SM9000?
The datasheet gives 3176 g and dimensions of 200 x 103 x 117 mm, with IP 65 and IP 67.
05How do I request a quotation for the SM9000?
Send the RFQ with part number SM9000, quantity and destination to Seven Star LLC.