Brand: Bronkhorst
Bronkhorst F-201AV EL-FLOW Select Gas Mass Flow Controller
Description
Overview
Bronkhorst F-201AV belongs to the EL-FLOW Select family of digital thermal mass flow controllers for gases, and it is the model to look at when flows of tens of litres per minute must be set and held accurately. The datasheet states a minimum full scale of 0.4 to 20 ln/min and a maximum full scale of 2 to 100 ln/min. Seven Star LLC is an independent supplier serving Oman and the GCC.
Key data of the Bronkhorst F-201AV
- Flow range: minimum 0.4 to 20 ln/min, maximum 2 to 100 ln/min.
- Accuracy: ±0.5 % of reading plus ±0.1 % of full scale.
- Repeatability: better than ±0.2 % of reading, or ±0.04 % of full scale if greater.
- Turndown: up to 1:187.5, or 1:50 in analog mode.
- Pressure rating: 64 barg; ingress protection IP40; weight 0.7 kg.
- Maximum Kv value: 6.6 x 10-2.
- Power: 15 to 24 Vdc, 3.5 W typical at 24 V, plus 0.9 W for a fieldbus.
What the numbers mean in practice
Take a plant engineer sizing a purge line. With a maximum full scale of 100 ln/min and a turndown of 1:187.5 the range of the digital instrument extends well below one percent of the top value, while the analog mode gives 1:50, so for a 100 ln/min unit a usable lower end of about 2 ln/min. The accuracy statement is split in two parts: the reading-proportional part (0.5 %) dominates at high flow, and the full-scale part (0.1 %) dominates near the bottom of the range.
The valve of the F-201AV uses the standard plunger construction of FFKM with a polyimide foil. Optional plunger materials are EPDM with PI foil, FDA and USP Class VI approved EPDM and FDA and USP Class VI approved FFKM. Body seals are FKM/Viton as standard, with EPDM, FFKM/Kalrez or FDA and USP Class VI compounds on request. Wetted parts are stainless steel 316L or comparable.
Control behaviour
Settling time in control is typically 1 to 2 seconds, with an option of 500 msec. Control stability is within ±0.1 % of full scale for 1 ln/min of nitrogen. Pressure sensitivity is 0.1 % of reading per bar for nitrogen and 0.01 % per bar for hydrogen. Temperature sensitivity is below 0.05 % FS per °C on zero and below 0.05 % Rd per °C on span, over an operating range of -10 to +70 °C. Leak integrity to the outside is tested below 2 x 10-9 mbar l/s helium.
Interfaces
RS232 is standard. Analog output is 0 to 5 (10) Vdc or 0 (4) to 20 mA, and the analog setpoint input uses the same ranges. Optional digital buses are PROFIBUS DP, CANopen, DeviceNet, PROFINET, EtherCAT, Modbus RTU, ASCII or TCP/IP, EtherNet/IP, POWERLINK and FLOW-BUS. The main power connector is a 9-pin D-sub (male); the function connector carries analog, RS232 and RS485 signals. Because the electronic platform matches the other EL-FLOW Select models, a plant that already uses a bus with the F-201CV can add a Bronkhorst F-201AV on the same network.
Where the F-201AV is used in the Gulf
Larger thermal controllers such as this are found in gas blending stations used to prepare calibration mixtures, in nitrogen blanketing test rigs, in catalyst test benches at petrochemical research labs and in desalination and water treatment pilot plants where air or oxygen has to be metered. Manufacturing users apply them in furnace atmosphere control and in coating or surface treatment lines. The multi-fluid memory, with up to eight stored calibration curves, helps when one bench runs several gases in turn.
How Bronkhorst F-201AV differs from other EL-FLOW Select variants
The F-201CV tops out at 25 ln/min and weighs 0.6 kg. The F-201AV extends the maximum to 100 ln/min and weighs 0.7 kg. Both are rated to 64 barg. The F-111B is a meter without a valve with a 100 barg rating. If your largest flow is above 25 ln/min and you need control, the F-201AV is the logical choice. If it is below that, the smaller model offers a finer lower range.
Installation
Install with the flow arrow in the flow direction. Use compression type or face-seal VCR/VCO couplings as ordered. The datasheet specifies a maximum error of 0.2 % at 1 bar for nitrogen when the device is turned 90 degrees from horizontal, so note the mounting position at order and keep it. Provide 30 minutes warm-up. Protect the IP40 housing from dust and splashing, and filter the gas ahead of the instrument.
Servicing
Check zero regularly, keep the gas clean and dry and avoid sudden pressure steps. When recalibration is due, send the instrument back for service; confirm the calibration gas and pressure when you order spare or replacement units so the new instrument arrives configured for the same process.
Ordering the Bronkhorst F-201AV
An accurate quotation for the Bronkhorst F-201AV needs a short list of process data: the gas, the required full scale between 2 and 100 ln/min, the upstream pressure and the pressure at the outlet. The difference between the two sets the pressure drop available over the valve, and with a maximum Kv value of 6.6 x 10-2 that drop decides whether the top flow can really be reached. Add the temperature of the gas, the tube size of the line and the interface the control system uses, and mention whether the optional 500 msec settling time matters for your process.
Practical scenarios
Consider a catalyst test bench that must switch between nitrogen, air and a reaction mix every shift. The eight stored calibration curves reduce this to a menu selection. Or consider a purge gas line feeding several analyzer cabinets in a petrochemical plant: a 100 ln/min controller keeps the purge rate constant while the supply pressure moves, because the thermal sensor measures mass flow and the valve closes the loop. A third case is a laboratory that blends two gases from two controllers; the matching electrical platform lets both devices talk to the same fieldbus master.
In every scenario, keep the installation clean. Particles from old cylinder regulators and rusty tubing are the most common reason for valve problems, and a small sintered filter before the controller is cheap insurance.
Related items you may also need: Magnetrol 030-9114-001 PC Board for Thermatel TG1 Sensor Housing | ABB Sensyflow FMT400-VTS/-VTCS | Sierra QuadraTherm 780i In-Line Thermal Mass Flow Meter | All Bronkhorst products.
Documents and quotation
The manufacturer datasheet for the Bronkhorst F-201AV is provided as a PDF with this listing. For a quotation, send the part number F-201AV, the gas, full-scale flow, inlet and outlet pressure, connection type, interface option and quantity. Seven Star LLC replies with availability and lead time for Oman and the GCC.
Datasheets & Documents
- Datasheet Bronkhorst F-201AV EL-FLOW Select Gas Mass Flow Controller Datasheet Download
Reviews (0)
How would you rate this product? Cancel reply
About brand
Bronkhorst is a Dutch maker of instruments for measuring and controlling small flows of gases and liquids, known as "the low-flow specialists". The company was founded in 1981 in Vorden, the Netherlands, and moved to Ruurlo in 1983, where its headquarters and campus are today. From thermal mass flow meters and controllers to Coriolis meters, its instruments are used in laboratories, analytical equipment, semiconductor and chemical process plants, and the company reports more than one million instruments in the field.
Bronkhorst company profile
| Founded | 1981, Vorden, the Netherlands |
|---|---|
| Headquarters | Ruurlo, the Netherlands |
| Size | More than 800 employees worldwide, about 600 of them at the headquarters; four production locations in Europe, Asia and America |
| Core technology | Thermal mass flow meters and controllers for gases, Coriolis mass flow meters for liquids and gases, ultrasonic flow meters, plus pressure and vapor flow instruments |
| Typical applications | Analytical instruments, gas blending, laboratory and pilot plants, chemical and energy transition processes, semiconductor and food and beverage production |
| Standards | ISO 9001 certified since 1992; CE, RoHS, REACH marking on the instruments listed here; optional ATEX Zone 2 on mini CORI-FLOW meters |
Milestones
- 1981: Bronkhorst is founded in Vorden, the Netherlands.
- 1987: The company is recognised with the King Willem I Plaque as the most innovative young company.
- 1992: ISO 9001 certification is achieved.
- 1997: M+W Instruments GmbH is acquired and becomes Bronkhorst Instruments in Germany.
- 2005: Bronkhorst USA becomes the first subsidiary outside Europe.
- 2019: An R&D office for MEMS sensors opens at the Technology Park in Enschede.
How a Bronkhorst thermal mass flow controller works
A thermal mass flow instrument sends a small part of the gas through a sensor tube that is heated. The flowing gas carries heat downstream, and the resulting temperature difference along the tube is proportional to the mass flow of the gas, so the reading does not depend on the line pressure or temperature the way a volumetric reading would. In a controller such as the EL-FLOW Select F-201CV, the measured value is compared with the setpoint and a proportional valve in the same block opens or closes until the two agree, typically within one to two seconds. Coriolis meters, such as the mini CORI-FLOW M12, M13 and M14, work differently: a vibrating tube is twisted by the mass moving through it, which allows direct mass and density measurement of liquids and gases. When selecting, match the principle to the fluid: thermal for clean, dry gases, Coriolis for liquids or mixed fluids.
Where Bronkhorst products are used in Oman and the GCC
Bronkhorst instruments suit laboratories and pilot plants at refineries, petrochemical sites, universities and water treatment utilities: calibration gas blending for gas detectors, carrier gas for analyzers, chemical dosing trials, fuel and reagent flow on test benches, and gas supply to process development rigs across the region.
Bronkhorst models we list
| Family | Examples |
|---|---|
| EL-FLOW Select thermal gas flow meters and controllers | F-201CV, F-111B, F-201AV |
| EL-FLOW Prestige high performance gas controllers | FG-201CV |
| mini CORI-FLOW low flow Coriolis meters | M12, M13, M14 |
Ordering Bronkhorst products: what we need
Seven Star LLC is an independent supplier serving Oman and the GCC. For a quotation send the exact model code, the fluid or gas, the full-scale flow, the working pressure and temperature, the process connection and the interface needed, such as analog, RS232 or a fieldbus. Instruments are calibrated to the process data, so complete details avoid a wrong build; for spares or replacements, add the serial number from the nameplate.
Facts checked against the Bronkhorst company website and published Bronkhorst datasheets, November 2026.
Browse Bronkhorst products by category: Thermal Mass Flow Meters & Controllers, Coriolis Mass Flowmeters.
FAQ
Questions about the Bronkhorst F-201AV EL-FLOW Select Gas Mass Flow
Answers from our engineering team on installation, compatibility and integration.
01What is the flow range of the Bronkhorst F-201AV?
The datasheet lists a minimum full scale of 0.4 to 20 ln/min and a maximum full scale of 2 to 100 ln/min, so one model covers a wide span of calibrations.
02How does the F-201AV differ from the F-201CV?
The F-201AV reaches 100 ln/min instead of 25 ln/min and weighs 0.7 kg instead of 0.6 kg. Both share the 64 barg rating, IP40 housing and 3.5 W typical power.
03What is the turndown ratio of the F-201AV?
Turndown is up to 1:187.5 in digital operation and 1:50 in analog mode, with accuracy of ±0.5 % of reading plus ±0.1 % of full scale.
04Which seals and wetted materials are used?
Wetted parts are stainless steel 316L or comparable. Seals are FKM/Viton as standard, with EPDM, FFKM/Kalrez and FDA or USP Class VI compounds as options.
05Does the F-201AV store several gas calibrations?
Yes, the datasheet states storage of a maximum of eight calibration curves, so a single instrument can be switched between calibrated gases.