Brand:

Additel ADT295 HART, Fieldbus and Profibus PA Communicator

Description

One handheld for HART, Fieldbus and Profibus PA devices

The Additel ADT295 is an Android-based field device communicator that configures, diagnoses and calibrates smart transmitters and valve positioners without carrying separate tools for each protocol. The base ADT295 handles HART, the ADT295-FF adds FOUNDATION Fieldbus and the ADT295-FF-PA also supports Profibus PA. Because it also measures and sources loop signals, one instrument can replace a communicator, a milliamp meter and a loop power supply on a typical commissioning round.

The 5 inch touchscreen is paired with eight function keys, and the datasheet states that it works with gloves, which matters on outdoor skids.

Highlights from the datasheet

  • More than 2,600 DD devices with free wireless DD updates.
  • Android operating system with 64 GB of storage.
  • Wi-Fi, Bluetooth, NFC and USB-C connectivity.
  • Support for one external ADT161 module.
  • ISO 17025 certificate with data and a 1-year warranty.

Technical specification

Parameter Specification
Models ADT295 (HART); ADT295-FF; ADT295-FF-PA
Processor and keys Android OS, 8 core; 1 power key and 8 function keys
Optional battery 10000 mAh high-capacity pack (standard pack is 6900 mAh)
Environment Rated accuracy 8 to 30 C; storage -20 to 70 C; charging 0 to 45 C; 0 to 95% RH non-condensing; up to 3,000 m
Port protection Typical 60 VAC withstand; automatic protection against accidental 220 V input
Compliance CE
Display 5 in, 720 x 1280 touchscreen plus 8 function keys, glove compatible
Operating system and storage Android; 64 GB
Voltage measure -1 to 1 V: 0.01% RD + 50 uV; 1 to 30 V: 0.01% RD + 1.5 mV
Current measure +/-30 mA: 0.01% RD + 1.5 uA
Current output 0 to 24 mA: 0.01% RD + 1.2 uA
Loop supply 24 V +/-5%, up to 30 mA
Battery 6900 mAh; about 20 h measuring, about 14 h with 24 V output; Type-C charging about 2.5 h
Enclosure and ruggedness IP54; 1 m drop; 5 g vibration; 8 g shock; 48 h salt spray
Operating temperature -20 to 50 C; instant operation, no warm-up
Connectivity Wi-Fi, Bluetooth, NFC, USB-C
Size and weight 192.3 x 96.5 x 55.3 mm; 0.8 kg
Expansion One external ADT161 module
Certificate and warranty ISO 17025 certificate with data; 1 year

Communication, pressure and file handling

The datasheet describes online and offline operating modes: the communicator can talk to an installed device during operation, and it also supports calibration, testing and troubleshooting in instrument shops. Standard functions include DD menu navigation, parameter configuration, test methods, data trending, customizable hotkeys, favorites, configuration storage and offline simulation. Connecting one Additel ADT161 digital pressure module lets you read reference pressure while still talking to the transmitter, which suits comparative calibration of pressure transmitters. The ADT161 family spans vacuum to 60,000 psi (4,200 bar) and differential ranges as low as 1 inH2O.

Onboard storage of 64 GB can hold manuals, wiring diagrams, work instructions and calibration records, and Bluetooth transfer shares files between team members without a network. Libraries update over Wi-Fi at no charge, and Additel states that product and application support is included with every ADT295.

Which model to order

Select ADT295 if every smart device on site is HART, which is the case for most refinery and utility 4-20 mA loops. Choose ADT295-FF when you also service FOUNDATION Fieldbus segments, and ADT295-FF-PA when Profibus PA instruments are present. Check the DD list for your transmitter and valve controller makes before ordering, because device support depends on the DD library.

Use in the field

Typical work includes loop checks during commissioning, trimming a pressure or temperature transmitter, reading diagnostic alerts and changing ranges. The 24 V supply powers a bench-top transmitter while the measure channels read the returned loop current. Treat the loop supply and outputs as electrical sources and observe local permit-to-work rules in hazardous areas, since the datasheet does not list a hazardous-area rating.

Display and ruggedness details

The screen is covered in tempered glass and the datasheet calls it sunlight-readable, with adjustable backlighting and selectable light and dark themes. The USB port is electrically isolated from the measurement terminals, so the unit can be connected to a computer during field testing. The ports are protected against accidental 220 V contact, a useful safeguard when a technician moves quickly between junction boxes.

Where it fits in GCC plants

Instrument technicians in oil and gas, petrochemical, water treatment and power plants spend much of their time with smart transmitters. A glove-friendly, IP54 communicator with a long battery suits hot, dusty shifts and long turnaround days.

Before ordering, list the transmitter and valve controller makes you service so that the DD library coverage can be checked, and state whether the optional 10000 mAh battery or an ADT161 pressure module is wanted. Ask for the model with the fieldbus protocols that match your plant.

Selection and installation notes

A communicator is used in the field, so check that the battery is charged, the protocol option you need is fitted, the correct device description files are loaded and the loop has the resistance the protocol requires. Connect across the loop as shown in the manual, take care in hazardous areas to use only approved instruments, and save the configuration of each device so that a replacement can be set up again quickly.

Information needed for a quotation

Send the full model code, the quantity, the delivery location in Oman or the GCC, the intended use and any accessories needed, so that the complete package can be quoted together.

Documents and supply information

The manufacturer data sheet is attached to this page and should be read together with the project specification and the site risk assessment before a purchase is confirmed. Ratings, standards and dimensions are quoted as printed on that sheet. Lead time, stock position and certificates available with the shipment are confirmed in the quotation, and we supply to Oman and the wider GCC region for industrial, utility and building services projects.

Included accessories

  • 9811-X adapter, 9705 battery and 9019 test cable kit.
  • 9031 hooks and 9052 USB cable.

Related instruments are under portable test and measurement. Send the protocols and device makes you need and we will quote; confirm the model code before ordering.

Datasheets & Documents
  • PDF Datasheet Additel 295 field device communicators datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Additel is an American manufacturer of pressure and temperature calibration equipment, headquartered in Brea, California, where it also runs an NVLAP-accredited calibration laboratory. Its range covers automated pressure calibrators and controllers, documenting process calibrators, digital pressure gauges, test pumps and dry well calibrators. Additel has offices in Utah, Beijing and Odense, Denmark, and its products are used in more than 100 countries.

Seven Star LLC is an independent supplier in Muscat. We list Additel instruments with their official datasheets for instrument and calibration teams in oil and gas, power, water and manufacturing in Oman and the GCC.

Additel company profile

Headquarters Brea, California, USA
Calibration laboratory NVLAP-accredited, Brea
Other offices Pleasant Grove (Utah), Beijing and Odense (Denmark)
Reach Products used in more than 100 countries
Product families ADT760A and ADT761A automatic handheld pressure calibrators, ADT762 hydraulic calibrator, ADT226 documenting process calibrator, ADT681A and ADT601Ex digital pressure gauges, ADT875 dry well calibrators, pressure test pumps

Additel product series

Pressure calibration equipment includes the ADT760A and 760S automatic handheld pressure calibrators, the ADT761A portable automatic calibrator, the ADT762 automated hydraulic calibrator and the ADT927 hydraulic test pump for 700 bar. Digital pressure gauges include the ADT681A wireless gauge with accuracy up to 0.01% of full scale and the ADT601Ex intrinsically safe gauge.

The ADT226 is a multifunction documenting process calibrator, and the ADT295 communicates with HART, Foundation Fieldbus and Profibus PA devices. Temperature calibration is covered by the ADT875 dry well calibrators and the ADT878 reference dry well calibrators.

Where Additel instruments are used

Field calibration of pressure and temperature transmitters, gauges, switches and RTDs, loop checks and documented calibration for quality audits, and calibration labs in plants and service companies.

Additel products we list

Family Examples
Portable Test & Measurement ADT226 Multifunction Documenting Process…, ADT760A / 760S Automatic Handheld Pressu…, ADT761A Portable Automatic Pressure Cali…, ADT762 Automated Hydraulic Pressure Cali…
Pressure Instrumentation ADT601Ex Intrinsically Safe Digital Pres…, ADT681A Wireless Digital Pressure Gauge
Temperature Instrumentation ADT875 Series Dry Well Temperature Calib…

Ordering Additel instruments: what we need

Send the model and pressure range, the medium (gas, oil or water), accuracy class and whether results must be documented. For dry wells give the temperature range, stability and insert size. We confirm the exact model code before ordering.

Facts checked against Additel's own catalogue and calibration services brochure, October 2026.

Frequently Asked Questions

Which protocols does the ADT295 support?
Three models: ADT295 for HART, ADT295-FF for HART plus FOUNDATION Fieldbus, and ADT295-FF-PA which adds Profibus PA.
How many devices are supported?
The datasheet states more than 2,600 DD devices, with free wireless DD updates.
How long does the battery last?
The 6900 mAh battery gives about 20 hours of measuring and about 14 hours with the 24 V output active, and charges in about 2.5 hours over Type-C.
Can it measure and source loop signals?
Yes. It measures voltage and current, sources 0 to 24 mA and supplies 24 V loop power at up to 30 mA.
How rugged is it?
The datasheet lists IP54, a 1 m drop, 5 g vibration, 8 g shock and a 48 h salt spray test, with operation from -20 to 50 C.