Hanna HI9829 Multiparameter Water Quality Meter

Description

A field multiparameter system for lakes, rivers, seawater and effluent

The HI9829 is a portable logging multiparameter system that monitors up to 14 water-quality parameters, seven measured and seven calculated. A microprocessor-based multisensor probe measures pH, ORP, turbidity, dissolved oxygen, conductivity, chloride, nitrate, ammonium and temperature, and the meter records them together. Hanna designed the system for field measurements of lakes, rivers and the sea, and the manual describes the sample environment as fresh, brackish or seawater.

The meter meets IP67 (30 minutes at 1 m depth) and the probe meets IP68 for continuous immersion. A model with a built-in 12-channel GPS receiver, the HI98290, is listed in the same manual and adds position tracking with a stated accuracy of 10 m.

Technical specifications

Parameter Specification
Meter models HI9829 (multiparameter meter); HI98290 (with GPS)
Probe models HI7609829 standard probe; HI7629829 logging probe; cable lengths by suffix (for example /10 = 10 m)
pH range / accuracy 0.00 to 14.00 pH / ±0.02 pH; ±600.0 mV
ORP range / accuracy ±2000.0 mV / ±1.0 mV
Dissolved oxygen 0.0 to 500.0 %; 0.00 to 50.00 mg/L; ±1.5 % of reading or ±1.0 %, whichever is greater, up to 300 %
Conductivity 0 to 200 mS/cm (absolute EC up to 400 mS/cm); ±1 % of reading or ±1 µS/cm
TDS / resistivity / salinity TDS 0 to 400,000 mg/L; resistivity 0 to 1.0000 MΩ·cm; salinity 0.00 to 70.00 PSU; seawater sigma 0.0 to 50.0
Turbidity 0.0 to 99.9 FNU and 100 to 1000 FNU; ±0.3 FNU or ±2 % of reading, whichever is greater
Ammonium-N ISE 0.02 to 200.0 mg/L, ±5 % of reading or 2 mg/L
Chloride ISE 0.6 to 200.0 mg/L, ±5 % of reading or 2 mg/L
Nitrate-N ISE 0.62 to 200.0 mg/L, ±5 % of reading or 2 mg/L
Atmospheric pressure 450.0 to 850.0 mmHg; ±3.0 mmHg within ±15 °C of calibration temperature
Temperature -5.00 to 55.00 °C; ±0.15 °C
Meter logging 44,000 records, logging interval 1 second to 3 hours; USB with HI929829 software
Probe logging (HI7629829) 35,000 measurements with all parameters logged; 140,000 with a single parameter
Probe protection / depth IP68; maximum depth 20 m; operating -5 to 55 °C
Meter power 4 x C-size alkaline or NiMH cells
Meter size / weight 221 x 115 x 55 mm / 750 g

Sensors, tracking and data handling

Every sensor is field-replaceable and carries a colour-coded cap, and the probe and sensors are recognised automatically. Calibration is automatic: five pH buffers, six conductivity standards, 0 and 100 % for oxygen, and 0, 20 and 200 FNU for turbidity, with the last five calibrations stored for GLP records. A measurement check helps the operator avoid erroneous readings, and auto-ranging applies to EC, ISE and turbidity.

The Fast Tracker Tag ID System uses iButton tags installed at sampling sites so that each reading is linked to a location, with up to 100 tags stored in the meter. Graphical display of logged data is available on the meter, and the USB interface moves records to a PC.

Specifying the system, and what to confirm

This is a configurable system. Meters, probes, sensors and accessories can be ordered in predefined configurations or individually, so the first decision is which sensors the monitoring programme needs, then the cable length, then standard or logging probe. For groundwater, tank or lake profiling where the probe should record autonomously, choose the logging probe; for spot checks, the standard probe is sufficient.

Confirm the pH/ORP, conductivity/turbidity, DO and ISE sensor codes with the supplier, the calibration standards required for turbidity and ISE work, and whether the GPS version is needed. Battery life depends on the configuration: Hanna quotes 280 hours with alkaline cells and no GPS for pH, ORP, DO and EC, falling to 190 hours when turbidity is also enabled.

How the Hanna HI9829 works

The multisensor probe goes into the water and each sensor reads its own parameter. The probe sends the readings to the meter. The meter shows them on the display and can log them with a time stamp. Calculated values such as salinity come from the measured values.

Where the HI9829 is used

  • Lakes and rivers
  • Seawater and brackish water
  • Effluent checks
  • Environmental surveys
  • Field sampling trips

Selection checklist for the HI9829

  • Parameters to measure
  • Meter with or without GPS
  • Probe type, standard or logging
  • Cable length
  • Ion selective sensors needed
  • Logging interval
  • Water depth
  • Calibration solutions

Installation, commissioning and maintenance notes for the HI9829

Fit the sensors into the probe and connect the probe to the meter. Calibrate the sensors with fresh solutions before the field trip. Rinse the probe after each use. Store the sensors as the manual describes. Replace the batteries when the display shows a low level.

Requesting a quotation for the Hanna HI9829

Please send the parameters, the cable length and the water type. Seven Star LLC serves industrial customers in Oman and the GCC and replies with a quotation based on the information provided.

HI9829 system type

The HI9829 is a portable logging multiparameter system.

HI9829 parameter count

It monitors up to 14 parameters, seven measured and seven calculated.

HI9829 measured parameters

It measures pH, ORP, turbidity, dissolved oxygen, conductivity and temperature.

HI9829 ion sensors

Chloride, nitrate and ammonium sensors are available.

HI9829 meter protection

The meter is rated IP67.

HI9829 probe protection

The probe is rated IP68 for continuous immersion.

HI9829 GPS model

The HI98290 adds a 12-channel GPS receiver.

HI9829 pH accuracy

The pH accuracy is plus or minus 0.02 pH.

HI9829 temperature range

The temperature range is minus 5 to plus 55 degrees C.

HI9829 turbidity range

The turbidity range is 0 to 1000 FNU.

HI9829 meter logging

The meter stores 44,000 records.

HI9829 logging interval

The logging interval runs from 1 second to 3 hours.

HI9829 probe depth

The maximum probe depth is 20 m.

HI9829 power

The meter runs on four C-size cells.

Water quality records for the HI9829

Record the calibration date, the site and the readings.

Choosing the right HI9829 probe

Choose the logging probe when the probe must record on its own.

Documents and quotation

The full instruction manual is attached. For a related multi-parameter option, see the YSI ProDSS meter, or browse Portable Test and Measurement. Use the RFQ form to list the meter, probe and sensor codes you need.

Datasheets & Documents
  • PDF Instruction Manual Hanna HI9829 / HI98290 Instruction Manual Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Hanna Instruments is a manufacturer of analytical instruments for pH, ORP, conductivity, dissolved oxygen, turbidity, chlorine and titration, founded in 1978 by Oscar and Anna Nardo. Its portable meters, benchtop meters, photometers and electrodes are designed to give laboratory-style results in plant and field conditions.

Water quality instruments

Waterproof portable meters such as the HI98190, HI98192, HI98193 and HI9829 cover pH, EC, dissolved oxygen and multiparameter logging, while photometers and turbidimeters address chlorine and turbidity checks in drinking water, wastewater, boilers, cooling circuits and process lines. Features such as CAL Check validation cuvettes, GLP calibration recall and on-board data logging are described in each Hanna manual.

Sourcing through Seven Star LLC

Seven Star LLC, an independent supplier in Muscat, lists Hanna Instruments meters with their manufacturer documents so that buyers in Oman and the GCC can compare ranges, calibration standards and kit contents before requesting a quotation.

Frequently Asked Questions

How many parameters does the HI9829 handle?
Up to 14 water-quality parameters, of which seven are measured and seven are calculated.
Which probe has autonomous logging?
The HI7629829 logging probe, which stores up to 35,000 measurements with all parameters logged.
What is the depth rating of the probe?
IP68 with a stated maximum depth of 20 m (66 ft).
Is a GPS version available?
Yes. The HI98290 adds a 12-channel GPS receiver with 10 m stated accuracy.
How big is the meter log?