The Michell EA-PDP-A is the Easidew PDP Dryer Portable fitted with the Easidew EA2-TX-M12 ceramic metal-oxide dew-point transmitter. It is a rugged IP65 case containing a stainless-steel sample block, an internal pressure sensor and a 3.5 inch touchscreen, built to prove compressed-gas quality at the outlet of a dryer before custody transfer. With a range of -100 to +20 °C dew point, EA-PDP-A is the variant that can demonstrate ISO 8573-1 moisture Class 1 through Class 4 compliance. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.

Dryer-outlet testing with EA-PDP-A

Most compliance checks are made on desiccant and refrigeration dryers, and gas treatment dryers are specified at pressure dew point. The instrument measures dew point, line pressure and temperature, and can display either pressure dew point or a calculated atmospheric dew point. A 2 GB SD card stores data as a TXT file that downloads through USB-C without special software, and the plot function draws dew point and pressure against time, handy for watching a desiccant bed dry down after start-up.

EA-PDP-A technical data

Part number EA-PDP-A
Dew-point sensor Easidew EA2-TX-M12, ceramic metal-oxide
Moisture range -100 to +20 °C (-148 to +68 °F) dew point
ISO 8573-1 moisture class Class 1 to Class 4
Accuracy ±2 °C (±3.6 °F) dew point over compensated temperature range
Response, dry to wet 5 minutes to T95 from -70 to +10 °C dew point
Calibration traceable 13-point certificate
Pressure internal sensor, up to 20 barg (290 psig)
Moisture scales dew point and ppmV
Battery rechargeable lithium-ion, up to 2 working days in power-save mode, USB-C charging
Inlet filter for sensor HDPE, under 10 µm
Sample block 304 stainless block, 316 stainless tube
Flow recommended 1 to 5 Nl/min, maximum 15 Nl/min
Case and weight IP65 closed, IP54 open; 3.6 kg; 275 x 125 x 250 mm closed

How EA-PDP-A differs from EA-PDP-B

Both versions share the case, display, 20 barg pressure sensor and datalogger. What differs is the sensor: EA-PDP-A uses the Easidew EA2-TX-M12 for a range down to -100 °C dew point, Class 1 capability and a 13-point calibration, while EA-PDP-B uses an SF82-TX-M12 polymer sensor limited to -60 to +60 °C, Classes 2 to 4 and a 9-point calibration. EA-PDP-A is slower on dry-to-wet response (5 minutes against 3 minutes over different ranges), but covers much lower moisture. For Class 1 work, the manufacturer’s pigtail accessory EA-PDP-PIG helps protect against moisture ingress at the outlet.

Package contents and spares for EA-PDP-A

The portable is supplied with a 2 m PTFE tube of 6 mm, ten inlet filter spares, a 1.5 m USB-C power and data cable, a screen stylus and a calibration certificate. Order codes include EA-PDP-CHA (a 3 A fast charger), EA-PDP-PIG (coiled stainless pigtail rated to 20 barg), EA2-HDPE (ten replacement HDPE guards), SSF-PF-10PK filter cartridges, PTFE-xx tube by the metre, adaptors PDP-2PK-FIT-A, -B and -C, and the sensor exchange code EA2-TX-M12-X-CAA. Recalibration is listed as EA2-TX-Recal.

Moisture technology behind EA-PDP-A

The EA2-TX-M12 inside EA-PDP-A is built on Michell’s ceramic metal-oxide sensor, the thin-film design that responds to the partial pressure of water vapour and is chosen for trace moisture. In this portable it is protected by an HDPE guard that stops particles under 10 µm and is fed from a sample block that removes 99.5 percent of particles of 0.1 µm and larger through a drop-in cartridge. Because the wetted path is all stainless steel (a 304 block with 316 tube), adsorption and desorption of moisture on the walls stays low, which matters when the target is a -70 °C Class 1 dew point. Moisture is displayed in dew point or ppmV, and the instrument computes atmospheric dew point from the measured pressure dew point and line pressure.

Power, display and data handling of EA-PDP-A

The lithium-ion battery conforms to Battery Regulation (EU) 2023/1542 and is charged through a USB-C connector with the 1.5 m cable provided, or faster with the optional 3 A charger which has four plug adaptors. A 3.5 inch LCD with a 640 x 480 pixel touchscreen is operated with the supplied stylus, even with gloves in cold plant rooms. The unit conforms to EN 61010-1:2010+A1:2019 and works from -20 to +50 °C ambient, with storage from -40 to +50 °C. Weight is 3.6 kg, so one technician can carry it through a site.

Service and calibration of EA-PDP-A

Two routes keep traceability: a sensor exchange (a reconditioned, calibrated sensor comes first and the old one is sent back) and a full recalibration of the portable. The service panel gives access to the dew-point sensor, SD card, sample-block filter and battery for fast field service.

Where EA-PDP-A is used in Oman and the GCC

Typical duties are refrigeration and desiccant dryers, nitrogen, oxygen and hydrogen generators, railway pneumatics, medical gas quality, inert gas blanketing and plastic-moulding dryers. Engineers on compressed-air networks in industrial estates and port facilities will recognise the benefit of a unit that survives drops and noisy electrical environments. The datasheet also lists hydrogen gas generators, benchtop gas generators and inert gas blanketing among the intended uses, and says the instrument is designed for high electrical noise, so it can be used close to compressors, variable-speed drives and switchgear without special precautions beyond normal good practice. A single technician can test an entire dryer bank in one visit and leave with a stored log of every outlet.

Documents and quotation

The manufacturer’s datasheet for Easidew PDP Dryer Portable is supplied in PDF with this page. To quote EA-PDP-A, send the part number, quantity, preferred outlet fittings (Legris, 6 mm Swagelok or 1/4 inch Swagelok) and any spares such as EA-PDP-CHA in your RFQ.

Other Models in the Michell Easidew Pdp Range

Seven Star LLC supplies models from the Michell Easidew Pdp range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Michell EA-PDP-B Easidew PDP Dryer Portable with SF82Humidity & Dew Point InstrumentsView details
Michell EA-PDP-A Easidew PDP Dryer Portable HygrometerHumidity & Dew Point InstrumentsYou are viewing this model
Datasheets & Documents
  • PDF Datasheet Michell EA-PDP-A Easidew PDP Dryer Portable Hygrometer Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Michell Instruments is a British manufacturer of humidity and dew point instruments, founded in 1974 in Cambridge, England, by Andrew Michell and incorporated on 12 September 1974. It worked with Cambridge University scientists on an aluminium oxide dew point sensor and later developed chilled mirror hygrometers and hydrocarbon dew point analysers. Michell is based in Ely, Cambridgeshire, and is part of Process Sensing Technologies (PST).

Seven Star LLC lists Michell dew-point transmitters, the Condumax II analyzer and its replacement sensor parts, and supplies them to gas plants and compressed air users in Oman by exact order code.

MICHELL Instruments company profile

Founded 1974, Cambridge, England, by Andrew Michell
Headquarters Ely, Cambridgeshire, United Kingdom, since 2007
Ownership Process Sensing Technologies (PST)
Technology Ceramic and aluminium oxide moisture sensors, chilled mirror dew point, Condumax hydrocarbon dew point (licensed from Shell)
Product families Easidew and Promet dew point transmitters, S8000 and Optidew chilled mirror hygrometers, Condumax II hydrocarbon dew point analysers, MDM300 portable dew point meters, oxygen analysers

Milestones

  • 1974: Andrew Michell founds Michell Instruments in Cambridge.
  • 2007: The company moves to Ely, Cambridgeshire.
  • 2024: Michell marks 50 years, now as part of PST.

Michell product families on this page

The Condumax II is an automatic hydrocarbon dew-point analyzer for natural gas. Several of its parts are listed separately: the CD2-SCR-002 hydrocarbon dew-point sensor cell, the CD2-WDC-002 water dew-point sensor assembly and the CD2-SX-WDC-002, which is the exchange version of that water dew-point sensor. The Easidew line is represented by the Easidew EA2 dew point transmitter and by the Easidew PRO I.S., an intrinsically safe moisture transmitter for hazardous areas. A ceramic impedance moisture sensor module is also listed, the sensing element behind these transmitters.

Where Michell analyzers are used

Hydrocarbon and water dew point are contract quality parameters for natural gas, so the Condumax II is found at gas processing plants, pipeline entry points and custody transfer stations. Easidew transmitters monitor dryers for instrument air and industrial gases, glove boxes, and dry gas feeds to processes that cannot tolerate water.

MICHELL Instruments products we list

Family Examples
Gas Analyzer Systems Easidew PRO I.S. Intrinsically Safe Mois…, Michell CD2-WDC-002 Water Dew-Point Sens…, Michell Condumax II Hydrocarbon Dew-Poin…, Easidew EA2 Dew Point Transmitter, Michell OptiPEAK TDL600 Moisture in Natu…, Michell Promet EExd Process Moisture Ana…, Michell S8000 Precision Chilled Mirror H…
Gas Analyzer Accessories MICHELL CD2-SX-WDC-002 Water Dew-Point S…, Michell Ceramic Impedance Moisture Senso…, Michell Condumax II CD2-SCR-002 Hydrocar…

Ordering Michell parts: what we need

For Condumax II parts send the analyzer serial number and the part code, such as CD2-SCR-002, and state if a new assembly or the exchange item is wanted. For Easidew transmitters we need the measuring span in dew point or ppm, the process connection thread, the output signal, the operating pressure, the gas and the area certification. A photo of the label on the old transmitter usually answers most of these in one go.

Facts checked against UK Companies House records and PST's Michell 50-year material, October 2026.

FAQ

Questions about the Michell EA-PDP-A Easidew PDP Dryer Portable

Answers from our engineering team on installation, compatibility and integration.

01Which sensor is fitted in EA-PDP-A?
EA-PDP-A is fitted with the Easidew EA2-TX-M12 ceramic metal-oxide dew-point transmitter with a -100 to +20 °C dew-point range.
02Can EA-PDP-A be used for ISO 8573-1 Class 1 dryers?
Yes. The datasheet states EA-PDP-A is compliant for ISO 8573-1 moisture Classes 1 to 4, while the polymer-sensor EA-PDP-B covers Classes 2 to 4.
03What pressure can EA-PDP-A measure?
It has an internal pressure sensor for line pressure up to 20 barg (290 psig), and the display can show pressure dew point or a calculated atmospheric dew point.
04How does EA-PDP-A store data?
It logs dew point, pressure, temperature, time and date to a 2 GB SD card as a TXT file that downloads by USB-C to a laptop or PC.
05How long does the EA-PDP-A battery last?
The rechargeable lithium-ion battery gives up to two working days in power-save mode and is charged through the USB-C connector.