Michell EA2-SAM Easidew Sampler for Dew-Point Sensors

Description

Michell EA2-SAM Easidew Sampler for pressure or atmospheric dew point

The Michell EA2-SAM is a low-cost, self-contained sampling system machined from a single stainless-steel block, designed to condition a gas sample for a dew-point transmitter in compressed air and industrial gas systems. It combines a connection port, a particulate filter, a flow-control valve and a 5/8 inch UNF sensor port in one body, supplied with a 6 mm PTFE vent tube and a multi-directional mounting bracket. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.

One block, three jobs

The monolithic design reduces joints, internal volume and wetted surface, which gives faster response and higher integrity than a sampler built from discrete parts. Gas enters through a quick-connect push-fit port that takes 6 mm O/D plastic tube (PTFE or FEP) or 1/4 inch Swagelok fittings, passes a filter, runs over the sensor, and leaves through the metering valve. The valve sets the flow at 1 to 5 l/min (2.12 to 10.59 scfh), the optimum for the sensor.

EA2-SAM specification table

Part number EA2-SAM
Gas flow rate 1 to 5 l/min (2.12 to 10.59 scfh)
Operating pressure 1 MPa (10 barg / 145 psig); high-pressure option available
Particulate filter 99.5 percent removal of 0.3 µm
Gas connections quick-connect for 6 mm O/D tube, or 1/4 inch Swagelok
Sample vent tube 0.5 m PTFE pig-tail supplied
Block and cap material 316 stainless steel
Flow control block aluminium
Couplings nickel-plated brass
Sensor port 5/8 inch UNF
Ingress protection IP66 (NEMA 4)
Operating temperature -40 to +60 °C
Storage temperature -40 to +70 °C, 0 to 95 percent RH non-condensing
Weight 1.1 kg; 1.3 kg with sensor fitted
Vacuum standard push fittings are not vacuum rated

Switching EA2-SAM between pressure and atmospheric measurement

At the factory the sampler is built for full line-pressure measurement, with the flow valve on the outlet. To measure at atmospheric pressure, swap the flow valve and the gas coupling so the valve sits at the inlet; it then lets the pressure down before the gas reaches the sensor. The outlet should always carry the 0.5 m PTFE pig-tail to prevent back-diffusion of ambient moisture, in either mode. Mounting is not position critical, and two M6 x 5 mm deep threaded holes at a 20 mm pitch let the block be fixed directly without the bracket.

How EA2-SAM differs from EA2-SAM-HP and other sampling systems

EA2-SAM is the 10 barg version with push fittings and a 0.3 µm cartridge. The order-code sheet lists EA2-SAM-HP as the 5/8 inch UNF high-pressure sampler kit, intended for atmospheric or pressure dew-point measurement up to 200 barg (2900 psig). For larger duties, the same manufacturer offers the ES20 and ES70 panel systems with gauges, flowmeters and enclosures, so EA2-SAM suits only the simple case where a single transmitter must be fed and no gauge or flowmeter is needed.

Typical applications for EA2-SAM

The datasheet names compressed air dryers, plastic moulding, ozone generators, medical gases, pneumatics, breathing air and welding gases. A maintenance team in an Omani manufacturing plant would use it where a permanent Easidew transmitter sits downstream of a dryer and the technician wants stable gas flow without building a bespoke tube assembly.

Materials and construction of EA2-SAM

The sample path is built from 316 stainless steel for the block and cap, so the wetted surface that touches the gas is the same corrosion-resistant grade used in most process analyzer systems. The flow-control block is aluminium, the couplings are nickel-plated brass, and the filter cartridge is sealed by an O-ring in the cap. The IP66 (NEMA 4) rating allows installation outdoors or in washdown areas, and the temperature limits of -40 to +60 °C operating and -40 to +70 °C storage cover the climate range met in Oman, from cold-store rooms to open compressor yards, although the sensor fitted in the port may impose its own narrower limit. The complete unit weighs about 1.1 kg without a sensor, a mass that a simple bracket and one pipe clamp can carry.

A frequent alternative to EA2-SAM is a home-built line. Many sites build a sample line from fittings, a filter housing and a needle valve. That approach works, but it adds dead volume and leak points, and each joint can hold moisture that biases a low dew-point reading for hours. EA2-SAM puts all the elements in one machined block with an internal volume kept small on purpose. The result is a measuring point that is quick to commission, repeatable between sites and easy to replace as a complete assembly when a service engineer needs to swap it.

Fitting a transmitter and commissioning EA2-SAM

Screw the transmitter into the 5/8 inch UNF port with its seal, connect the sample line, fit the pig-tail and open the valve slowly until flow is in range. Allow time for the dry-down before logging dew point, since a damp sample line influences the first minutes. The filter is reached via a cap with an O-ring seal; other cartridge ratings can be supplied to order.

Maintenance of EA2-SAM

Replace the filter when pressure drop or contamination increases, inspect the cap O-ring, and verify that no part of the line has been exposed to a vacuum, because the push fittings are not vacuum rated. A pack of cartridges is listed in the accessories as SSF-PF-10PK.

Documents and quotation

The Easidew Sampler datasheet is provided as a PDF. For a quotation on EA2-SAM, send an RFQ with the part number, quantity, expected line pressure and the transmitter that will be mounted in it.

Selection Notes for the MICHELL Instruments EA2-SAM

When comparing the MICHELL Instruments EA2-SAM with alternatives, start from the duty it must perform: the measured or controlled variable, the range and accuracy you need, the environment at the mounting point and the interface to your control system. Check the electrical and mechanical connections against the installed equipment, and confirm any hazardous-area or ingress-protection requirements for the location. If any point is unclear, send a photo of the nameplate and a short description of the application, and we will compare it with the MICHELL Instruments humidity & dew point instruments range and confirm the correct order code before you commit.

Datasheets & Documents
  • PDF Datasheet Michell EA2-SAM Easidew Sampler for Dew-Point Sensors Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Michell Instruments is a British manufacturer of humidity and dew point instruments, founded in 1974 in Cambridge, England, by Andrew Michell and incorporated on 12 September 1974. It worked with Cambridge University scientists on an aluminium oxide dew point sensor and later developed chilled mirror hygrometers and hydrocarbon dew point analysers. Michell is based in Ely, Cambridgeshire, and is part of Process Sensing Technologies (PST).

Seven Star LLC lists Michell dew-point transmitters, the Condumax II analyzer and its replacement sensor parts, and supplies them to gas plants and compressed air users in Oman by exact order code.

MICHELL Instruments company profile

Founded 1974, Cambridge, England, by Andrew Michell
Headquarters Ely, Cambridgeshire, United Kingdom, since 2007
Ownership Process Sensing Technologies (PST)
Technology Ceramic and aluminium oxide moisture sensors, chilled mirror dew point, Condumax hydrocarbon dew point (licensed from Shell)
Product families Easidew and Promet dew point transmitters, S8000 and Optidew chilled mirror hygrometers, Condumax II hydrocarbon dew point analysers, MDM300 portable dew point meters, oxygen analysers

Milestones

  • 1974: Andrew Michell founds Michell Instruments in Cambridge.
  • 2007: The company moves to Ely, Cambridgeshire.
  • 2024: Michell marks 50 years, now as part of PST.

Michell product families on this page

The Condumax II is an automatic hydrocarbon dew-point analyzer for natural gas. Several of its parts are listed separately: the CD2-SCR-002 hydrocarbon dew-point sensor cell, the CD2-WDC-002 water dew-point sensor assembly and the CD2-SX-WDC-002, which is the exchange version of that water dew-point sensor. The Easidew line is represented by the Easidew EA2 dew point transmitter and by the Easidew PRO I.S., an intrinsically safe moisture transmitter for hazardous areas. A ceramic impedance moisture sensor module is also listed, the sensing element behind these transmitters.

Where Michell analyzers are used

Hydrocarbon and water dew point are contract quality parameters for natural gas, so the Condumax II is found at gas processing plants, pipeline entry points and custody transfer stations. Easidew transmitters monitor dryers for instrument air and industrial gases, glove boxes, and dry gas feeds to processes that cannot tolerate water.

MICHELL Instruments products we list

Family Examples
Gas Analyzer Systems Easidew PRO I.S. Intrinsically Safe Mois…, Michell CD2-WDC-002 Water Dew-Point Sens…, Michell Condumax II Hydrocarbon Dew-Poin…, Easidew EA2 Dew Point Transmitter, Michell OptiPEAK TDL600 Moisture in Natu…, Michell Promet EExd Process Moisture Ana…, Michell S8000 Precision Chilled Mirror H…
Gas Analyzer Accessories MICHELL CD2-SX-WDC-002 Water Dew-Point S…, Michell Ceramic Impedance Moisture Senso…, Michell Condumax II CD2-SCR-002 Hydrocar…

Ordering Michell parts: what we need

For Condumax II parts send the analyzer serial number and the part code, such as CD2-SCR-002, and state if a new assembly or the exchange item is wanted. For Easidew transmitters we need the measuring span in dew point or ppm, the process connection thread, the output signal, the operating pressure, the gas and the area certification. A photo of the label on the old transmitter usually answers most of these in one go.

Facts checked against UK Companies House records and PST's Michell 50-year material, October 2026.

FAQ

Questions about the Michell EA2-SAM Easidew Sampler for Dew-Point

Answers from our engineering team on installation, compatibility and integration.

01What does EA2-SAM contain?
EA2-SAM is a stainless-steel block with a 0.3 µm filter cartridge, push-in inlet and outlet fittings, a metering valve, a 6 mm PTFE vent tube and a mounting bracket.
02What pressure rating does EA2-SAM have?
The standard EA2-SAM is rated to 1 MPa (10 barg, 145 psig). A high-pressure kit is listed separately as EA2-SAM-HP in the order-code sheet.
03Which sensor thread does EA2-SAM accept?
It has a 5/8 inch UNF sensor port to take Michell impedance sensors and transmitters, and it weighs 1.1 kg, or 1.3 kg with a sensor fitted.
04Can EA2-SAM be used under vacuum?
No, not with the standard fittings. The datasheet states that the standard push fittings are not vacuum rated, so use a different sampler for sub-atmospheric work.
05How is atmospheric measurement set up on EA2-SAM?
Swap the flow-control valve and the inlet gas coupling so the valve is at the inlet. The valve then reduces the gas to atmospheric pressure before it reaches the sensor.