Michell MCPM4-1 Multi-Channel Process Monitor

Description

Michell MCPM4-1 multi-channel process monitor for moisture and oxygen

The Michell MCPM4-1 is the mains-powered four-channel rack monitor of the Multi-Channel Process Monitor family. It is a 19 inch, 3RU unit for the safe area that collects data from intrinsically safe Michell sensors in the hazardous zone: Promet I.S. for moisture in gas, Liquidew I.S. for moisture in liquids and the Minox-i transmitter for oxygen. Integrated galvanic isolation-type I.S. barriers, a 7 inch colour touchscreen and per-channel outputs mean a single box replaces separate monitors for each measuring point. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.

What each MCPM4-1 channel provides

Each channel has independent measurement electronics, sensors and sampling block, so maintenance on one channel does not interrupt the others. Per channel there are three configurable 4 to 20 mA outputs (maximum load 500 Ω), two process alarm relays with user-programmable set points and one additional system alarm, with Form C contacts rated 30 V DC and 5 A non-inductive. Alarm indicators on the display follow NAMUR 102. Serial communication uses RS485 Modbus RTU, and an Ethernet port adds Modbus TCP/IP.

MCPM4-1 specification table

Part number MCPM4-1
Enclosure 19 inch rack, 3RU; 132 x 483 x 375 mm (h x w x d)
Power supply 85 to 264 V AC, 47/63 Hz, maximum 30 VA (MCPM4-2 uses 18 to 28 V DC)
Display 7 inch colour touchscreen LCD
Analog outputs three 4 to 20 mA per channel, user-configured for parameter, unit and range
Alarms two relays per channel plus a system alarm
Digital RS485 Modbus RTU, Ethernet Modbus TCP/IP
Data logging SD card up to 32 GB FAT16/FAT32
I.S. barriers galvanic isolation type, integrated
Environment indoor safe area, 0 to +50 °C, under 90 percent rh
Interconnection cable twisted pair, screened, single pair (two pairs with pressure sensor)
Moisture calculation ISO or IGT#8 for natural gas

How MCPM4-1 differs from MCPM4-2 and larger monitor variants

The order-code sheet defines two bases: MCPM4-1 for 85 to 265 V AC and MCPM4-2 for 18 to 28 V DC, both with the same barriers, display and outputs. The datasheet also describes six-channel capability, so a four-channel unit leaves blank channels for later expansion. Each channel is then configured: B1 is a Promet I.S. gas sensor, B2 to B5 add a pressure transmitter of 0 to 69 or 0 to 138 barg for dynamic pressure compensation with ATEX/IECEx or North American certification, B6 and B7 are Liquidew I.S. liquid measurement, and B8 is the oxygen monitor for a Minox-i with % or ppmV range sensors. The mains-powered AC base MCPM4-1 is the usual choice in plants with a safe-area control room and utility power.

Applications of MCPM4-1 in oil, gas and petrochemical plants

Natural gas production and processing, pipeline drying, offshore export gas, fiscal metering and custody transfer, gas storage, hydrogen production and storage, LNG terminals, ethylene and polyethylene refineries and gas generation industries are all listed uses. For Oman and GCC facilities this means one cabinet near the control room can serve glycol outlet dryness, sales-gas quality and a nitrogen blanket oxygen check at the same time.

Displays and data on MCPM4-1

The screen presents moisture content or dew point (user choice) and analysis pressure, with configurable tag numbers per channel. Moisture can be read as ppmV, as natural-gas ppmV, lb/MMSCF or mg/m3, or as dew point in °C or °F; pressure in psig or barg; oxygen in percent or ppmV. Pressure compensation can use a fixed programmed value or a dynamic input from the optional sensor. An integral database holds Henry’s Law constants for liquid measurements, and users can add their own saturation constants.

Installing and wiring MCPM4-1

A multi-channel monitor saves panel space and wiring when several analyzers are needed. Mount the unit in a rack with at least 100 mm clear at the rear for cables and vents. The rear panel separates hazardous-area and safe-area terminals and has a single RS485 port. Run screened twisted-pair cable from each sensor, two pairs when a pressure sensor is used, and keep the instrument in a safe indoor area between 0 and +50 °C.

Sampling and companion sensors for MCPM4-1

Sensors usually sit on an ES70-type sampling system. The Promet I.S. combines the ceramic metal-oxide moisture sensor with sample conditioning, and the Minox-i is a loop-powered, galvanic fuel-cell oxygen transmitter with a 0 to 25 percent standard range. See the ES70G-M listing for a panel that can mount these. Because the sensor channels are intrinsically safe, MCPM4-1 itself stays in the safe area and only the screened sensor cables cross into the hazardous zone.

Measurement technology behind MCPM4-1 channels

The monitor does not measure moisture itself; it powers and reads the intrinsically safe sensors in the field. Promet I.S. uses Michell’s ceramic metal-oxide moisture sensor on a sample-conditioning system, and Liquidew I.S. relies on the same sensing element with a Henry’s Law calculation: moisture concentration in ppmW is the saturation concentration of the liquid at measurement temperature multiplied by the ratio of actual to saturation water vapour pressure. Pre-programmed saturation values cover common hydrocarbon liquids, with user entries for other fluids. Minox-i uses a galvanic fuel cell for oxygen from 0 to 25 percent. MCPM4-1 therefore gives one consistent interface, calibration reminders and alarm philosophy across gas, liquid and oxygen channels, with a data log on SD card that records parameters and error codes.

Documents and quotation

The multi-channel process monitor datasheet and order-code sheet are available as PDF references. For a quotation on MCPM4-1, send an RFQ with the part number, the configuration of each channel (B to E codes), quantity and the hazardous-area standard required.

Datasheets & Documents
  • PDF Datasheet Michell MCPM4-1 Multi-Channel Process Monitor Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Michell Instruments is a British manufacturer of humidity and dew point instruments, founded in 1974 in Cambridge, England, by Andrew Michell and incorporated on 12 September 1974. It worked with Cambridge University scientists on an aluminium oxide dew point sensor and later developed chilled mirror hygrometers and hydrocarbon dew point analysers. Michell is based in Ely, Cambridgeshire, and is part of Process Sensing Technologies (PST).

Seven Star LLC lists Michell dew-point transmitters, the Condumax II analyzer and its replacement sensor parts, and supplies them to gas plants and compressed air users in Oman by exact order code.

MICHELL Instruments company profile

Founded 1974, Cambridge, England, by Andrew Michell
Headquarters Ely, Cambridgeshire, United Kingdom, since 2007
Ownership Process Sensing Technologies (PST)
Technology Ceramic and aluminium oxide moisture sensors, chilled mirror dew point, Condumax hydrocarbon dew point (licensed from Shell)
Product families Easidew and Promet dew point transmitters, S8000 and Optidew chilled mirror hygrometers, Condumax II hydrocarbon dew point analysers, MDM300 portable dew point meters, oxygen analysers

Milestones

  • 1974: Andrew Michell founds Michell Instruments in Cambridge.
  • 2007: The company moves to Ely, Cambridgeshire.
  • 2024: Michell marks 50 years, now as part of PST.

Michell product families on this page

The Condumax II is an automatic hydrocarbon dew-point analyzer for natural gas. Several of its parts are listed separately: the CD2-SCR-002 hydrocarbon dew-point sensor cell, the CD2-WDC-002 water dew-point sensor assembly and the CD2-SX-WDC-002, which is the exchange version of that water dew-point sensor. The Easidew line is represented by the Easidew EA2 dew point transmitter and by the Easidew PRO I.S., an intrinsically safe moisture transmitter for hazardous areas. A ceramic impedance moisture sensor module is also listed, the sensing element behind these transmitters.

Where Michell analyzers are used

Hydrocarbon and water dew point are contract quality parameters for natural gas, so the Condumax II is found at gas processing plants, pipeline entry points and custody transfer stations. Easidew transmitters monitor dryers for instrument air and industrial gases, glove boxes, and dry gas feeds to processes that cannot tolerate water.

MICHELL Instruments products we list

Family Examples
Gas Analyzer Systems Easidew PRO I.S. Intrinsically Safe Mois…, Michell CD2-WDC-002 Water Dew-Point Sens…, Michell Condumax II Hydrocarbon Dew-Poin…, Easidew EA2 Dew Point Transmitter, Michell OptiPEAK TDL600 Moisture in Natu…, Michell Promet EExd Process Moisture Ana…, Michell S8000 Precision Chilled Mirror H…
Gas Analyzer Accessories MICHELL CD2-SX-WDC-002 Water Dew-Point S…, Michell Ceramic Impedance Moisture Senso…, Michell Condumax II CD2-SCR-002 Hydrocar…

Ordering Michell parts: what we need

For Condumax II parts send the analyzer serial number and the part code, such as CD2-SCR-002, and state if a new assembly or the exchange item is wanted. For Easidew transmitters we need the measuring span in dew point or ppm, the process connection thread, the output signal, the operating pressure, the gas and the area certification. A photo of the label on the old transmitter usually answers most of these in one go.

Facts checked against UK Companies House records and PST's Michell 50-year material, October 2026.

FAQ

Questions about the Michell MCPM4-1 Multi-Channel Process Monitor

Answers from our engineering team on installation, compatibility and integration.

01What is MCPM4-1?
MCPM4-1 is the 85 to 265 V AC four-channel version of the Michell Multi-Channel Process Monitor, a 19 inch rack unit for Promet I.S., Liquidew I.S. and Minox-i sensors.
02What outputs does each MCPM4-1 channel have?
Each channel has three configurable 4 to 20 mA outputs, two programmable alarm relays and a share of the system alarm, plus RS485 Modbus RTU and Ethernet Modbus TCP/IP.
03Does MCPM4-1 include I.S. barriers?
Yes. Galvanic isolation-type intrinsically safe barriers are integrated into the monitor, with hazardous-area and safe-area connections segregated on the rear panel.
04How does MCPM4-1 differ from MCPM4-2?
Both share the same 7 inch display, barriers and outputs; MCPM4-1 runs on 85 to 265 V AC while MCPM4-2 runs on 18 to 28 V DC.
05Can MCPM4-1 monitor liquids and oxygen?
Yes. Channels can be set for moisture in gas (Promet I.S.), moisture in liquid (Liquidew I.S.) or oxygen (Minox-i), in any combination.