MICHELL Instruments Cermet II Dew-Point Hygrometer, -100 to +20 C

Description

Overview

The MICHELL Instruments Cermet II is an on-line hygrometer that pairs a smart ceramic moisture sensor with a separate panel-mount monitor, so dew point can be read at a convenient location up to 1200 m (3930 ft) from the sample point. It is calibrated from minus 100 to plus 20 degrees Celsius dew point and works at pressures from vacuum up to 40 MPa, making it a flexible choice for dried compressed air, welding and industrial gases, glove boxes, medical gases and sulfur hexafluoride switchgear. Seven Star LLC is an independent supplier serving Oman and the GCC.

Components of the MICHELL Instruments Cermet II system

Sensor

The MICHELL Instruments Cermet II sensor uses Michell’s Advanced Ceramic Moisture Sensor with an on-board processor, which supports interchangeability and stable performance. It is housed in stainless steel, rated NEMA 4 and IP66, and operates from minus 40 to plus 60 degrees Celsius, compensated over minus 20 to plus 40. Process connection is a 5/8 inch UNF parallel thread; the standard cable is 2 m long.

Monitor

The monitor is a self-contained display in a 1/8 DIN panel case with IP54 (NEMA 12) protection on the front panel, upgradable to IP66 (NEMA 4) with an optional cover. It carries two isolated 10 A form C alarm relays rated 240 V AC or 24 V DC, a 4-20 mA output as standard with a maximum load of 500 ohm, and an RS232 port. Its universal supply accepts 85 to 265 V AC at 50 or 60 Hz, or 95 to 370 V DC.

MICHELL Instruments Cermet II specifications

Measurement range -100 to +20 C (-148 to +68 F) dew point
Accuracy Plus or minus 1 C from -60 to +20 C dew point; plus or minus 2 C from -100 to -60 C
Resolution 0.1 C from -80 to +20 C; 1 C from -100 to -80 C
Operating pressure Vacuum to 40 MPa (400 barg, 5801 psig) maximum
Flow rate / velocity 1 to 5 l/min; 0 to 10 m/s
Measurement units C dp, F dp, ppmV, lb/MMscf (natural gas), g/m3 (natural gas)
Secondary variable Pressure from an optional external transducer, 0-45 MPa, 4-20 mA input
Sensor calibration Traceable to NIST and UKAS
Sensor enclosure Stainless steel, NEMA 4 and IP66
Analyzer operating temperature 0 to +50 C
Sensor operating temperature -40 to +60 C (storage -40 to +109 C)
Cable length 2 m standard, 1200 m maximum
Optional outputs 0-20 mA, 0-10 V, RS485; 1 or 2 extra 5 A relays (4 total)

Technology: how the ceramic sensor works

The ceramic tile is a sandwich: a base electrode on a ceramic substrate, a porous hygroscopic insulating layer less than one micron thick, and a permeable gold film electrode thinner than 0.1 micron. Water has a very high dielectric constant compared with the active layer and the carrier gas, so adsorbed molecules change the impedance, and the response is rapid both when the gas dries and when it gets wetter.

Where it is used in Oman and the GCC

Typical duties in the Gulf include checking the outlet of heat-regenerative and adsorption dryers that supply instrument air, monitoring pipeline drying by vacuum and dry gas purging, verifying industrial gas purity, and watching SF6 in high-voltage switchgear. Pressure compensation is useful when the plant reports moisture in ppmV or in natural gas units, because the dew point of a gas varies with pressure; a live pressure transducer or a manual pressure factor can supply the correction.

How Cermet II differs from other Michell dew point instruments

  • MICHELL Instruments Cermet II has a separate sensor and monitor, up to 1200 m apart; Easidew transmitters are single-body loop-powered devices.
  • Its range of minus 100 to plus 20 C dew point and 40 MPa pressure rating exceeds the portable MDM300, which is limited to 350 barg.
  • It provides two standard alarm relays, expandable to four, plus ppmV and natural gas unit conversion.

Choosing units, alarms and outputs

The display can show degrees Celsius or Fahrenheit dew point, ppmV with a user selectable range and resolution, pounds per million standard cubic feet for natural gas, or grams per cubic metre. Digital processing linearizes the sensor signal and applies active or passive pressure compensation. Alarm relays can be set over the full operating range with variable hysteresis in any chosen engineering unit, so a dryer-failure alarm can be written in the same units used in the plant specification. For a purchase description, write MICHELL Instruments Cermet II together with the cable length, relay count and output options, which lets the supplier confirm the exact build against the datasheet.

Installation and commissioning

Fit the sensor in a sample block or directly in the line using the 5/8 inch UNF thread, keep flow between 1 and 5 l/min where a sample cell is used, and run the cable to the panel monitor. Set the units, alarm points with their hysteresis, and the 4-20 mA scaling at the front panel. Alarms can also flag a sensor or cable fault. Enter a pressure factor or connect a transducer if concentration units are needed.

Maintenance

Sensor calibrations are traceable to NIST and UKAS standards, and the on-board intelligence means a replacement sensor can be swapped without reprogramming the monitor. Plan periodic recalibration according to the application, and keep the sensor protected from liquid water and contaminants such as oil mist.

Related products and ordering notes

When requesting a price for MICHELL Instruments Cermet II, give the dew point range, pressure, cable length, and whether extra relays, a 0-10 V output or RS485 are needed.

Related items you may also need: Michell MCPM4-1 Multi-Channel Process Monitor | Michell Condumax II CD2-SCR-002 Hydrocarbon Dew-Point Sensor Cell | Condumax II CD2-SYS-I Analyzer with Sampling System | All MICHELL Instruments products.

Points to confirm for the MICHELL Instruments Cermet II

  • If the MICHELL Instruments Cermet II is a like-for-like replacement, read the full code from the installed nameplate and compare it with this page.
  • For each MICHELL Instruments Cermet II request we confirm the exact code and lead time in writing.

Details needed to quote the MICHELL Instruments Cermet II

The MICHELL Instruments Cermet II is a model family that Michell builds to a configuration, so the orderable unit is fixed by the options below and not by the model name alone.

  • Measuring range and the gas or sample the MICHELL Instruments Cermet II will analyse.
  • Output signals and communication needed by your control system, for example 4-20 mA, relays or a digital protocol.
  • Supply voltage and the mounting style required on site.
  • Hazardous area classification of the installation location, if any.
  • Sample handling and process connection details, including pressure, temperature and any sampling system or accessories.

Send these details, or the full code from the nameplate of an existing unit, and the quotation will confirm the exact configuration before order.

Documents and quotation

The datasheet for the MICHELL Instruments Cermet II is attached on this page. To request availability, send an RFQ to Seven Star LLC with the part number MICHELL Instruments Cermet II, the quantity and your destination in Oman or the GCC.

Datasheets & Documents
  • PDF Datasheet MICHELL Instruments Cermet II Dew-Point Hygrometer, -100 to +20 C Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Michell Instruments is a British manufacturer of humidity and dew point instruments, founded in 1974 in Cambridge, England, by Andrew Michell and incorporated on 12 September 1974. It worked with Cambridge University scientists on an aluminium oxide dew point sensor and later developed chilled mirror hygrometers and hydrocarbon dew point analysers. Michell is based in Ely, Cambridgeshire, and is part of Process Sensing Technologies (PST).

Seven Star LLC lists Michell dew-point transmitters, the Condumax II analyzer and its replacement sensor parts, and supplies them to gas plants and compressed air users in Oman by exact order code.

MICHELL Instruments company profile

Founded 1974, Cambridge, England, by Andrew Michell
Headquarters Ely, Cambridgeshire, United Kingdom, since 2007
Ownership Process Sensing Technologies (PST)
Technology Ceramic and aluminium oxide moisture sensors, chilled mirror dew point, Condumax hydrocarbon dew point (licensed from Shell)
Product families Easidew and Promet dew point transmitters, S8000 and Optidew chilled mirror hygrometers, Condumax II hydrocarbon dew point analysers, MDM300 portable dew point meters, oxygen analysers

Milestones

  • 1974: Andrew Michell founds Michell Instruments in Cambridge.
  • 2007: The company moves to Ely, Cambridgeshire.
  • 2024: Michell marks 50 years, now as part of PST.

Michell product families on this page

The Condumax II is an automatic hydrocarbon dew-point analyzer for natural gas. Several of its parts are listed separately: the CD2-SCR-002 hydrocarbon dew-point sensor cell, the CD2-WDC-002 water dew-point sensor assembly and the CD2-SX-WDC-002, which is the exchange version of that water dew-point sensor. The Easidew line is represented by the Easidew EA2 dew point transmitter and by the Easidew PRO I.S., an intrinsically safe moisture transmitter for hazardous areas. A ceramic impedance moisture sensor module is also listed, the sensing element behind these transmitters.

Where Michell analyzers are used

Hydrocarbon and water dew point are contract quality parameters for natural gas, so the Condumax II is found at gas processing plants, pipeline entry points and custody transfer stations. Easidew transmitters monitor dryers for instrument air and industrial gases, glove boxes, and dry gas feeds to processes that cannot tolerate water.

MICHELL Instruments products we list

Family Examples
Gas Analyzer Systems Easidew PRO I.S. Intrinsically Safe Mois…, Michell CD2-WDC-002 Water Dew-Point Sens…, Michell Condumax II Hydrocarbon Dew-Poin…, Easidew EA2 Dew Point Transmitter, Michell OptiPEAK TDL600 Moisture in Natu…, Michell Promet EExd Process Moisture Ana…, Michell S8000 Precision Chilled Mirror H…
Gas Analyzer Accessories MICHELL CD2-SX-WDC-002 Water Dew-Point S…, Michell Ceramic Impedance Moisture Senso…, Michell Condumax II CD2-SCR-002 Hydrocar…

Ordering Michell parts: what we need

For Condumax II parts send the analyzer serial number and the part code, such as CD2-SCR-002, and state if a new assembly or the exchange item is wanted. For Easidew transmitters we need the measuring span in dew point or ppm, the process connection thread, the output signal, the operating pressure, the gas and the area certification. A photo of the label on the old transmitter usually answers most of these in one go.

Facts checked against UK Companies House records and PST's Michell 50-year material, October 2026.

FAQ

Questions about the MICHELL Instruments Cermet II Dew-Point Hygrometer,

Answers from our engineering team on installation, compatibility and integration.

01What dew point range does the Cermet II cover?
The Cermet II measures -100 to +20 C dew point, with accuracy of plus or minus 1 C from -60 to +20 C and plus or minus 2 C from -100 to -60 C.
02How far can the Cermet II sensor be from its monitor?
The sensor and monitor can be located up to 1200 m (3930 ft) apart, using the standard 2 m cable plus extension as required.
03What outputs and alarms does the Cermet II have?
The monitor has a 4-20 mA output and RS232 as standard, two isolated 10 A form C relays, and options for 0-10 V, RS485 and two additional relays.
04What pressure can the Cermet II sensor withstand?
The sensor operates from vacuum up to 40 MPa (400 barg, 5801 psig), with a 5/8 inch UNF parallel thread process connection.
05Can the Cermet II compensate for pressure?
Yes. It accepts a 4-20 mA signal from an optional external pressure transducer or a manual pressure factor to correct ppmV, lb/MMscf and g/m3 readings.