Brand: Panametrics
Panametrics MISP2 (MIS Probe 2) Aluminum Oxide Moisture Probe
Description
Panametrics MISP2 overview
The Panametrics MISP2, also written MIS Probe 2 or MIS 2 Probe, is an aluminum oxide moisture sensor probe that belongs to the Moisture Image Series. It measures moisture concentration in gases and in non-aqueous liquids from trace levels up to ambient levels, and it is coupled to a Panametrics hygrometer or analyzer through an interconnecting cable. The MISP2 is a familiar replacement and spare item in plants that already run Panametrics moisture analyzers, which is why buyers often search for it by its exact model name.
What sets the MISP2 apart from a plain sensor element is the electronics module attached to the probe. A built-in microprocessor gives 16-bit resolution so the probe can follow changes of parts per billion in moisture concentration, and it continuously monitors and compensates the electronics for long-term stability. Calibration data is stored in nonvolatile EEPROM, so data entry is automatic after installation and is protected against a power outage. Each aluminum oxide sensor is individually calibrated against known moisture concentrations that are traceable to NIST or NPL.
Key features of the MISP2
- Intrinsically safe design for the Moisture Image Series probe, avoiding explosion-proof housings, purges or external zener requirements in IEC Zone 0 areas.
- Three-function capability: moisture sensor plus optional built-in temperature and pressure sensors.
- 16-bit resolution electronics with nonvolatile calibration storage.
- Calibrations traceable to NIST or NPL for each individually calibrated sensor.
- Requires only unshielded twisted-pair cabling, so the probe can sit up to 3000 ft (914 m) from the analyzer.
- Threaded and VCR mounting configurations are available.
- Operation from vacuum up to 5000 psig (345 bar) with no minimum flow rate required.
MISP2 technical specifications
| Parameter | Value |
|---|---|
| Sensor type | Aluminum oxide moisture sensor probe |
| Overall range capability | -110 to 60 deg C (-166 to 140 deg F) |
| Standard range option | -80 to 20 deg C (-112 to 68 deg F) with data to -110 deg C |
| Ultra-low range option | -110 to -50 deg C (-166 to -58 deg F) |
| High range option | -80 to 60 deg C (-112 to 140 deg F) |
| Calibrated accuracy at 25 deg C | Plus or minus 2 deg C above -100 deg C; plus or minus 3 deg C below -100 deg C |
| Repeatability | Plus or minus 0.2 deg C above -100 deg C; plus or minus 0.5 deg C below -100 deg C |
| Operating pressure | 5 micron Hg to 5000 psig (345 bar), limited by optional pressure sensor |
| Flow range | Gases: static to 10,000 cm/s at 1 atm; liquids: static to 10 cm/s at 1 g/cc |
| Built-in temperature sensor | Nonlinear NTC thermistor, -30 to 70 deg C, plus or minus 0.5 deg C |
| Built-in pressure ranges | 30-300, 50-500, 100-1000, 300-3000 or 500-5000 psig; accuracy 1 percent of full scale |
| Intrinsic safety | ATEX II 1 G Ex ia IIC T4; CSA C US Class I, Division 1, Groups A, B, C and D, T4 |
| Probe to analyzer separation | 3000 ft (0.9 km) maximum recommended; AWG 22 unshielded twisted pair |
| Compatible instruments | moisture.IQ analyzer and PM880 portable hygrometer |
Values above are taken from the manufacturer datasheet for the MISP2. The pressure sensor and the dew point range are chosen at ordering, so confirm the combination you need before requesting a quotation.
Typical applications in Oman and the GCC
In Oman and across the GCC the MISP2 is typically used where trace moisture decides product quality or equipment safety. Natural gas conditioning and dehydration units, petrochemical gas streams, industrial gas production, furnace and heat-treating atmospheres, power generation and air dryers all appear in the application list of the manufacturer. In gas plants the probe is mounted at the sampling point so that the measurement reflects the process rather than a long sample line, and in non-aqueous liquid service the low maximum liquid velocity of 10 cm/s should be respected in the sample system design.
Selection and compatibility notes
When choosing a MISP2, specify the dew point calibration range that matches your process, the pressure range of the built-in sensor, the mounting style (threaded or VCR) and the cable length. The probe is intended for use with the moisture.IQ analyzer and the PM880 portable hygrometer, which use the temperature and pressure signals to convert dew point into ppm, pounds per million standard cubic feet and relative humidity. Check which analyzer generation you have before ordering, because each probe family needs the matching cable and input module.
Installation and commissioning
Install the MISP2 in a sample cell or directly in the line at a point that sees representative, dry-down-friendly gas. Keep sample tubing short and clean, avoid dead legs, and use materials suited to trace moisture work. Connect the probe with unshielded twisted-pair AWG 22 cable to a compatible analyzer, and follow the intrinsic safety wiring instructions of the analyzer manual in hazardous areas. After connection the stored calibration data is read automatically, so no manual entry of calibration constants is needed.
Maintenance and service life
Aluminum oxide sensors drift slowly with time and exposure, so a periodic recalibration or exchange against a recent calibration is good practice. Keep the sensor protected from liquid contamination and from condensation during shutdowns, and store spare probes with their protective covers fitted. The manufacturer states operating temperature of -20 to 60 deg C and storage of -30 to 70 deg C for the moisture sensor, so avoid leaving a probe in a hot closed vehicle or a heated panel.
Related products and ordering notes
Related Panametrics moisture items include the moisture.IQ multichannel analyzer, the PM880 portable hygrometer, the pro.IQ packaged system and the HygroPro transducer element already listed on this site. When requesting the MISP2 please state whether you are replacing an existing probe, and quote the label data from the old one if it is still available. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.
Quick answers about the MISP2
- What is the Panametrics MISP2? The MISP2, also called MIS Probe 2, is an intrinsically safe aluminum oxide moisture probe from the Moisture Image Series that can include built-in temperature and pressure sensors.
- How far can the MISP2 be from the analyzer? The manufacturer recommends up to 3000 ft (0.9 km) using unshielded twisted pair AWG 22 cable, and suggests consulting the factory for longer distances.
- Which instruments work with the MISP2? The datasheet lists the moisture.IQ analyzer and the PM880 portable hygrometer as compatible instruments for the MIS Probe 2.
Documents and quotation
A manufacturer datasheet for the MISP2 is available from Seven Star LLC on request and is attached to this listing where provided. To receive a quotation, send the RFQ with the part number MISP2, the quantity required and your delivery location, together with any hazardous area, range or configuration details from your project specification.
Datasheets & Documents
- Datasheet Panametrics MISP2 (MIS Probe 2) Aluminum Oxide Moisture Probe Datasheet Download
Reviews (0)
How would you rate this product? Cancel reply
About brand
Panametrics is an American maker of ultrasonic flowmeters and moisture, oxygen and gas analysers. It was founded in 1960 and was based in Waltham, Massachusetts, when GE bought it in 2002. It later became a Baker Hughes business, and since 1 January 2026 it has belonged to Crane Company. Its long-running site in Shannon, Ireland, opened in 1973.
Seven Star LLC lists Panametrics flowmeters and analyzers for buyers in Oman: DigitalFlow gas and flare meters, clamp-on and portable liquid meters, and trace moisture analyzers for natural gas and process gases. We are an independent supplier, so each inquiry is checked against the manufacturer's order string before a quotation goes out.
Panametrics company profile
| Founded | 1960, Massachusetts, USA |
|---|---|
| Main sites | Billerica, Massachusetts, and Shannon, Ireland (since 1973) |
| Ownership | Crane Company since 1 January 2026; previously GE (from 2002) and Baker Hughes |
| Key technology | Transit-time ultrasonic flow measurement and aluminium oxide moisture sensing |
| Product families | TransPort PT900 portable and DigitalFlow fixed ultrasonic flowmeters for liquid, gas and flare gas, aluminium oxide moisture probes and hygrometers, Aurora laser moisture analysers, XMO2 oxygen and XMTC thermal conductivity transmitters |
Milestones
- 1960: Panametrics is founded.
- 1973: The Shannon, Ireland, site is established.
- 2002: GE acquires Panametrics.
- 2026: Crane Company completes its purchase of Baker Hughes' Precision Sensors & Instrumentation, including Panametrics.
Panametrics product families on this page
The listed items follow the manufacturer's own grouping:
- Gas and flare flow: DigitalFlow GF868 flare gas meter, GM868 and XGM868i gas flow transmitters
- Liquid flow: XMT1000 transmitter, AT600 AquaTrans clamp-on meter, TransPort PT900 portable meter
- Moisture: moisture.IQ multichannel analyzer, HygroPro II aluminum oxide transmitter, Aurora TDLAS analyzer for natural gas
Older units may carry GE Panametrics or GE Sensing on the nameplate. They belong to the same product lines, and the model string is read the same way.
Where Panametrics flowmeters and moisture analyzers are used
Flare meters report flared volumes on production sites, refineries and LNG plants. Clamp-on and portable ultrasonic meters check water, hydrocarbon liquid and utility lines without cutting the pipe. Aluminum oxide and TDLAS moisture analyzers watch dehydration units, gas pipelines, hydrogen and industrial gas streams. The manufacturer also names chemical processing, energy and water and wastewater among its markets, which matches what GCC operators ask us for.
Panametrics products we list
Ordering Panametrics equipment: what we need
A Panametrics model is a long configuration string, and the transducers or probes are ordered separately. Please send the full model and serial number of the electronics, the transducer or moisture probe part number, pipe size, schedule and material, fluid, pressure and temperature, flow range, hazardous area approval and output or protocol. For a flare meter add the flare gas composition range and velocity range. For moisture, give the dew point range and whether sample system components are required.
Facts checked against GE's 2002 acquisition release, Crane Company's January 2026 completion release and Irish press, October 2026.
FAQ
Questions about the Panametrics MISP2 (MIS Probe 2) Aluminum Oxide
Answers from our engineering team on installation, compatibility and integration.
01What is the Panametrics MISP2?
The MISP2, also called MIS Probe 2, is an intrinsically safe aluminum oxide moisture probe from the Moisture Image Series that can include built-in temperature and pressure sensors.
02How far can the MISP2 be from the analyzer?
The manufacturer recommends up to 3000 ft (0.9 km) using unshielded twisted pair AWG 22 cable, and suggests consulting the factory for longer distances.
03Which instruments work with the MISP2?
The datasheet lists the moisture.IQ analyzer and the PM880 portable hygrometer as compatible instruments for the MIS Probe 2.
04What pressure can the MISP2 withstand?
The probe operates from 5 micron Hg up to 5000 psig (345 bar), limited by the optional built-in pressure sensor range that is selected at ordering.
05Does the probe store its own calibration data?
Yes. Calibration data is kept in nonvolatile EEPROM in the probe electronics, so it loads automatically and survives a power outage.