Brand: Tektronix
Tektronix Keithley DMM7510 7.5-Digit Sampling Multimeter
Description
DMM7510 is a Keithley 7.5-digit graphical sampling multimeter from Tektronix: a precision DMM with a 14 ppm basic one-year DC voltage accuracy and a built-in 18-bit, 1 MS/s digitizer, shown on a five-inch capacitive touchscreen.
A reference-grade meter that can also draw waveforms
Resolution runs from 3.5 to 7.5 digits. The digitizing functions use the same ranges as the DC voltage and current functions, so a transient can be captured with the full dynamic range of the meter, and the voltage digitizer keeps the 10 GΩ input impedance option to limit loading of the circuit under test.
The buffer holds more than 11 million readings in standard mode or 27.5 million in compact mode, and the graphing utility compares up to four reading buffers on one display. Autocalibration reduces drift with temperature and time, and the datasheet gives two-year specifications for longer calibration cycles.
Key features as printed in the datasheet
- 100 mV, 1 Ω and 10 µA lowest ranges for sleep-mode currents and low-level signals.
- Offset-compensated ohms, four-wire and dry-circuit functions.
- GPIB, USB-TMC and LXI-compliant LAN, with a front USB memory port.
- TSP script language plus SCPI; TSP-Link expansion.
- DMM7510-RACK version without a handle for rack use.
Technical specification
DC voltage accuracy ±(ppm of reading + ppm of range), 1 year
| Parameter | Specification |
|---|---|
| 100 mV (10 nV resolution) | 18 + 9 |
| 1 V (100 nV) | 15 + 2 |
| 10 V (1 µV) | 14 + 1.2 |
| 100 V (10 µV) | 40 + 5 (22 + 5 within 30 days of autocalibration) |
| 1000 V (100 µV) | 40 + 5 |
| 10 V, 24 hour | 2 + 0.7 |
| ADC linearity | 1.0 ppm of reading + 1.0 ppm of range |
| Input impedance | >10 GΩ or 10 MΩ ±1 % (100 mV to 10 V); 10 MΩ ±1 % (100 V, 1000 V) |
| Rejection | NMRR 110 dB at 5 PLC with line sync; CMRR 140 dB at 1 and 5 PLC |
Resistance and protection
| Parameter | Specification |
|---|---|
| Resistance 1 Ω range, 1 year | 30 + 50 ppm (test current 10 mA) |
| Resistance 1 kΩ | 24 + 3 ppm (1 mA) |
| Resistance 1 MΩ | 30 + 4 ppm (10 µA) |
| Resistance 100 MΩ | 2000 + 30 ppm |
| Input protection | 1010 V DC (715 V RMS AC) on HI and LO; 350 V on sense terminals |
| Current fuses | 3.5 A 1 kV fast-blow (3 A range) and 11 A 1 kV fast-blow (10 A range) |
| Common-mode voltage | 500 V peak LO to chassis |
Digitizer, digital I/O and general
| Parameter | Specification |
|---|---|
| Digitizer | 1 MS/s, 18-bit, voltage or current |
| Digital I/O | Six lines on a 9-pin D; 0.7 V low, 3.7 V high; 5 V pin 500 mA |
| Interfaces | LAN (LXI 1.4 Core 2011), GPIB, USB-TMC, USB host |
| Display | 5 in. capacitive touch TFT, 800 x 480 |
| Power | 100 to 240 V; 50/60 Hz auto-sensed; 60 VA |
| Size, bench | 106 x 255 x 425 mm; without handle 88 x 213 x 410 mm (H x W x D) |
| Shipping weight | 4.08 kg with bumpers and handle; 3.63 kg without |
| Environment | 0 to 50 °C, ≤80 % RH at 35 °C, up to 2000 m; storage -30 to 70 °C |
| Warm-up | 90 minutes to rated accuracy |
| Safety as printed | NRTL listed to UL61010-1 and CSA C22.2 No 61010-1; EU Low Voltage and EMC Directives; MIL-PRF-28800F Class 3 vibration |
Ordering and services
Order DMM7510 for the bench version or DMM7510-RACK for the rack version with no handle. Included are the quick start guide, power cord, 1 m USB cable and 3 m LAN cable. Extended warranties (2, 3 or 5 years) and KeithleyCare standard, data and ISO 17025 calibration plans are listed as services. Test leads, Kelvin probes and rack kits are separate.
Operating notes
Plan for the 90-minute warm-up to rated accuracy and run autocalibration when ambient temperature changes, because the best DC specification applies within 30 days of autocalibration and within ±5 °C of the autocalibration temperature. Above 500 V, add 0.02 ppm per volt to the reading error. All specifications apply to front or rear terminals except the 10 A range, which is rear terminals only.
Calibration labs and electronics R&D
The very low resistance and current ranges suit bonding and contact-resistance checks, battery-powered device current profiling and transmitter calibration reference work. A calibration laboratory serving oil, gas and utilities customers in the region can use it as a check standard for lower-class meters, subject to its own accreditation. It is an indoor laboratory instrument with no ingress or hazardous-area rating printed.
Leads, probes and software
Keithley lists the 1754 universal lead kit, 5804 safety lead kit, 5805 and 5808 Kelvin probes, 5806 Kelvin clips, CA-18-1 shielded banana cable, GPIB cables and adapters, 8501-1 trigger link cable and 4299-8 and 4299-9 rack kits, plus KickStart and TSP Toolkit software. See the Portable Test and Measurement category.
Selection and installation notes
A bench multimeter is selected from the measurement accuracy, the resolution and the interfaces needed. Allow the warm-up time, use good-quality leads, keep the inputs inside the stated limits and use remote sense or shielding where low-level signals are measured. Plan the calibration interval according to your quality system.
Information needed for a quotation
Send the full model code, the quantity, the delivery location in Oman or the GCC, the intended use and any accessories needed, so that the complete package can be quoted together.
Documents and supply information
The manufacturer data sheet is attached to this page and should be read together with the project specification and the site risk assessment before a purchase is confirmed. Ratings, standards and dimensions are quoted as printed on that sheet. Lead time, stock position and certificates available with the shipment are confirmed in the quotation, and we supply to Oman and the wider GCC region for industrial, utility and building services projects.
Frequently asked questions
What is the basic DC accuracy of the DMM7510?
14 ppm of reading plus 1.2 ppm of range on the 10 V range over one year.
What does the built-in digitizer offer?
18-bit resolution at up to 1 MS/s for voltage or current waveforms.
How many readings can it store?
More than 11 million in standard mode or 27.5 million in compact mode.
How long should it warm up?
The datasheet requires 90 minutes to reach rated accuracy.
Does Seven Star LLC act as a Tektronix distributor?
No, we are an independent supplier in Muscat and supply Tektronix items on request rather than acting as a Tektronix channel partner.
Documents and quotation
Download the official Tektronix datasheet for DMM7510 from this page. As an independent supplier based in Muscat, Seven Star LLC supplies Tektronix items on request: use the quote form with the quantity you need. Check every value in the document against your own specification before you order. Browse similar items in the category.
Datasheets & Documents
- Datasheet Keithley DMM7510 datasheet (PDF) Download
Reviews (0)
How would you rate this product? Cancel reply
About brand
Tektronix is an American test and measurement company, founded in 1946 as Tekrad and headquartered in Beaverton, Oregon. It designs oscilloscopes, signal sources, multimeters, power analyzers and source measure instruments, and its Keithley line covers precision bench meters and SourceMeter units.
Oscilloscopes, meters and signal sources
Tektronix TBS2000, 2 Series, 3 Series MDO and 4 Series B scopes serve electronics labs, service benches and training centres, while the Keithley DMM6500, DMM7510 and 2460 and the DMM4050 and PA1000 handle precision and power measurement.
Supply from Muscat
Seven Star LLC, an independent supplier in Muscat, quotes Tektronix instruments for Oman and the GCC with the official datasheet beside each listing.
Related products
Frequently Asked Questions
14 ppm of reading plus 1.2 ppm of range on the 10 V range over one year.
18-bit resolution at up to 1 MS/s for voltage or current waveforms.
More than 11 million in standard mode or 27.5 million in compact mode.
The datasheet requires 90 minutes to reach rated accuracy.
No, we are an independent supplier in Muscat and supply Tektronix items on request rather than acting as a Tektronix channel partner.