Michell SF72-TX Rail Dew-Point Transmitter

Description

Michell SF72-TX rail-approved dew-point transmitter

The Michell SF72-TX is a compact two-wire dew-point transmitter whose 316 stainless-steel body and glass-to-metal seal were developed for railway pneumatics and other vibrating, high-pressure environments. Standard output is 4 to 20 mA scaled -60 to +60 °C dew point, with alternative scaling available on request. The datasheet describes its key measuring points as -50 to +10 °C dew point, the range where an air dryer on a train, compressor skid or industrial plant must prove it is working. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.

Built for shock, vibration and pressure

Rail approval is the feature that separates SF72-TX from its sister dew-point probes. Shock and vibration testing follows IEC 61373:1999-01 (EN 50155), while RFI and EMC emissions and immunity are certified to EN 50121-3-2:2016. The pressure rating is exceptional for a sensor of this size: 45 MPa (450 barg, 6526 psig) operating, with an overpressure rating of twice that, 90 MPa (900 barg, 13053 psig). The polymer-based sensor is calibrated on a high-volume traceable system, which is why the transmitter is offered in OEM quantities.

SF72-TX specification summary

Part number SF72-TX
Measurement range -60 to +60 °C (-76 to +140 °F) dew point
Accuracy ±2 °C (±3.6 °F) dew point
Repeatability ±0.2 °C (±0.36 °F) dew point
Long-term stability less than 1 percent per year at reference conditions
Calibration traceable 8-point certificate
Output 4 to 20 mA, 2-wire current source
Supply voltage 12 to 28 V DC
Load resistance maximum 250 Ω at 12 V, 500 Ω at 24 V
Ambient operating temperature -40 to +50 °C (gas -40 to +60 °C)
Operating pressure 45 MPa (450 barg / 6526 psig)
Flow 1 to 5 Nl/min in a standard sampling block; 0 to 10 m/s direct insertion
Housing and protection 316 stainless steel, IP66
Process connection G1/2 inch BSP, 316 stainless steel
Electrical connection M12, 4 pin, A coded
Dimensions and weight 119 mm long including filter x ø27 mm; 150 g

Wiring, diagnostics and communication of SF72-TX

Only pins 3 and 4 of the M12 plug are used for the loop: pin 3 is the 4 to 20 mA signal and pin 4 the positive supply. The programmed fault states are useful for a control system: a sensor fault drives 23 mA, under-range dew point drives 4 mA and over-range drives 20 mA. A two-wire RS485 Modbus RTU channel is provided for digital diagnostics, and the optional SF72-CK communication kit is the tool for setting it up. Note that the universal communications kit EAU-CK in the manufacturer’s order codes has a 5-way M12 lead that is explicitly not suitable for the SF72 rail M12 4-pin connector.

How SF72-TX differs from other Michell dew-point transmitters

Compare SF72-TX with the other probes that appear in the same manufacturer documents. It has a 4-pin A-coded M12 connector that carries the rail approval, an operating rating of 450 barg, and a G1/2 inch BSP process thread, whereas the Easidew-type sensors in the EA2-SAM sampler and the Cermet order codes use a 5/8 inch UNF thread. Its guard is an HDPE type for particles under 10 µm, and the sample block that fits is the SB-12 with 1/8 inch NPT ports, not the 5/8 inch UNF blocks of the Easidew family. The communications kit is also specific: SF72-CK rather than the universal EAU-CK. Always confirm thread and connector before ordering, because a transmitter that does not match the block cannot be adapted in the field.

Where SF72-TX is applied

Rail vehicles (compressed air for brakes and doors), industrial dryers, compressors and mobile equipment use SF72-TX because it tolerates constant vibration. In Oman and the GCC, high-pressure gas skids, tyre-inflation and breathing-air systems, and port equipment can use the same sensor where reliability under vibration is needed. The datasheet’s emphasis on response speed, reliability and long-term data stability suits control loops that act on dew point.

Why dew point matters on rail and compressed-air systems

Moisture that survives a dryer condenses in cold pipework, corrodes valves and freezes in brake and door circuits, so a transmitter at the dryer outlet is the first warning of a failed desiccant bed or refrigerant stage. SF72-TX keeps working near the freezing point of water vapour and also well below it, covering -60 to +60 °C dew point with ±2 °C accuracy and ±0.2 °C repeatability. The long-term stability figure of under 1 percent per year at reference conditions means calibration drift stays modest between services. A thermal compensation characterised over the operating range corrects the sensor for gas temperature, so a transmitter mounted on a hot engine room manifold still reads correctly inside its -40 to +60 °C gas limit.

Installation notes for SF72-TX

Mount the transmitter in the SB-12 sample block with the supplied bonded seal or into a G1/2 inch BSP port for direct insertion in flow up to 10 m/s. Keep the connector accessible, fit the HDPE guard, and wire with a supply of 12 to 28 V DC and a load within the 250 Ω at 12 V limit. Output scaling beyond -60 to +60 °C can be requested before dispatch.

Spares, accessories and service for SF72-TX

The accessory table lists EA2-HDPE (pack of 10 guards), SB-12 (G1/2 inch 316 stainless sample block with 1/8 inch NPT ports), GEN-LABEL (two replacement labels), SF72-CK (communication kit), DP-MC (BS EN 10204 3.1 material certificate) and BS-12-PK5 (five bonded seals/O-rings). A recalibration service for the transmitter is listed as READINGS.

Documents and quotation

The SF72 Rail datasheet in PDF form is supplied with this listing. For a quotation on SF72-TX, send an RFQ with the part number, quantity, any non-standard output scaling, and accessories such as SB-12 or SF72-CK.

Datasheets & Documents
  • PDF Datasheet Michell SF72-TX Rail Dew-Point Transmitter Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Michell Instruments is a British manufacturer of humidity and dew point instruments, founded in 1974 in Cambridge, England, by Andrew Michell and incorporated on 12 September 1974. It worked with Cambridge University scientists on an aluminium oxide dew point sensor and later developed chilled mirror hygrometers and hydrocarbon dew point analysers. Michell is based in Ely, Cambridgeshire, and is part of Process Sensing Technologies (PST).

Seven Star LLC lists Michell dew-point transmitters, the Condumax II analyzer and its replacement sensor parts, and supplies them to gas plants and compressed air users in Oman by exact order code.

MICHELL Instruments company profile

Founded 1974, Cambridge, England, by Andrew Michell
Headquarters Ely, Cambridgeshire, United Kingdom, since 2007
Ownership Process Sensing Technologies (PST)
Technology Ceramic and aluminium oxide moisture sensors, chilled mirror dew point, Condumax hydrocarbon dew point (licensed from Shell)
Product families Easidew and Promet dew point transmitters, S8000 and Optidew chilled mirror hygrometers, Condumax II hydrocarbon dew point analysers, MDM300 portable dew point meters, oxygen analysers

Milestones

  • 1974: Andrew Michell founds Michell Instruments in Cambridge.
  • 2007: The company moves to Ely, Cambridgeshire.
  • 2024: Michell marks 50 years, now as part of PST.

Michell product families on this page

The Condumax II is an automatic hydrocarbon dew-point analyzer for natural gas. Several of its parts are listed separately: the CD2-SCR-002 hydrocarbon dew-point sensor cell, the CD2-WDC-002 water dew-point sensor assembly and the CD2-SX-WDC-002, which is the exchange version of that water dew-point sensor. The Easidew line is represented by the Easidew EA2 dew point transmitter and by the Easidew PRO I.S., an intrinsically safe moisture transmitter for hazardous areas. A ceramic impedance moisture sensor module is also listed, the sensing element behind these transmitters.

Where Michell analyzers are used

Hydrocarbon and water dew point are contract quality parameters for natural gas, so the Condumax II is found at gas processing plants, pipeline entry points and custody transfer stations. Easidew transmitters monitor dryers for instrument air and industrial gases, glove boxes, and dry gas feeds to processes that cannot tolerate water.

MICHELL Instruments products we list

Family Examples
Gas Analyzer Systems Easidew PRO I.S. Intrinsically Safe Mois…, Michell CD2-WDC-002 Water Dew-Point Sens…, Michell Condumax II Hydrocarbon Dew-Poin…, Easidew EA2 Dew Point Transmitter, Michell OptiPEAK TDL600 Moisture in Natu…, Michell Promet EExd Process Moisture Ana…, Michell S8000 Precision Chilled Mirror H…
Gas Analyzer Accessories MICHELL CD2-SX-WDC-002 Water Dew-Point S…, Michell Ceramic Impedance Moisture Senso…, Michell Condumax II CD2-SCR-002 Hydrocar…

Ordering Michell parts: what we need

For Condumax II parts send the analyzer serial number and the part code, such as CD2-SCR-002, and state if a new assembly or the exchange item is wanted. For Easidew transmitters we need the measuring span in dew point or ppm, the process connection thread, the output signal, the operating pressure, the gas and the area certification. A photo of the label on the old transmitter usually answers most of these in one go.

Facts checked against UK Companies House records and PST's Michell 50-year material, October 2026.

FAQ

Questions about the Michell SF72-TX Rail Dew-Point Transmitter

Answers from our engineering team on installation, compatibility and integration.

01What makes SF72-TX different from a standard dew-point probe?
SF72-TX is rail certified: EN 50121-3-2:2016 for EMC and IEC 61373:1999-01 (EN 50155) for shock and vibration, in a 316 stainless IP66 body.
02What pressure can SF72-TX withstand?
Operating pressure is 45 MPa (450 barg, 6526 psig), with an overpressure rating of twice that, 90 MPa (900 barg, 13053 psig).
03How is SF72-TX wired?
Use the M12 4-pin A-coded connector: pin 3 carries the 4 to 20 mA signal and pin 4 the positive supply from 12 to 28 V DC. RS485 Modbus RTU is provided for diagnostics.
04What output does SF72-TX give on a fault?
The programmed diagnostics are 23 mA for a sensor fault, 4 mA for under-range dew point and 20 mA for over-range dew point.
05Which sample block fits SF72-TX?
The datasheet lists SB-12, a G1/2 inch 316 stainless steel block with 1/8 inch NPT inlet and outlet ports, plus BS-12-PK5 seals for the process connection.