Chauvin Arnoux CA 6541 1 kV Insulation Tester

Description

CA 6541 is a battery-powered 1 kV megohmmeter from Chauvin Arnoux for site insulation testing at 50, 100, 250, 500 and 1000 V DC, with a measuring range up to 4 TΩ, continuity at over 200 mA, resistance, capacitance and voltage measurement in a rugged case with a back-lit display and bargraph.

Insulation, continuity and PI testing in one site case

According to the manual, the CA 6541 and CA 6543 are portable megohmmeters in a rugged construction-site casing with a cover. They measure voltage, insulation, continuity, resistance and capacitance, and show the result numerically and as a bargraph. The CA 6541 runs on eight LR14 (C) cells; the CA 6543 has a rechargeable NiMH pack, can run directly from the mains, and adds memory, a real-time clock and an RS232 interface for PC control and printing.

The CA 6541 is therefore the simpler of the two: a nine-position rotary switch picks the function, six keys handle smoothing, alarm, timer, the DAR and PI ratios and the R(t) curve function, and one START/STOP button runs the test. Automatic detection of AC or DC voltage on the terminals inhibits or stops a measurement, and the item under test is discharged automatically at the end.

Features listed in the manual

  • Five test voltages: 50, 100, 250, 500 and 1000 V DC, nominal current above 1 mA at the nominal voltage.
  • Insulation resistance from 2 kΩ at 50 V to 4 TΩ at 1000 V.
  • DAR and PI calculation from 0.000 to 9.999.
  • Continuity at over 200 mA with lead compensation limited to 5 Ω, plus resistance measurement.
  • Capacitance measurement after discharge, 0.005 to 4.999 µF.
  • Digital filtering, programmable alarm thresholds, timer and defective-fuse detection.
  • Automatic discharge of residual high voltage.

Technical specification

Insulation resistance

Parameter Specification
Test voltages Un 50, 100, 250, 500, 1000 V DC
Off-load voltage Up to 1.1 x Un ±5 V
Short-circuit current Below 6 mA DC
Range at 50 V 2 kΩ to 200 GΩ
Range at 100 V 4 kΩ to 400 GΩ
Range at 250 V 10 kΩ to 1 TΩ
Range at 500 V 20 kΩ to 2 TΩ
Range at 1000 V 40 kΩ to 4 TΩ
Intrinsic uncertainty ±(5 % L + 3 ct) up to 399.9 GΩ; ±(15 % L + 10 ct) above
Discharge to 25 V 2.8 s from 1000 V, 2.2 s from 500 V, 1.7 s from 250 V, 1 s from 100 V, 0.5 s from 50 V
DAR and PI 0.000 to 9.999, uncertainty 5 % L

Voltage, continuity, resistance and capacitance

Parameter Specification
Voltage 1 to 1000 V, DC and 16 to 420 Hz, ±(1 % L + 3 ct), input about 750 kΩ
DC voltage after test 25 to 1000 V, ±(1 % L + 3 ct)
Continuity 0.01 to 39.99 Ω, ±(3 % L ± 4 ct), over 200 mA up to 20 Ω
Continuity open-circuit voltage Below 12.4 V DC
Resistance 0.01 Ω to 399.9 kΩ in five ranges, ±(3 % L + 3 ct)
Capacitance 0.005 to 4.999 µF, ±(10 % L + 1 ct)
Overvoltage 1200 V AC and DC for 10 s between + and –

General specification

Parameter Specification
Power supply 8 x 1.5 V LR14 (C) batteries
Life per set Typically 21,000 insulation measurements of 5 s with a 20 s pause at the nominal load
Operating range -10 to +55 °C, 10 to 80 % RH, indoor or outdoor without rain
Storage (no batteries) -40 to +70 °C
Altitude, pollution degree Below 2000 m, degree 2
Dimensions (L x W x H) 240 x 185 x 110 mm
Weight About 3.4 kg
Protection IP 53, IK 04
Safety IEC/EN 61010-2-034, 600 V CAT III; IEC 61557 parts 1, 2, 4 and 10; EMC IEC/EN 61326-1
Fuses F1 (+ terminal) 2.5 A, 1.2 kV, 8 x 50 mm; F2 (G terminal) 0.1 A, 660 V, 6.3 x 32 mm

CA 6541 or CA 6543

The ranges and test voltages are the same in both models. Choose the CA 6541 for routine maintenance work where results are written down on the job and battery cells are convenient; choose the CA 6543 if you need stored measurements (about 1,500 insulation tests are quoted for it), a real-time clock and PC or printer connection. Confirm the model code before ordering because the two share this manual.

Setup and precautions

The CA 6541 is supplied in a cardboard box with a shoulder bag containing red and blue 1.5 m safety leads, a black guarded lead, three crocodile clips, a test probe, eight LR14 cells, a quick start guide, a safety data sheet and a verification certificate. To fit the cells, switch to OFF with all terminals disconnected, release the four captive screws, remove the front panel and mind the seal.

The manual specifies leads and instrument for 600 V in category III, tells the user to use only the leads and accessories supplied, to keep fingers behind the guard and to fit the correct fuse type.

Typical work in Oman and the GCC

Electrical contractors and maintenance teams use a 1 kV megohmmeter for insulation checks on low-voltage cables, switchboards, motors and heaters, and the PI and DAR ratios help judge motor windings over time. The -10 to +55 °C operating range suits outdoor site work in summer; as the manual notes for all insulation testing, take care when the temperature is near the dew point because humidity affects the reading.

Accessories and spares

The manual lists a remote-control probe, a set of two test probes, and a set of three 3 m safety leads (red, blue and guarded black) as accessories; spares include 1.5 m lead sets, crocodile clips, accessory bag no. 6 and both fuse types. For the CA 6543 it also lists a serial printer and adapter. Related instruments are in the Portable Test & Measurement category.

Frequently asked questions

Which test voltages does the CA 6541 provide?

50, 100, 250, 500 and 1000 V DC, each with a nominal current above 1 mA.

How high a resistance can it read?

Up to 4 TΩ at 1000 V, 2 TΩ at 500 V, 1 TΩ at 250 V, 400 GΩ at 100 V and 200 GΩ at 50 V.

Can it measure continuity?

Yes, from 0.01 to 39.99 Ω at over 200 mA, with test lead compensation limited to 5 Ω.

Does it calculate PI and DAR?

Yes. The DAR and PI ratios are calculated over the range 0.000 to 9.999.

Does Seven Star LLC act as a Chauvin Arnoux distributor?

No, we are an independent supplier in Muscat and supply Chauvin Arnoux items on request rather than acting as a Chauvin Arnoux channel partner.

Documents and quotation

Need CA 6541? The Chauvin Arnoux user manual is attached to this listing. Seven Star LLC (independent supplier, Muscat) can quote the item and arrange supply on request; list quantity and site location in your enquiry. Verify the code, dimensions and approvals in the document before ordering. Other products are in the category page.

Datasheets & Documents
  • PDF Datasheet CA 6541 / CA 6543 user manual (PDF) Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Chauvin Arnoux is a French maker of electrical test and measuring instruments, founded in Paris in 1893 by engineer Raphaël Chauvin, who was joined in partnership by René Arnoux in 1894. The company built galvanometers in its early years and a universal tester in 1927. Chauvin Arnoux is still family-run, with its head office in Asnières-sur-Seine near Paris.

Seven Star LLC is an independent supplier in Muscat. We list Chauvin Arnoux testers for electrical contractors and maintenance teams in Oman and the GCC.

Chauvin Arnoux company profile

Founded 1893, Paris, France, by Raphaël Chauvin
Partner René Arnoux, from 1894
Headquarters Asnières-sur-Seine, France
Ownership Family-run group
Product families C.A 6116 installation testers, C.A 6547 and C.A 6555 insulation testers, C.A 8336 and C.A 8345 power quality analysers, F400 series clamp meters, earth testers, Metrix multimeters and oscilloscopes

Milestones

  • 1893: Raphaël Chauvin founds the company in Paris.
  • 1894: René Arnoux becomes a partner.
  • 1927: The company introduces a universal tester.
  • 2018: Chauvin Arnoux marks 125 years.

Chauvin Arnoux product series

Insulation testers include the CA 6541 1 kV tester, the CA 6505 5 kV tester for up to 10 TΩ and the CA 6550 10 kV tester. Earth testers include the CA 6422 2P/3P tester, the CA 6471 earth and resistivity tester and the CA 6417 earth loop clamp meter.

The CA 6240 is a 10 A micro-ohmmeter for contact and joint resistance, the CA 6116N a multifunction installation tester, the CA 5273 a TRMS digital multimeter and the CA 1954 a thermal imaging camera.

Where Chauvin Arnoux instruments are used

Insulation testing of cables, motors and transformers, earthing and lightning protection checks, contact resistance of breakers and busbars, testing of new installations, and thermal inspection of panels.

Chauvin Arnoux products we list

Family Examples
Portable Test & Measurement CA 6240 10 A Micro-Ohmmeter, CA 6417 Earth Loop Clamp Meter, CA 6422 Earth Tester 2P/3P up to 2 kΩ, CA 6471 Earth and Resistivity Tester, CA 6505 5 kV Insulation Tester up to 10 TΩ, CA 6541 1 kV Insulation Tester, CA 6550 10 kV Insulation Tester

Ordering Chauvin Arnoux instruments: what we need

Send the model, the accessories or leads required and whether a calibration certificate is needed.

Facts checked against Chauvin Arnoux's anniversary material and French heritage records, October 2026.

Frequently Asked Questions

Which test voltages does the CA 6541 provide?
50, 100, 250, 500 and 1000 V DC, each with a nominal current above 1 mA.
How high a resistance can it read?
Up to 4 TΩ at 1000 V, 2 TΩ at 500 V, 1 TΩ at 250 V, 400 GΩ at 100 V and 200 GΩ at 50 V.
Can it measure continuity?
Yes, from 0.01 to 39.99 Ω at over 200 mA, with test lead compensation limited to 5 Ω.
Does it calculate PI and DAR?
Yes. The DAR and PI ratios are calculated over the range 0.000 to 9.999.
Does Seven Star LLC act as a Chauvin Arnoux distributor?
No, we are an independent supplier in Muscat and supply Chauvin Arnoux items on request rather than acting as a Chauvin Arnoux channel partner.