Brand: H2scan
H2scan HY-OPTIMA 700B General Purpose In-Line Hydrogen Analyzer
Description
A compact 24 V hydrogen analyzer for non-hazardous process lines
HY-OPTIMA 700B is the general purpose member of the H2scan in-line hydrogen analyzer range. It is a small solid-state unit, 236 mm long and 0.4 kg, that reads hydrogen directly in a process gas stream without reference gas, carrier gas or consumables, and with no cross sensitivity to other gases according to the datasheet. The 700B series contains models 710B, 720B, 730B and 740B.
Communication is by a 4-20 mA analog output or by RS232 or RS422 serial, and the unit includes a 1 A / 30 VDC SPDT relay output. Power is 10 to 26 VDC at 10 W, which suits an analyzer cabinet or skid supply. The adapter fitting is 1/2 in MNPT.
The datasheet states that once installed and field calibrated the analyzer typically needs only a quick calibration about every three months with readily available primary standard gases spanning the expected range.
Technical specifications
| Parameter | Specification |
|---|---|
| Operating pressure at analyzer | Recommended 0.95-1.1 ATM absolute (14.0-16.1 psia); maximum 2 ATM absolute (29.4 psia) |
| Process gas temperature | -20 to 60°C |
| Flow rate | 0.1 to 10 slpm |
| Operating humidity | Below 95% RH, non-condensing |
| Calibration interval | 90 days |
| Analog output | 4-20 mA |
| Serial output | RS232 or RS422 |
| Relays | 1 A / 30 VDC SPDT |
| Input voltage and power | 10-26 VDC; 10 W |
| Dimensions | 236.22 x 86.36 x 35.56 mm (9.3 x 3.4 x 1.4 in) |
| Weight | 0.4 kg (0.8 lbs) |
| Adapter fitting | 1/2 in MNPT |
| Operating temperature | -20 to 55°C (-4 to 131°F) |
| Storage temperature | -40 to 80°C (-40 to 176°F) |
| Certifications | UL, CE (as listed on the datasheet) |
Model 710B, 720B, 730B and 740B compared
Models 710B, 730B and 740B are intended for dry gas streams where hydrogen is always present and can be exposed to hydrogen continuously. Model 720B is for processes where hydrogen appears only occasionally, for short periods, in an air, oxygen or nitrogen background, such as during a leak or upset.
| Model | H2 range | H2 must be present | CO / H2S limits | T90 response | Accuracy | Calibration gas background |
|---|---|---|---|---|---|---|
| 710B | 0.1% to 10% | Yes | CO below 100 ppm; H2S below 20 ppm | Below 90 s | 0.15% | N2 |
| 720B | 0.4% to 5% | No | 0; 0 | Below 60 s | 0.3% | O2, N2 |
| 730B | 0.5% to 100% | Yes | CO below 100 ppm; H2S below 1000 ppm | Below 60 s | 0.3% (to 10%); 1.0% (10 to 100%) | N2 |
| 740B | 0.5% to 100% | Yes | CO up to 20%; H2S up to 3% | Below 90 s | 0.3% (to 10%); 1.0% (10 to 100%) | N2 |
All accuracy figures are absolute, assume a dry stream at 25°C and constant pressure, and are in addition to errors in the calibration gases.
Where the 700B fits in oil, gas and power plants
The datasheet lists fuel cells (anode and cathode side), hydrogen in natural gas or biomethane, air separation, steam methane reforming, electrolysis process streams, laboratory research, metals annealing, semiconductors and oil hydrogenation. In a GCC context this points to hydrogen blending trials, electrolyzer skids, annealing furnaces and utilities where the analyzer sits in a non-classified, sheltered location.
Because the 700B is general purpose, it is not an alternative to the explosion proof 2700 series when the analyzer itself must be installed in a classified area; for that case compare the gas analyzer systems range or ask us for the 2700 or the intrinsically safe 5330 series.
Commissioning checklist and how it compares with sister analyzers
Before installing, confirm four things from the datasheet: the stream is dry (all performance figures assume a dry stream at 25°C and constant pressure), carbon monoxide and hydrogen sulfide stay under the limit of the chosen model, the line pressure sits between 0.95 and 1.1 ATM absolute for best results, and the flow through the sensor is between 0.1 and 10 slpm. If the stream pressure varies, note that the 700B figures assume constant pressure whereas the 2700 series figures assume pressure compensation, which is a practical reason to prefer the 2700 on variable-pressure lines.
Against the other H2scan process analyzers, the 700B is the smallest and lightest at 0.4 kg and the only one in this group powered from a nominal 24 V supply with RS232 or RS422 serial. The explosion proof 2700 series weighs 2.7 kg and runs from 90 to 240 VAC, while the HY-OPTIMA 5232 weighs 0.75 kg, adds self-calibration and Modbus RS-485, and is limited to 0.4 to 5% hydrogen.
Plan for the three-monthly calibration visit: primary standard gases spanning the operating range are required, and the background gas must match the model (nitrogen for 710B, 730B and 740B; oxygen or nitrogen for 720B).
Sample system and calibration planning
Because the 700B works best at 0.95 to 1.1 ATM absolute and 0.1 to 10 slpm, it is usually installed after a pressure-reducing sample conditioning panel rather than directly in a high-pressure line. General engineering note (not from the datasheet): the panel normally includes a filter, a pressure regulator, a flow indicator and a vent line, plus a connection for calibration gas so that the 90-day check can be done without removing the analyzer. Keep the process gas above its dew point all the way to the sensor, since the humidity limit is below 95% RH non-condensing.
For the quarterly calibration, choose primary standard gases that bracket the normal operating concentration, and log each result so that drift trends can be seen. In hot plant areas in Oman and the GCC, place the analyzer inside a shaded or air-conditioned analyzer cabinet so that the electronics and the sample stay within their temperature limits during summer.
Frequently asked questions
What models make up the 700B series? Models 710B, 720B, 730B and 740B with hydrogen ranges from 0.1% up to 100%.
What outputs does the HY-OPTIMA 700B provide? A 4-20 mA analog output, RS232 or RS422 serial, and a 1 A / 30 VDC SPDT relay.
What supply does it need? 10 to 26 VDC at 10 W input power.
Is it explosion proof? No. The 700B is the general purpose version; the datasheet lists UL and CE certifications, not hazardous-area ratings.
How big is the unit? 236.22 x 86.36 x 35.56 mm and 0.4 kg, with a 1/2 in MNPT adapter fitting.
Documents and quotation
The H2scan datasheet is attached. Seven Star LLC is an independent supplier based in Muscat. To request a quotation, send the model (710B to 740B), expected hydrogen range, background gas and signal requirements.
Datasheets & Documents
- Datasheet H2scan HY-OPTIMA 700B Series datasheet (document 90000145) Download
Reviews (0)
How would you rate this product? Cancel reply
About brand
H2scan is an American maker of solid-state hydrogen sensors and analysers, founded in 2002 and based in Valencia, California. Its first product commercialised a wide-range hydrogen sensor developed at Sandia National Laboratories. H2scan's monitors measure hydrogen directly in process streams, battery rooms and power transformers.
Seven Star LLC lists H2scan hydrogen process analyzers, area monitors and transformer monitors for buyers in Oman and the GCC, and we answer each inquiry against the manufacturer's own datasheet.
H2scan company profile
| Founded | 2002, California |
|---|---|
| Headquarters | Valencia, California |
| Technology | Solid-state hydrogen sensing, licensed from Sandia National Laboratories |
| Product families | HY-OPTIMA in-line hydrogen process analysers (700B, 2700, 5232, 5330), HY-ALERTA fixed area hydrogen monitors (600B, 2620), GEN5 transformer monitors |
H2scan product families on this page
The listed items follow the manufacturer's grouping:
- HY-OPTIMA: in-line hydrogen process analyzers (2700, 700B, 5232, 5330)
- HY-ALERTA and HY-GUARD: fixed hydrogen area monitors for plant buildings and battery rooms
- GRIDSCAN and Sentinel PRO: dissolved hydrogen and power quality monitoring for transformers
Where H2scan equipment is used
H2scan names refineries, petrochemical plants, industrial gas supply, hydrogen production, electrolyzers and fuel cells, battery rooms, energy storage, data centers and utility transformers among its markets. In Gulf installations these map to hydrotreater and reformer streams, substation battery rooms, hydrogen-cooled generators and transformer fleets.
H2scan products we list
Ordering H2scan equipment: what we need
Send the model number, hydrogen range, background gas, hazardous-area requirement, supply voltage and output type. For transformer monitors add the oil type and the protocol, Modbus or DNP3.
Facts checked against Sandia National Laboratories news releases and H2scan announcements, October 2026.
Related products
Frequently Asked Questions
Models 710B, 720B, 730B and 740B with hydrogen ranges from 0.1% up to 100%.
A 4-20 mA analog output, RS232 or RS422 serial, and a 1 A / 30 VDC SPDT relay.
10 to 26 VDC at 10 W input power.
No. The 700B is the general purpose version; the datasheet lists UL and CE certifications, not hazardous-area ratings.
236.22 x 86.36 x 35.56 mm and 0.4 kg, with a 1/2 in MNPT adapter fitting.