Michell MDM300-IS Intrinsically Safe Portable Hygrometer

Description

Michell MDM300-IS intrinsically safe portable hygrometer

The Michell MDM300-IS is the hazardous-area version of the MDM300 portable dew-point hygrometer, designed for engineers who must take moisture readings inside classified zones on gas plants, refineries and offshore platforms. Its coding is II 1G Ex ia IIC T4 Ga under ATEX and UKCA, Ex ia IIC T4 Ga under IECEx and 0Ex ia IIC T4 Ga under TR CU, all for -20 to +50 °C, and cQPSus covers Class I Division 1 Groups A, B, C, D plus Class I Zone 0 AEx/Ex ia IIC Ga T4. A related MDM300-IS-INMETRO variant adds Brazilian certification.

Compared with a general-purpose hygrometer, the unit trades some speed for safety. Michell quotes a T95 of 30 minutes or less to -60 °C dew point, and the blue-case instrument uses a PTFE seal, borosilicate glass and ceramic in the gas path alongside 316L stainless steel. The sensor is the same ceramic metal-oxide tile used across the MDM300 series, with a thin active layer under one micron. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.

Why choose MDM300-IS for zoned areas

Using a non-certified meter at a Zone 0 or Division 1 sample point is not an option, and carrying a gas-tight wet-lab alternative is slow. MDM300-IS is built so a maintenance team can sample natural gas pipelines, dehydration outlets, petrochemical lines and offshore gas with an instrument certified for the zone. The hazardous-area charger is not certified, so charge in the safe area and carry the instrument out.

MDM300-IS specification table

Part number MDM300-IS
ATEX / UKCA marking II 1G Ex ia IIC T4 Ga (-20 to +50 °C)
IECEx marking Ex ia IIC T4 Ga (-20 to +50 °C)
TR CU marking 0Ex ia IIC T4 Ga
North America Class I Div 1 Groups A B C D T4; Class I Zone 0 AEx/Ex ia IIC Ga T4
Accuracy, dew point ±1 °C from -60 to +20 °C; ±2 °C from -100 to -60 °C
Response T95 in 30 minutes or less to -60 °C dew point
Units °C, °F, K; ppmV, g/m3, lb/mmscf for natural gas; ppmW and g/kg for air, N2, H2, CO2, SF6; % rh
Flow across sensor 0.2 to 0.5 Nl/min
Wetted materials 316L stainless steel, PTFE seal, borosilicate glass, ceramic
Battery life up to 24 hours typical (NiMH 4.8 V)
Maximum pressure 350 barg (5076 psig)
Enclosure and weight IP66 / NEMA 4; 1.5 kg
Dimensions 218 x 170 x 90 mm (d x w x h)

How MDM300-IS differs from MDM300-STD and its variants

Against the standard model, MDM300-IS halves the battery life to about 24 hours, doubles the dry-down time to 30 minutes, narrows the flow window to 0.2 to 0.5 Nl/min and adds 150 g of weight. In exchange it gains hydrogen as a supported gas for ppmW and g/kg conversion, and natural-gas units of lb/mmscf, g/m3 and ppmV. The INMETRO version is the same instrument with an added Brazilian certificate, while the hazardous-area carry case code is MDM300-IS-CASE and the bag is MDM300-IS-BAG.

Applications of MDM300-IS in Oman and GCC energy facilities

In Oman’s gas gathering and treatment plants, MDM300-IS suits spot checks at dehydration units, fuel gas headers and pipeline pigging stations. Refinery and petrochemical teams can use it on hydrogen, nitrogen and CO2 supply lines, while power utilities use it for SF6 and quench gas. Because it holds 8 MB of logs, a round of readings across an offshore deck can be downloaded later via Bluetooth at about 5 m.

Batteries, sensors and spares for MDM300-IS

Two battery codes exist: MDM300-BAT-IS for units made before serial number 165024 (around December 2019) and MDM300-BNW-IS for serial numbers from 165024 onward. Always match the pack to the serial number. The calibrated sensor assembly for this model is MDM300-SEN-IS, with a service-exchange option MDM300-SEN-X-IS and a recalibration service MDM300-IS-Recal covering 13 points between -100 and +20 °C dew point.

Using and maintaining MDM300-IS safely

Confirm the zone and gas group against the markings on the instrument and the certificates before entry. Keep the gas flow in the narrow 0.2 to 0.5 Nl/min window, because a higher flow degrades dew-point accuracy on this model. After use, dry the path, check the PTFE seal and orifice, and store the unit with its case. The user is responsible for ensuring the whole installation meets local standards for explosive atmospheres.

Sampling hardware that pairs with MDM300-IS

Sample conditioning decides whether the reading is true. The ceramic-sensor handheld performs best with a short, clean, dry stainless or PTFE line, a particulate filter upstream and a metering valve that holds flow inside the narrow 0.2 to 0.5 Nl/min window. The manufacturer’s accessory list offers fitting packs, flexible tubing assemblies of 0.5 m and 1.8 m, and a pump assembly, and it points to application kits that provide an out-of-the-box sampling system for popular duties. For pressurised lines, reduce to a safe pressure with a regulator rated for the gas before the orifice, and vent the outlet to a safe location.

FAQ summary for MDM300-IS

Is the charger certified? No. How fast is it? Thirty minutes to reach -60 °C T95. Which gases? Air, nitrogen, hydrogen, CO2 and SF6 in the datasheet’s conversions, plus natural gas units. Which pressure? Up to 350 barg.

Documents and quotation

The manufacturer’s MDM300 datasheet in PDF form is available with this listing. For a quotation on MDM300-IS, send an RFQ stating the part number, quantity, area classification (ATEX, IECEx or North American), and whether you need the hazardous-area case MDM300-IS-CASE, spare battery or sensor assemblies.

Datasheets & Documents
  • PDF Datasheet Michell MDM300-IS Intrinsically Safe Portable Hygrometer Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Michell Instruments is a British manufacturer of humidity and dew point instruments, founded in 1974 in Cambridge, England, by Andrew Michell and incorporated on 12 September 1974. It worked with Cambridge University scientists on an aluminium oxide dew point sensor and later developed chilled mirror hygrometers and hydrocarbon dew point analysers. Michell is based in Ely, Cambridgeshire, and is part of Process Sensing Technologies (PST).

Seven Star LLC lists Michell dew-point transmitters, the Condumax II analyzer and its replacement sensor parts, and supplies them to gas plants and compressed air users in Oman by exact order code.

MICHELL Instruments company profile

Founded 1974, Cambridge, England, by Andrew Michell
Headquarters Ely, Cambridgeshire, United Kingdom, since 2007
Ownership Process Sensing Technologies (PST)
Technology Ceramic and aluminium oxide moisture sensors, chilled mirror dew point, Condumax hydrocarbon dew point (licensed from Shell)
Product families Easidew and Promet dew point transmitters, S8000 and Optidew chilled mirror hygrometers, Condumax II hydrocarbon dew point analysers, MDM300 portable dew point meters, oxygen analysers

Milestones

  • 1974: Andrew Michell founds Michell Instruments in Cambridge.
  • 2007: The company moves to Ely, Cambridgeshire.
  • 2024: Michell marks 50 years, now as part of PST.

Michell product families on this page

The Condumax II is an automatic hydrocarbon dew-point analyzer for natural gas. Several of its parts are listed separately: the CD2-SCR-002 hydrocarbon dew-point sensor cell, the CD2-WDC-002 water dew-point sensor assembly and the CD2-SX-WDC-002, which is the exchange version of that water dew-point sensor. The Easidew line is represented by the Easidew EA2 dew point transmitter and by the Easidew PRO I.S., an intrinsically safe moisture transmitter for hazardous areas. A ceramic impedance moisture sensor module is also listed, the sensing element behind these transmitters.

Where Michell analyzers are used

Hydrocarbon and water dew point are contract quality parameters for natural gas, so the Condumax II is found at gas processing plants, pipeline entry points and custody transfer stations. Easidew transmitters monitor dryers for instrument air and industrial gases, glove boxes, and dry gas feeds to processes that cannot tolerate water.

MICHELL Instruments products we list

Family Examples
Gas Analyzer Systems Easidew PRO I.S. Intrinsically Safe Mois…, Michell CD2-WDC-002 Water Dew-Point Sens…, Michell Condumax II Hydrocarbon Dew-Poin…, Easidew EA2 Dew Point Transmitter, Michell OptiPEAK TDL600 Moisture in Natu…, Michell Promet EExd Process Moisture Ana…, Michell S8000 Precision Chilled Mirror H…
Gas Analyzer Accessories MICHELL CD2-SX-WDC-002 Water Dew-Point S…, Michell Ceramic Impedance Moisture Senso…, Michell Condumax II CD2-SCR-002 Hydrocar…

Ordering Michell parts: what we need

For Condumax II parts send the analyzer serial number and the part code, such as CD2-SCR-002, and state if a new assembly or the exchange item is wanted. For Easidew transmitters we need the measuring span in dew point or ppm, the process connection thread, the output signal, the operating pressure, the gas and the area certification. A photo of the label on the old transmitter usually answers most of these in one go.

Facts checked against UK Companies House records and PST's Michell 50-year material, October 2026.

FAQ

Questions about the Michell MDM300-IS Intrinsically Safe Portable

Answers from our engineering team on installation, compatibility and integration.

01Which hazardous-area approvals does MDM300-IS hold?
The datasheet lists ATEX/UKCA II 1G Ex ia IIC T4 Ga, IECEx Ex ia IIC T4 Ga, TR CU 0Ex ia IIC T4 Ga and cQPSus Class I Division 1 and Zone 0 coding.
02How long to reach -60 °C dew point with MDM300-IS?
Michell quotes a T95 of 30 minutes or less to -60 °C (-76 °F) dew point at atmospheric pressure, slower than the 15 minutes of MDM300-STD.
03Can the MDM300-IS charger be used in the hazardous area?
No. The intelligent charger supplied with the instrument is not certified for hazardous-area use, so charge the battery in a safe area.
04Which battery pack fits MDM300-IS?
Units with serial numbers before 165024 use MDM300-BAT-IS, while serial number 165024 onward use MDM300-BNW-IS. Check the serial label before ordering.
05What flow should be set on MDM300-IS?
The flow across the sensor should be 0.2 to 0.5 Nl/min. The wetted parts are 316L stainless steel, PTFE seal, borosilicate glass and ceramic.