Brand:

Magnaflux Ultragel II Ultrasonic Couplant

Description

A general-purpose ultrasonic couplant gel for flaw detection and thickness gauging

Magnaflux Ultragel II is a medium-viscosity ultrasonic couplant from the Sonotech range of Magnaflux. The manufacturer describes it as a high-performance couplant for flaw detection and sizing, thickness gauging, flow metering and acoustic emission testing. It is a bright blue, transparent, thixotropic gel. It stays where it is put on vertical and overhead surfaces, then thins under the movement of the probe so that the transducer slides easily during a scan.

Ultrasound at the frequencies used in NDT does not cross even a thin layer of air. A couplant fills the gap between the transducer face and the test surface so that sound can pass into the part and the echo can return. Ultragel II does this with an acoustic impedance higher than water. The data sheet says this reduces surface noise, and the gel also fills small depressions on rough or corroded surfaces.

According to the data sheet, Ultragel II is approved by Pratt & Whitney (PMC 4384) and meets nuclear-grade limits for halogen and sulfur content. Both are below 50 ppm.

Features listed by the manufacturer

  • Thixotropic, medium-viscosity gel that clings to vertical and overhead surfaces
  • Fills depressions in rough surfaces and gives good transducer lubrication
  • Higher acoustic impedance to reduce surface noise
  • Slow drying, for longer inspections or long-term coupling
  • High corrosion protection for the inspected part
  • Good surface wetting on most metals and composites
  • Nuclear grade: halogens and sulfur each below 50 ppm
  • Suitable for hydrogen embrittlement testing (ASTM F519)
  • Silicone-free and water soluble, so it rinses off with water

Technical specification

Parameter Specification
Product type Ultrasonic couplant, general purpose gel
NDT method Ultrasonic testing (UT)
Appearance Transparent gel, bright blue
Comparative viscosity 5 on the manufacturer’s 0 to 10 scale (medium gel)
Viscosity (comparison guide) About 80,000 cps, Helipath spindle E at 1.5 rpm
Composition notes Contains glycerin and propylene glycol; no silicone
Halogens Below 50 ppm
Sulfur Below 50 ppm
Water soluble Yes
Usage temperature -10 to 210 °F / -23 to 99 °C
Storage temperature 50 to 86 °F / 10 to 30 °C, do not freeze
Compatibility Most composites and metals; may darken or discolour magnesium
Specifications listed API, ASME, AWS, ASTM F519, ASTM F945 or PWA 36604, MCL E-205 Type II, Pratt & Whitney PMC 4384
Data sheet revision February 2024

The data sheet notes that couplant integrity and acoustic performance may decline outside the usage temperature limits. The Magnaflux comparison guide gives the temperature range for thickness gauging and lower maximum temperatures for flaw inspection.

Pack sizes and part numbers

Pack Part number Typical use
12 fl oz / 354 mL bottle, case of 12 25-912 Field kits and thickness gauge cases
1 gal / 3.78 L cubitainer 25-901 Inspection crews and workshop benches
5 gal / 18.9 L cubitainer 25-905 Shutdown campaigns and refilling bottles
55 gal / 208 L drum 25-955 Automated scanning and large users

The data sheet lists the 1 gal and 5 gal cubitainers for shipment to Europe, the Middle East and Africa. The bottles and the drum are listed for North America, Asia and Oceania, so these packs need a separate check of availability before a quotation for Oman.

Where Ultragel II is used

The manufacturer lists flaw detection, flaw sizing, thickness gauging, flow metering, acoustic emission testing, weld inspection, rough surfaces, vertical and overhead surfaces, aerospace and nuclear inspections, composites, turbine blades and aircraft wheel maintenance. In oil, gas, refining and power plants in Oman and the GCC, this means corrosion and wall-thickness surveys on piping and vessels, pulse-echo and phased array weld inspection, and the clamp-on transducers of ultrasonic flowmeters. Because the gel clings to the surface, it suits work on pipe racks and the shells of tanks, where a thin liquid would run off before the reading is taken.

For long-term coupling, such as a clamp-on flowmeter left in place for a survey, slow drying is the property that matters. The data sheet names slow drying as a benefit for extended inspection time.

Choosing a couplant: Ultragel II compared with other general-purpose gels

The Magnaflux comparison guide puts Ultragel II at the top of its general-purpose group for corrosion inhibition and acoustic performance. The table below uses the guide’s thickness-gauging temperature ranges.

Couplant Temperature range Notes from the guide
Ultragel II -23 to 99 °C Highest corrosion inhibition and acoustic score in the group; ASTM F519, ASTM F945/PWA 36604, PWA 36700, BAC 5968, BAC 5980, BAC 5439 listed
Soundsafe -18 to 93 °C Similar viscosity, PWA 36700
Sonotrace Gr 30 -4 to 79 °C BAC 5968
Echogel Gr 20 -3 to 60 °C ASTM F519, BAC 5439

For hot surfaces above 99 °C, such as in-service piping, a high-temperature couplant is needed instead. Magnaflux lists separate products for that service. For normal-incidence shear wave probes, a very viscous shear-wave couplant is required, because ordinary gels do not carry shear waves.

Application and removal

The data sheet instruction is short: apply a small amount of couplant to the transducer or the inspection area before the measurement. Remove it afterwards with a water rinse (warm or hot water is recommended), isopropyl alcohol or 100% ethyl alcohol. General practice, not from the sheet: wipe the surface clean of loose scale before applying the gel, use the same couplant for calibration on the reference block as on the part, and clean the area before painting or welding, since residue can affect coating adhesion.

Storage and handling

Store the couplant in its original container, out of direct sunlight, with the container closed when not in use. Do not freeze it. The data sheet also says never to pour unused couplant back into the original container. Where pumps or valves dispense bulk couplant, wash them between drums so that new product is not contaminated. In the GCC summer, keep stock in an air-conditioned store, because the maximum storage temperature is 30 °C.

Material certification

For austenitic stainless steel, nickel alloys and titanium, inspection codes such as ASME Section V limit the halogen and sulfur content of couplants and ask for a record of the batch. Magnaflux publishes a batch certificate for Ultragel II, and the data sheet gives both contaminants as below 50 ppm. Ask for the batch certificate with the order when your procedure requires it, and keep it in the inspection file with the batch number printed on the container.

What to include in a quotation request

Send the part number from the pack table, the quantity, the delivery location in Oman or elsewhere in the GCC, and whether a batch certificate is required. If the couplant will be used on hot surfaces or with shear-wave probes, state the surface temperature and probe type so that we can confirm Ultragel II is the right product or propose another couplant.

Related products

Ultragel II is used with ultrasonic thickness gauges, flaw detectors and calibration blocks. Browse the NDT Accessories range, or the wider Non-Destructive Testing (NDT) category, for related equipment.

Documents and quotation

The manufacturer data sheet (revised February 2024) is attached for download. The values on this page are taken from it and from the Magnaflux couplant comparison guide; confirm the pack size and any specification your project requires before ordering. Seven Star LLC is an independent supplier in Muscat; send the part number and quantity for a quotation.

Datasheets & Documents
  • PDF Datasheet Magnaflux Ultragel II Product Data Sheet (manufacturer PDF, revised February 2024) Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Magnaflux is a non-destructive testing company based in Glenview, Illinois, USA, whose liquid penetrant and magnetic particle products are used to find surface and near-surface flaws in welds, castings, forgings and machined parts.

Seven Star LLC is an independent supplier in Muscat. We list Magnaflux consumables and inspection tools for NDT companies, fabrication yards and maintenance workshops in Oman and the GCC.

Magnaflux product series

Penetrant products include the Spotcheck system of SKC-S cleaner and remover, SKL-SP2 visible red dye penetrant and SKD-S2 solvent-based developer, the ZL-60D level 2 and ZL-67 level 3 Zyglo water-washable fluorescent penetrants and the ZP-9F developer aerosol.

Magnetic particle products include the 7HF oil-based visible particle aerosol, the 14AM fluorescent particle suspension and the 20B fluorescent premix, applied with an electromagnetic yoke such as the Y-7 AC/DC yoke.

Where Magnaflux products are used

Weld inspection on pipelines, vessels and structures, shutdown inspections in refineries and plants, incoming inspection of castings and forgings, and aerospace and automotive part inspection to ASTM, ISO and aerospace specifications.

Ordering Magnaflux products: what we need

Send the product name and package (aerosol, litre or drum), the quantity and the inspection standard or specification you work to. For yokes give the supply voltage and whether AC, DC or both are required.

Frequently Asked Questions

Is Ultragel II water soluble?
Yes. Ultragel II is water soluble and contains no silicone. The data sheet says to remove it with a water rinse, preferably warm or hot water, or with isopropyl or 100% ethyl alcohol.
What temperature range does Ultragel II cover?
The data sheet gives a usage range of -10 to 210 F (-23 to 99 C). Couplant integrity and acoustic performance may decline beyond these limits, so hot in-service surfaces need a high-temperature couplant.
Can Ultragel II be used on stainless steel and nickel alloys?
Yes. The data sheet lists halogens and sulfur each below 50 ppm and describes the gel as nuclear grade. A batch certificate is published, which procedures such as ASME Section V commonly require for these materials.
Which specifications does Ultragel II meet?
The data sheet lists API, ASME, AWS, ASTM F519, ASTM F945 or PWA 36604, MCL E-205 Type II and Pratt & Whitney PMC 4384.
What pack sizes are available?
The data sheet lists a 12 fl oz bottle case of 12 (25-912), a 1 gal cubitainer (25-901), a 5 gal cubitainer (25-905) and a 55 gal drum (25-955). The 1 gal and 5 gal packs are listed for the Middle East.