Michell EPD Intrinsically Safe Field Display

Description

Michell EPD intrinsically safe field display for Easidew I.S. transmitters

The Michell EPD is a loop-powered four-digit local indicator for the Easidew PRO I.S. and Easidew I.S. dew-point transmitters. It sits in the hazardous area beside the sensor, takes its power from the 4 to 20 mA loop and shows moisture in ppmV, ppmW, °C dp or °F dp, whichever unit label is specified for the associated transmitter. The digits are 34 mm high, the enclosure is glass-reinforced plastic rated IP66, and the certification covers ATEX Group II Category 1GD (Ex ia IIC T5 Ga), IECEx, UKCA and FM. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.

Display and loop behaviour of EPD

The display is a non-multiplexed LCD with four digits, and the dew-point resolution is 0.1 °C or °F. For concentration units the resolution steps with the value: 1 for 1000 to 9999, 0.1 for 100.0 to 999.9, 0.01 for 10.00 to 99.99 and 0.001 for 0.000 to 9.999. Zero and span are adjustable between 0 and ±9999 with 4 mA and 20 mA input respectively. Over-range or under-range shows 9999 or -9999 with all decimal points flashing. The voltage drop is below 1.2 V at 20 °C and below 1.3 V at -40 °C, rising to below 5 V with the loop-powered green backlight.

EPD technical data

Part number EPD
Input 4 to 20 mA loop powered
Display LCD, 4 digits, 34 mm high
ATEX Group II Category 1GD, Ex ia IIC T5 Ga, Ta -40 to +70 °C
IECEx Ex ia IIC T5 Ga, Tamb -40 to +70 °C
FM Class I, II, III Division 1 Groups A to G, T4 at 70 °C
Operating temperature -40 to +70 °C (FM version -20 to +60 °C)
Storage temperature -40 to +85 °C
Humidity up to 95 percent at 40 °C, non-condensing
Enclosure IP66 GRP, three M20 x 1.5 entries
Terminals screw clamp for 0.5 to 1.5 mm2
Weight 1.7 kg
Alarms (option) two isolated solid-state outputs, Ron 5 Ω plus 0.7 V maximum, Roff 1 MΩ minimum

Building the EPD order code

The product is configured by features. Feature A is the base model EPD; Feature B is the range, either B1 (-100 to +20 °C dew-point input) or BX with the user-defined zero and full scale in °C, °F or ppmV (for example BX(0/100P) for 0 to 100 ppmV) which must match the transmitter; Feature C is the backlight (C1 none, C2 backlit); Feature D is the tag plate (D1 none, D2 stainless steel up to 12 characters); Feature E is alarms (E1 none, E2 dual solid-state); Feature F is the keypad (F1 none, F2 external); and Feature G is certification (G1 ATEX gas and IECEx, G2 ATEX gas and dust with IECEx, G3 adds FM and cFM with 22.25 mm plain cable entries). The datasheet’s own example is EPD + BX(0/100P) + C1 + D2 + E1 + F1 + G1.

How EPD differs from other Michell display solutions

EPD is not a transmitter and cannot measure by itself; it only reads the loop current of a connected Easidew PRO I.S. or Easidew I.S. unit. The existing Easidew PRO XP display kit is a different item for the explosion-proof family. Within the EPD range, differences are limited to the option letters, such as the FM version, which has cable entries changed from M20 to plain 22.25 mm and a narrower temperature range of -20 to +60 °C.

Installation and wiring of EPD

The display is wired in series in the loop. In the hazardous area it sits between the Easidew PRO I.S. transmitter and the I.S. galvanic isolator (available with the transmitter) located in the safe area, through which the 4 to 20 mA circuit continues. Mount on a surface through two M6 clearance holes or on an 80 mm outside-diameter pipe with the EPD-PMK stainless bracket kit. The EPD-CAB kit supplies two cable glands and 0.5 m of BS5308 crimped cable.

Where EPD is used

The unit is placed beside moisture transmitters on natural gas dehydration, LNG, refinery and petrochemical sample lines, so that an operator sees dew point locally without a trip to the control room. It can also be mounted inside a sampling system enclosure.

Safety limits and what EPD does not do

EPD is a certified display, not a barrier. The datasheet’s connection diagram shows an I.S. galvanic isolator in the safe area between the hazardous-area loop and the control system, and the isolator can be purchased with the Easidew PRO I.S. transmitter. The alarm outputs are isolated solid-state switches that comply with the requirements for simple apparatus, so they can switch a safe-area circuit only through suitable certified interfaces, and they are fully independent of one another. The installation as a whole remains the user’s responsibility and must follow local rules for explosive atmospheres.

Display ergonomics and tag information

The 34 mm digits are legible from several metres, the optional green backlight is powered from the loop itself with no extra wiring, and a stainless legend plate of up to 12 characters carries the tag or application. The optional external membrane keypad lets the span and zero be configured without opening the lid, which matters in a hazardous area. A terminal compartment inside the case, with screw clamps for 0.5 to 1.5 mm2 conductors, and three M20 x 1.5 cable entries make field wiring straightforward, and the 1.7 kg unit is small enough to hang on a sample-system enclosure wall.

Documents and quotation

The intrinsically safe field display datasheet is available as a PDF with this listing. To quote EPD, send an RFQ with the full order string (range, backlight, tag plate, alarms, keypad, certification), quantity and any EPD-PMK or EPD-CAB accessories.

Spares and Lifecycle Considerations

For critical service, many plants hold one MICHELL Instruments EPD as an operational spare rather than waiting for a failure. When you plan spares, look at how many identical humidity & dew point instruments units are installed, how long a replacement would take to arrive, and the cost of an unplanned stop. Store spare units in their original packaging in a dry, temperature-controlled area and record the part number and serial number in your maintenance system. We can quote single units or a small spares package and confirm the lead time for each item in writing.

Datasheets & Documents
  • PDF Datasheet Michell EPD Intrinsically Safe Field Display Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Michell Instruments is a British manufacturer of humidity and dew point instruments, founded in 1974 in Cambridge, England, by Andrew Michell and incorporated on 12 September 1974. It worked with Cambridge University scientists on an aluminium oxide dew point sensor and later developed chilled mirror hygrometers and hydrocarbon dew point analysers. Michell is based in Ely, Cambridgeshire, and is part of Process Sensing Technologies (PST).

Seven Star LLC lists Michell dew-point transmitters, the Condumax II analyzer and its replacement sensor parts, and supplies them to gas plants and compressed air users in Oman by exact order code.

MICHELL Instruments company profile

Founded 1974, Cambridge, England, by Andrew Michell
Headquarters Ely, Cambridgeshire, United Kingdom, since 2007
Ownership Process Sensing Technologies (PST)
Technology Ceramic and aluminium oxide moisture sensors, chilled mirror dew point, Condumax hydrocarbon dew point (licensed from Shell)
Product families Easidew and Promet dew point transmitters, S8000 and Optidew chilled mirror hygrometers, Condumax II hydrocarbon dew point analysers, MDM300 portable dew point meters, oxygen analysers

Milestones

  • 1974: Andrew Michell founds Michell Instruments in Cambridge.
  • 2007: The company moves to Ely, Cambridgeshire.
  • 2024: Michell marks 50 years, now as part of PST.

Michell product families on this page

The Condumax II is an automatic hydrocarbon dew-point analyzer for natural gas. Several of its parts are listed separately: the CD2-SCR-002 hydrocarbon dew-point sensor cell, the CD2-WDC-002 water dew-point sensor assembly and the CD2-SX-WDC-002, which is the exchange version of that water dew-point sensor. The Easidew line is represented by the Easidew EA2 dew point transmitter and by the Easidew PRO I.S., an intrinsically safe moisture transmitter for hazardous areas. A ceramic impedance moisture sensor module is also listed, the sensing element behind these transmitters.

Where Michell analyzers are used

Hydrocarbon and water dew point are contract quality parameters for natural gas, so the Condumax II is found at gas processing plants, pipeline entry points and custody transfer stations. Easidew transmitters monitor dryers for instrument air and industrial gases, glove boxes, and dry gas feeds to processes that cannot tolerate water.

MICHELL Instruments products we list

Family Examples
Gas Analyzer Systems Easidew PRO I.S. Intrinsically Safe Mois…, Michell CD2-WDC-002 Water Dew-Point Sens…, Michell Condumax II Hydrocarbon Dew-Poin…, Easidew EA2 Dew Point Transmitter, Michell OptiPEAK TDL600 Moisture in Natu…, Michell Promet EExd Process Moisture Ana…, Michell S8000 Precision Chilled Mirror H…
Gas Analyzer Accessories MICHELL CD2-SX-WDC-002 Water Dew-Point S…, Michell Ceramic Impedance Moisture Senso…, Michell Condumax II CD2-SCR-002 Hydrocar…

Ordering Michell parts: what we need

For Condumax II parts send the analyzer serial number and the part code, such as CD2-SCR-002, and state if a new assembly or the exchange item is wanted. For Easidew transmitters we need the measuring span in dew point or ppm, the process connection thread, the output signal, the operating pressure, the gas and the area certification. A photo of the label on the old transmitter usually answers most of these in one go.

Facts checked against UK Companies House records and PST's Michell 50-year material, October 2026.

FAQ

Questions about the Michell EPD Intrinsically Safe Field Display

Answers from our engineering team on installation, compatibility and integration.

01Which transmitters is EPD designed for?
EPD is made for the Easidew PRO I.S. and Easidew I.S. dew-point transmitters. It reads their 4 to 20 mA loop and displays the value in ppmV, ppmW, °C dp or °F dp.
02How is EPD powered?
EPD is loop powered. Its voltage drop is below 1.2 V at 20 °C and below 5 V when the optional green backlight is fitted.
03What certification does EPD carry?
The datasheet lists ATEX Group II Category 1GD Ex ia IIC T5 Ga, IECEx Ex ia IIC T5 Ga, UKCA and FM Class I, II, III Division 1 approvals.
04Does the EPD display range have to match the transmitter?
Yes. The zero, full-scale value and unit in the BX range code must be the same as the associated transmitter, for example BX(0/100P) for 0 to 100 ppmV.
05What mounting accessories exist for EPD?
The datasheet lists EPD-PMK, a stainless-steel pipe-mount kit for 80 mm pipe, and EPD-CAB, a connection kit with two cable glands and 0.5 m of cable.