Servomex S2500980C ZnSe Window for SpectraExact 2500

Description

Servomex S2500980C zinc selenide window explained

Servomex S2500980C is the zinc selenide window, ZnSe, in the Window Spares table of the SpectraExact 2500 series manual. A 2500 cell uses two such windows, each 6 mm thick, to let the infrared or visible beam through a sample held in a 316 stainless steel body. Zinc selenide is not the default; it is one of the special materials the manual says some applications require, alongside barium fluoride. The part number you need is therefore decided by the Manufacturing Data Sheet printed at the front of the manual for your particular analyser.

We describe the part so that planners can see both the benefits and the safety implications. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.

The ZnSe warning every maintainer should read

The manual carries a dedicated warning for zinc selenide. If a ZnSe window has been damaged by acidic attack, for example because acidic condensate or organic acids entered the cell, some of the material is likely to have been reduced to selenium metal, which is described as highly toxic. On no account may attacked ZnSe be handled with bare fingers; it must be bagged and disposed of according to local regulations. That rule is the strongest reason to inspect the old window, the PTFE gasket and the end bosses carefully whenever S2500980C is changed.

Parameter Value
Part number S2500980C
Material Zinc selenide (ZnSe)
Table 6.5 Window Spares
Window thickness 6 mm
Standard alternative S2500980A, calcium fluoride
Cleaning agents Cotton buds or lens tissue with detergent or isopropanol, no abrasives
Handling Cotton gloves or finger cots
Removal aid Hand aspirator 2387-0514
Locking ring tool Cell key in Special Tools Kit S2500979
Analyser power 115/230 V or 100/200 V AC plus or minus 15 percent, 50/60 Hz

How S2500980C differs from sibling windows

  • Against S2500980B (BaF2): ZnSe carries the special acid-attack warning for selenium formation, while BaF2 is flagged simply as toxic.
  • Against S2500980A (CaF2): the standard window is non-toxic and reasonably robust.
  • Against S2500980G (sapphire) and S2500980D (quartz): those are for different wavelength bands and carry no toxicity note in the manual.

Where ZnSe is used and why

ZnSe passes a broad infrared band, which makes it a fit for certain long-wave measurements, where the source units in Table 6.3 include a long-wave ZnSe version, S2500931A. Analysers built with ZnSe windows normally also carry a ZnSe source or matching detector selection, so check the data sheet for the whole optical chain, not only the window.

Safe removal and refitting of S2500980C

Take the whole cell to a clean workshop bench. Dismantle in this order: remove the end bosses, unscrew the locking ring with the cell key, withdraw the five wavy washers, the spacer and the O ring, then lift the window squarely with a hand aspirator while the outlet is sealed. Dispose of any damaged ZnSe as hazardous waste. When refitting, float the new window onto the PTFE gasket squarely, use fresh seals, and tighten the locking ring in stages with pauses. Leak test with a manometer.

Preventing acid attack on the new window

Check the sample conditioning to ensure no acidic condensate reaches the cell. The manual suggests end boss scrubbers or purges to protect the optics, and states that if the analyser is dismantled the scrubbers must not remain exposed to normal atmosphere for more than a few minutes. Flush the end bosses with dry nitrogen before new scrubbers are fitted.

Stocking S2500980C

A window is a long-lead specialty item, so we suggest asking for a quote before a planned shutdown. Include seals kit S2500982A or S2500982B in the same request.

Typical sites

Refineries, petrochemical units and gas processing plants in the Gulf commonly use infrared analysers of this type for hydrocarbon-bearing streams, and the window material must be agreed against the process chemistry at the project stage.

Remember the instrument limits when judging a replacement: performance is stated as an intrinsic error below plus or minus 1 percent fsd, repeatability of plus or minus 0.5 percent fsd and linearity typically below plus or minus 1 percent fsd. A freshly fitted window should restore the reference detector signals towards their initial values; the manual treats signals under 50 percent of initial as the trigger for cell cleaning or window replacement, and aggressive samples can cause general corrosion or pitting of the glass-like surfaces.

Because ZnSe is brittle, support it squarely when floating it into place and never use the cell inlet and outlet stubs as levers. The locking ring has a coated thread to prevent seizure, so no PTFE tape is needed on it, although PTFE tape 1835-3026 appears elsewhere in the sample cell spares list for fittings.

When the old window is out, inspect the cell bore and the end bosses for deposits. Record the part number S2500980C and the date in the analyser log so that the next engineer knows which window material is in place.

In short, S2500980C is a specialist spare that deserves the same care at the purchasing stage as at the fitting stage. Confirm the data sheet, quote S2500980C exactly, and plan disposal of the old window before it arrives.

Documents and quotation

The Servomex SpectraExact 2500 series manual, part number 02500005D, is the document we attach for Servomex S2500980C; its section 6 spares tables are the source of the part number and description. Because zinc selenide is toxic when degraded, we ask that you confirm the window type on your data sheet before ordering. To request a quotation, send S2500980C, the quantity, the analyser serial number and the Final Specification Sheet reference if you have it, and your delivery location in Oman or the GCC. Each RFQ is answered with availability and lead time.

Other Models in the Servomex 2500 Range

Seven Star LLC supplies models from the Servomex 2500 range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Servomex S2500931A Long Wave ZnSe Source Unit for SpectraExact 2500Gas Analyzer SystemsView details
Servomex S2501905C CVDFET Detector Unit for SpectraExact 2500Gas Analyzer SystemsView details
Servomex S2500980B BaF2 Window for SpectraExact 2500Gas Analyzer SystemsView details
Servomex S1200922 CW Boss Scrubber H2O/CO2 for SpectraExact 2500Gas Analyzer SystemsView details
Servomex S2501911A Power Control PCB for SpectraExact 2500Gas Analyzer SystemsView details
Servomex S2500980C ZnSe Window for SpectraExact 2500Gas Analyzer SystemsYou are viewing this model
Datasheets & Documents
  • PDF Datasheet Servomex S2500980C ZnSe Window for SpectraExact 2500 Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Servomex is a British specialist in gas analysis, founded in 1952 by Richard Steel, John Crockett and Ronald Mudie; the name comes from servomechanism. It first made electronic products, then built its first paramagnetic oxygen analyser in 1961 after work with Charles Munday on a paramagnetic sensor. Servomex is based in Crowborough, East Sussex, and is part of Spectris.

Seven Star LLC lists Servomex process, combustion, emissions and portable gas analyzers, plus accessories, and sources them for refineries, petrochemical plants and industrial gas producers in Oman.

SERVOMEX company profile

Founded 1952, United Kingdom
Founders Richard Steel, John Crockett and Ronald Mudie
Headquarters Jarvis Brook, Crowborough, East Sussex
Ownership Spectris plc, since 1999
Key technologies Paramagnetic oxygen sensing (1961), infrared analysis (1969), zirconia oxygen sensors (1981)
Product families MultiExact and SERVOPRO MultiExact analysers, SERVOTOUGH OxyExact and Oxy 1800 oxygen analysers, SERVOTOUGH Laser 3 Plus TDL analysers, SERVOFLEX portable analysers, zirconia combustion analysers

Milestones

  • 1952: Servomex is founded.
  • 1961: The first Servomex paramagnetic oxygen analyser is introduced.
  • 1981: Servomex introduces zirconia oxygen sensors.
  • 1999: Spectris acquires Servomex.

Servomex analyzer families listed

Servomex groups its analyzers under family names that indicate where they can be used. SERVOTOUGH covers analyzers built for hazardous areas and field mounting, SERVOPRO covers units for general purpose areas and analyzer houses, and SERVOFLEX is the portable line. On this page:

  • SERVOTOUGH OxyExact 2200: paramagnetic oxygen analyzer for process control and safety
  • SERVOTOUGH SpectraExact 2500: photometric analyzer for process gas components
  • SERVOTOUGH Laser 3 Plus: compact tunable diode laser analyzer for combustion
  • SERVOPRO 4900 Multigas: multi-component analyzer for emissions monitoring
  • SERVOPRO NanoTrace DF-550E: ultra-trace oxygen analyzer for high purity gases
  • SERVOFLEX Micro i.s. 5100: intrinsically safe portable oxygen analyzer

A flow alarm relay box for the 2700 flue gas analyzer is also listed.

Where Servomex gas analyzers are used

Typical duties are oxygen measurement for inerting and blanketing, vapor recovery and flare header safety, combustion control on fired heaters and reformers, purity control in air separation and electronic gases, and continuous emissions monitoring. In the GCC the main users are refining, petrochemicals, fertilizer, LNG and industrial gas plants.

SERVOMEX products we list

Family Examples
Gas Analyzer Systems SERVOPRO 4900 Multigas Analyzer, SERVOPRO NanoTrace DF-550E Ultra-Trace O…, SERVOTOUGH Laser 3 Plus Combustion TDL G…, SERVOTOUGH OxyExact 2200 Oxygen Analyzer, ServoTough SpectraExact 2500 Gas Analyzer
Portable Test & Measurement Micro i.s. 5100 Portable Oxygen Analyzer
Gas Analyzer Accessories Tough Flue Gas 2700 Flow Alarm Relay Box

Ordering Servomex analyzers and spares: what we need

A Servomex gas analyzer is built to an order code that fixes the measured gas, range, sample wetted materials, outputs and hazardous area approval. For replacements and spare parts, send the model, the serial number and the order code from the data plate. For a new Servomex oxygen analyzer or multigas unit, send the full stream composition, the component and range to be measured, sample pressure and temperature, the area classification, power supply and communication needs. We check the configuration against Servomex documentation before quoting.

Facts checked against Servomex's own 70-year history and company profiles, October 2026.

FAQ

Questions about the Servomex S2500980C ZnSe Window for SpectraExact 2500

Answers from our engineering team on installation, compatibility and integration.

01What is Servomex S2500980C?
S2500980C is the zinc selenide (ZnSe) window in Table 6.5 of the SpectraExact 2500 series manual.
02What is the main hazard with ZnSe windows?
If acidic condensate or organic acids enter the cell, ZnSe can be reduced to selenium metal, which is highly toxic and must never be touched with bare fingers.
03How should damaged ZnSe be disposed of?
The manual says to bag it and dispose of it according to local regulations.
04Which spare cell kit is used when refitting?
Fit new O rings from a cell seals kit, S2500982A (Viton and PTFE) or S2500982B (Chemraz).
05Is S2500980C the standard window?
No, the standard material is calcium fluoride, S2500980A.