Servomex S1806989B Solvent Resistant High-Flow O2 Transducer

Description

What Servomex S1806989B combines

Servomex S1806989B is the last item in the sample option table for the Oxy 1800 solvent resistant / high-flow cell, described in the manual as the solvent resistant high-flow O2 transducer assembly. It is the transducer for the toughest version of the analyser: one that must resist solvent vapour attack and also handle a fast sample stream. Buyers sometimes request it by the shorthand 1806, but the complete number with the B suffix is what must be quoted.

This text helps procurement teams tell the four Oxy 1800 transducers apart. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.

Specification summary for S1806989B

Parameter Value
Part number S1806989B
Cell type Solvent resistant, high-flow
Seal material Chemraz, as used for the solvent tolerant transducer
Response time T90 Less than 5 seconds
Flow rate effect Less than 0.2 percent O2 over 50 to 70 litres per hour
Noise Less than 0.04 percent O2 peak to peak
Repeatability Less than 0.1 percent of reading or 0.05 percent O2
Linearity Less than 0.05 percent O2
Output signals 4-20 mA isolated, 600 ohm maximum; 0-1 Vdc non-isolated
Alarm relays 250 Vac 3 A or 28 Vdc 1 A, non-inductive

The four transducers side by side

Parameter Value
S1800989B Standard materials, base flow window 50 to 250 ml/min
S1802989B Chemraz seals, base flow window
S1804989B Standard materials, 50 to 70 l/h high flow
S1806989B Chemraz seals and 50 to 70 l/h high flow

The table above is the shortest way to see why S1806989B costs more engineering than the others: it unites two sets of requirements that the other three satisfy separately.

Where the combination is needed

A common use is oxygen monitoring in a solvent vapour recovery system, where the stream is fast because it comes straight from a header, and in blanketing systems at petrochemical terminals that handle volatile liquids. The same cell can be justified in packaging or coating lines that use inert gas atmospheres with solvent carryover. Where a plant in the GCC uses an aromatic or ketone solvent, the question to ask is whether the elastomer rating covers that chemical; the manual only states that Chemraz is the solvent tolerant material, so confirm the species with the manufacturer if a particular chemical is critical.

Sample system planning

High flow means the upstream pressure and flow device must be designed for 50 to 70 litres per hour at the cell, with the vent unrestricted at 11.5 to 18 psia. The fine filter must hold back particles above 3 micrometres. Because the sample is aggressive, add the enclosure purge when the sample is toxic or corrosive: air or inert gas at 100 to 200 ml per minute, running before any cover is opened.

Installing and commissioning S1806989B

Work to service manual 01800002B. After exchange, run the analyser for four hours before the sample arrives to avoid condensation, then set zero with at least 99.9 percent nitrogen and span with air at 20.9 percent O2. Verify the voltage and current outputs on the terminal board: at span of the 0 to 25 percent O2 range the manual expects 0.838 V and 17.41 mA, for example, when air at 20.95 percent is applied.

Stocking guidance

Because S1806989B is an uncommon build, hold at least one on the shelf only if the analyser is critical. A reasonable kit also includes S1802966 tubes, the fuse kit S1800987 and an enclosure seals kit S1800986.

Risks and limits of S1806989B

Combining two special properties does not remove the usual limits. The analyser still carries a rating of installation category II with an over-voltage impulse withstand up to 2500 V, a supply of 100 to 240 Vac plus or minus 10 percent at 50 or 60 Hz and not more than 50 VA, and electrical terminations for 0.5 to 2.5 square millimetre conductors. Connections to the analogue output terminals must not exceed 30 V RMS or 60 Vdc to earth, and screened cables with the screen earthed at the gland are required for EMC compliance.

The analyser must be connected to protective earth, with an isolating switch or circuit breaker within easy reach of the operator and marked as the disconnecting device. These points apply to every Oxy 1800, but they matter most when a solvent-laden plant has stricter electrical audits.

What to send with an RFQ for S1806989B

  • Analyser serial number or nameplate photograph.
  • Sample composition and the solvent expected.
  • Flow rate at the cell in litres per hour.
  • Required delivery location in Oman or the wider GCC.

Remember also that older paperwork may call this analyser family xendos 1800 or 1802; the number S1806989B belongs to the newer quick start table, so quote it as written.

In day-to-day purchasing, S1806989B is best confirmed against the analyser build record. Compare S1806989B with S1804989B and S1802989B in the same table, check that the tube kit already fitted matches the solvent resistant option, and ask for the part by its full number, S1806989B, so the supplier does not offer a near neighbour. A transducer that is wrong for the cell body cannot be corrected by recalibration, which is why the suffix letter in S1806989B matters.

Documents and quotation

The Servomex SERVOTOUGH Oxy 1800 quick start manual, part number 01800003B revision 4, is the document we attach for Servomex S1806989B; it carries the spares table, the sample condition table and the performance figures quoted on this page. Because this is the combined variant, please include your analyser serial number so we can confirm the build. To request a quotation, send the part number S1806989B, the quantity, the analyser serial number if you have it, and the delivery location in Oman or elsewhere in the GCC. We reply to each RFQ with availability and lead time.

Other Models in the Servomex Oxy1800 Range

Seven Star LLC supplies models from the Servomex Oxy1800 range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Servomex S1806989B Solvent Resistant High-Flow O2 TransducerGas Analyzer SystemsYou are viewing this model
Servomex S1802989B Solvent Resistant O2 Transducer Oxy 1800Gas Analyzer SystemsView details
Servomex S1800989B Standard O2 Transducer for Oxy 1800Gas Analyzer SystemsView details
Servomex S1804989B High-Flow O2 Transducer for Oxy 1800Gas Analyzer SystemsView details
Servomex S1800912A 4.5 Digit Display PCB for Oxy 1800Gas Analyzer SystemsView details
Servomex S1802966 Hastelloy C-276 Sample Tubes Kit Oxy 1800Gas Analyzer SystemsView details
Datasheets & Documents
  • PDF Datasheet Servomex S1806989B Solvent Resistant High-Flow O2 Transducer Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Servomex is a British specialist in gas analysis, founded in 1952 by Richard Steel, John Crockett and Ronald Mudie; the name comes from servomechanism. It first made electronic products, then built its first paramagnetic oxygen analyser in 1961 after work with Charles Munday on a paramagnetic sensor. Servomex is based in Crowborough, East Sussex, and is part of Spectris.

Seven Star LLC lists Servomex process, combustion, emissions and portable gas analyzers, plus accessories, and sources them for refineries, petrochemical plants and industrial gas producers in Oman.

SERVOMEX company profile

Founded 1952, United Kingdom
Founders Richard Steel, John Crockett and Ronald Mudie
Headquarters Jarvis Brook, Crowborough, East Sussex
Ownership Spectris plc, since 1999
Key technologies Paramagnetic oxygen sensing (1961), infrared analysis (1969), zirconia oxygen sensors (1981)
Product families MultiExact and SERVOPRO MultiExact analysers, SERVOTOUGH OxyExact and Oxy 1800 oxygen analysers, SERVOTOUGH Laser 3 Plus TDL analysers, SERVOFLEX portable analysers, zirconia combustion analysers

Milestones

  • 1952: Servomex is founded.
  • 1961: The first Servomex paramagnetic oxygen analyser is introduced.
  • 1981: Servomex introduces zirconia oxygen sensors.
  • 1999: Spectris acquires Servomex.

Servomex analyzer families listed

Servomex groups its analyzers under family names that indicate where they can be used. SERVOTOUGH covers analyzers built for hazardous areas and field mounting, SERVOPRO covers units for general purpose areas and analyzer houses, and SERVOFLEX is the portable line. On this page:

  • SERVOTOUGH OxyExact 2200: paramagnetic oxygen analyzer for process control and safety
  • SERVOTOUGH SpectraExact 2500: photometric analyzer for process gas components
  • SERVOTOUGH Laser 3 Plus: compact tunable diode laser analyzer for combustion
  • SERVOPRO 4900 Multigas: multi-component analyzer for emissions monitoring
  • SERVOPRO NanoTrace DF-550E: ultra-trace oxygen analyzer for high purity gases
  • SERVOFLEX Micro i.s. 5100: intrinsically safe portable oxygen analyzer

A flow alarm relay box for the 2700 flue gas analyzer is also listed.

Where Servomex gas analyzers are used

Typical duties are oxygen measurement for inerting and blanketing, vapor recovery and flare header safety, combustion control on fired heaters and reformers, purity control in air separation and electronic gases, and continuous emissions monitoring. In the GCC the main users are refining, petrochemicals, fertilizer, LNG and industrial gas plants.

SERVOMEX products we list

Family Examples
Gas Analyzer Systems SERVOPRO 4900 Multigas Analyzer, SERVOPRO NanoTrace DF-550E Ultra-Trace O…, SERVOTOUGH Laser 3 Plus Combustion TDL G…, SERVOTOUGH OxyExact 2200 Oxygen Analyzer, ServoTough SpectraExact 2500 Gas Analyzer
Portable Test & Measurement Micro i.s. 5100 Portable Oxygen Analyzer
Gas Analyzer Accessories Tough Flue Gas 2700 Flow Alarm Relay Box

Ordering Servomex analyzers and spares: what we need

A Servomex gas analyzer is built to an order code that fixes the measured gas, range, sample wetted materials, outputs and hazardous area approval. For replacements and spare parts, send the model, the serial number and the order code from the data plate. For a new Servomex oxygen analyzer or multigas unit, send the full stream composition, the component and range to be measured, sample pressure and temperature, the area classification, power supply and communication needs. We check the configuration against Servomex documentation before quoting.

Facts checked against Servomex's own 70-year history and company profiles, October 2026.

FAQ

Questions about the Servomex S1806989B Solvent Resistant High-Flow O2

Answers from our engineering team on installation, compatibility and integration.

01Which Oxy 1800 variant uses Servomex S1806989B?
S1806989B fits the Oxy 1800 built with the cell that is both solvent resistant and high-flow, the fourth transducer in the solvent resistant / high-flow sample option table.
02How is S1806989B different from S1804989B?
S1804989B is high-flow with standard materials, while S1806989B adds the solvent resistant construction, which uses Chemraz rather than Viton in the cell.
03What response time is quoted for the high-flow family?
The performance table gives a T90 of less than 5 seconds and noise below 0.04 percent O2 for high-flow transducers.
04Is a tube kit needed with S1806989B?
For corrosive samples the manual lists the Hastelloy C-276 inlet and outlet tube kit S1802966 in the same option table.
05Can S1806989B go in a hazardous area analyser?
No, the Oxy 1800 is for non-hazardous locations and non-flammable samples.