Brand:

Servomex S1800912A 4.5 Digit Display PCB for Oxy 1800

Description

Servomex S1800912A: the 4.5 digit readout board

Servomex S1800912A is the complete 4.5 digit display printed circuit board for the SERVOTOUGH Oxy 1800 analyser. It belongs to the common parts table of the xendos 1800 Series spares list, where it sits next to the 3.5 digit board S1800902A. The practical difference is on the front panel: the standard analyser shows 0.0 to 100.0 percent O2, while analysers with the optional 4.5 digit LCD show 0.00 to 100.00 percent O2.

This description is written to help the buyer avoid ordering the wrong digit count. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.

Data from the manual

Parameter Value
Part number S1800912A
Description 4.5 digit display pcb complete, 1 off
Display range 0.00 to 100.00 percent O2
Display resolution 0.01 percent O2
Standard alternative S1800902A, 3.5 digit, 0.0 to 100.0 percent, 0.1 percent resolution
Display type Backlit LCD
Front panel indicators Power, flow, alarm, and range LEDs
Alarm hysteresis 0.2 to 0.3 percent O2
Cover fastening Four M6 bolts, 5 mm hex key
Ingress protection IP66, NEMA 4X

Who needs the 4.5 digit version

Trace-adjacent users who watch small changes around a purity level, for example an oxygen purity of 99.95 percent, benefit from the extra digit, since 0.01 percent steps give operators a clearer view of trends than 0.1 percent steps. The extra digit does not make the measurement more accurate, because the specified intrinsic error is unchanged, but it helps when logging, trending or commissioning.

Difference from S1800902A

  • S1800902A: three and a half digits, one decimal place.
  • S1800912A: four and a half digits, two decimal places.
  • Both are listed as common parts for the Oxy 1800 and neither carries the housekeeping electronics, which are on S1800911.

Replacement procedure

Isolate the power at the external switch or breaker, as required by the electrical safety notes of the manual, then remove the four M6 bolts from the hinged left cover. The display cover opens with the screw at the middle of its right side. Handle the board with static precautions and refit the same ribbon connections. When finished, close the covers fully and restore power; the green Power LED shows supply is present. Check the display reads sensibly with the sample off and test the zero with nitrogen.

After replacement

Check the range LEDs and the alarm set points. The alarm set point range equals the full scale of the display, 0 to 100 percent O2, and is independent of the analogue output scaling. Repeat zero and span if the board change was combined with a transducer service.

Related parts

Common parts in the same table include the housekeeping board S1800911, the terminals board S1800913C, the switch mode power supply module 2822-2028 and the anti-static glass window 3950-6087.

Alarm and range behaviour that the display board shows

On an analyser using S1800912A the digital display shows concentration in 0.01 steps, while the yellow range LEDs show which of the two analogue output ranges is active. Pressing the Alarm Set button lights one alarm LED and shows that alarm setpoint on the display; pressing it again shows the second. Both alarms can be set from 0 to 100 percent O2 and the display LED flashes if an alarm is active, clearing automatically when oxygen returns to normal. The extra decimal place makes it easier to set alarms close to a nominal value, for example near 20.90 percent in an air-quality monitoring setup.

Pre-order checklist for S1800912A

  • Count the decimals on the current display: one decimal means S1800902A, two means S1800912A.
  • Check whether the analyser serial label shows the optional 4.5 digit build.
  • Decide whether the terminals board S1800913C or the housekeeping board S1800911 also needs attention.
  • Record whether the front window 3950-6087 is scratched, since it can be replaced in the same visit.

Installation steps for S1800912A

  • Switch off and isolate the supply with the external disconnect.
  • Remove the four M6 bolts on the left hand cover using a 5 mm key.
  • Open the hinged display cover with the screw at the middle of the right hand side.
  • Swap the board, reconnect carefully and close both covers.
  • Restore power, confirm the Power LED is green and check the readout against a known gas.

The manual advises that the electrical installation must be safe for the extremes of conditions in the operating environment, so also check that the housing seals from the enclosure seals kit S1800986 are in good condition before closing.

A final purchasing tip: S1800912A is a complete board, so it is quoted as one item with no separate parts. Because S1800902A and S1800912A look nearly identical, a photograph of the front display with its decimal places is the fastest proof of which one you hold. Quote S1800912A together with the analyser serial number when you send the RFQ, and mention if the front window 3950-6087 should be quoted at the same time.

If the analyser is later upgraded or serviced by another team, keep a note that S1800912A is fitted, because the same front window and cover seals serve both display boards and the difference will not be visible from outside the housing.

Documents and quotation

The Servomex SERVOTOUGH Oxy 1800 quick start manual, part number 01800003B revision 4, is the document we attach for Servomex S1800912A; it carries the spares table, the sample condition table and the performance figures quoted on this page. Please confirm whether your unit shows two or three decimals on the display so we can match the 3.5 or 4.5 digit board. To request a quotation, send the part number S1800912A, the quantity, the analyser serial number if you have it, and the delivery location in Oman or elsewhere in the GCC. We reply to each RFQ with availability and lead time.

Other Models in the Servomex Oxy1800 Range

Seven Star LLC supplies models from the Servomex Oxy1800 range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Servomex S1806989B Solvent Resistant High-Flow O2 TransducerGas Analyzer SystemsView details
Servomex S1802989B Solvent Resistant O2 Transducer Oxy 1800Gas Analyzer SystemsView details
Servomex S1800989B Standard O2 Transducer for Oxy 1800Gas Analyzer SystemsView details
Servomex S1804989B High-Flow O2 Transducer for Oxy 1800Gas Analyzer SystemsView details
Servomex S1800912A 4.5 Digit Display PCB for Oxy 1800Gas Analyzer SystemsYou are viewing this model
Servomex S1802966 Hastelloy C-276 Sample Tubes Kit Oxy 1800Gas Analyzer SystemsView details
Datasheets & Documents
  • PDF Datasheet Servomex S1800912A 4.5 Digit Display PCB for Oxy 1800 Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Servomex is a British specialist in gas analysis, founded in 1952 by Richard Steel, John Crockett and Ronald Mudie; the name comes from servomechanism. It first made electronic products, then built its first paramagnetic oxygen analyser in 1961 after work with Charles Munday on a paramagnetic sensor. Servomex is based in Crowborough, East Sussex, and is part of Spectris.

Seven Star LLC lists Servomex process, combustion, emissions and portable gas analyzers, plus accessories, and sources them for refineries, petrochemical plants and industrial gas producers in Oman.

SERVOMEX company profile

Founded 1952, United Kingdom
Founders Richard Steel, John Crockett and Ronald Mudie
Headquarters Jarvis Brook, Crowborough, East Sussex
Ownership Spectris plc, since 1999
Key technologies Paramagnetic oxygen sensing (1961), infrared analysis (1969), zirconia oxygen sensors (1981)
Product families MultiExact and SERVOPRO MultiExact analysers, SERVOTOUGH OxyExact and Oxy 1800 oxygen analysers, SERVOTOUGH Laser 3 Plus TDL analysers, SERVOFLEX portable analysers, zirconia combustion analysers

Milestones

  • 1952: Servomex is founded.
  • 1961: The first Servomex paramagnetic oxygen analyser is introduced.
  • 1981: Servomex introduces zirconia oxygen sensors.
  • 1999: Spectris acquires Servomex.

Servomex analyzer families listed

Servomex groups its analyzers under family names that indicate where they can be used. SERVOTOUGH covers analyzers built for hazardous areas and field mounting, SERVOPRO covers units for general purpose areas and analyzer houses, and SERVOFLEX is the portable line. On this page:

  • SERVOTOUGH OxyExact 2200: paramagnetic oxygen analyzer for process control and safety
  • SERVOTOUGH SpectraExact 2500: photometric analyzer for process gas components
  • SERVOTOUGH Laser 3 Plus: compact tunable diode laser analyzer for combustion
  • SERVOPRO 4900 Multigas: multi-component analyzer for emissions monitoring
  • SERVOPRO NanoTrace DF-550E: ultra-trace oxygen analyzer for high purity gases
  • SERVOFLEX Micro i.s. 5100: intrinsically safe portable oxygen analyzer

A flow alarm relay box for the 2700 flue gas analyzer is also listed.

Where Servomex gas analyzers are used

Typical duties are oxygen measurement for inerting and blanketing, vapor recovery and flare header safety, combustion control on fired heaters and reformers, purity control in air separation and electronic gases, and continuous emissions monitoring. In the GCC the main users are refining, petrochemicals, fertilizer, LNG and industrial gas plants.

SERVOMEX products we list

Family Examples
Gas Analyzer Systems SERVOPRO 4900 Multigas Analyzer, SERVOPRO NanoTrace DF-550E Ultra-Trace O…, SERVOTOUGH Laser 3 Plus Combustion TDL G…, SERVOTOUGH OxyExact 2200 Oxygen Analyzer, ServoTough SpectraExact 2500 Gas Analyzer
Portable Test & Measurement Micro i.s. 5100 Portable Oxygen Analyzer
Gas Analyzer Accessories Tough Flue Gas 2700 Flow Alarm Relay Box

Ordering Servomex analyzers and spares: what we need

A Servomex gas analyzer is built to an order code that fixes the measured gas, range, sample wetted materials, outputs and hazardous area approval. For replacements and spare parts, send the model, the serial number and the order code from the data plate. For a new Servomex oxygen analyzer or multigas unit, send the full stream composition, the component and range to be measured, sample pressure and temperature, the area classification, power supply and communication needs. We check the configuration against Servomex documentation before quoting.

Facts checked against Servomex's own 70-year history and company profiles, October 2026.

FAQ

Questions about the Servomex S1800912A 4.5 Digit Display PCB for Oxy

Answers from our engineering team on installation, compatibility and integration.

01What does Servomex S1800912A do?
It is the complete 4.5 digit display board for the Oxy 1800, giving the optional 0.00 to 100.00 percent O2 readout.
02How does it differ from S1800902A?
S1800902A is the 3.5 digit board with 0.1 percent resolution; S1800912A is the 4.5 digit board with 0.01 percent resolution.
03Does S1800912A change the analyser accuracy?
No. The display resolution changes, but the specified intrinsic error stays less than 0.2 percent of reading or 0.05 percent O2.
04Where is the display board located?
It sits behind the display window in the hinged left cover, which opens after removing the four M6 bolts.
05Do I need the housekeeping board too?
Not necessarily; the housekeeping pcb S1800911 is a separate part.