Brand:

Servomex S1800989B Standard O2 Transducer for Oxy 1800

Description

Servomex S1800989B standard O2 transducer: what the part is

Servomex S1800989B is the standard oxygen transducer assembly for the Servomex SERVOTOUGH Oxy 1800, the safe-area paramagnetic oxygen analyser that older documents call the xendos 1800. In the manual the assembly sits in the table headed xendos 1800 sample option – standard, with the description Standard O2 Transducer Assy. It is the measuring heart of the analyser: the part that turns the magnetic susceptibility of oxygen in the sample into an electrical signal that the housekeeping board scales to 0.0 to 100.0 percent O2. When an Oxy 1800 drifts beyond what calibration can correct, or the cell has been wetted by condensate, this is the assembly a service team changes, and S1800989B is the number to quote.

This page was written for planners and buyers who need to identify the right transducer before raising a purchase request. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.

What S1800989B is made of and what it touches

The manual lists the materials in contact with the sample for a basic analyser as borosilicate glass, nickel, platinum, platinum/iridium alloy, 316 stainless steel, and Viton for the standard transducer or Chemraz for the solvent tolerant one. S1800989B is the Viton version. That detail decides compatibility: Viton is fine for clean, dry, non-corrosive streams such as air separation purity checks, nitrogen blanketing and inert gas monitoring, but it is the wrong choice where aromatic solvents or aggressive vapours are present.

Parameter Value
Part number S1800989B
Description in manual Standard O2 Transducer Assy., 1 off
Sample option Standard cell with 316 stainless steel tube kit S1800966
Measurement range 0.0 – 100.0 percent O2
Analogue output ranges Two ranges from 0-2.5, 5, 10, 25 and 100 percent O2
Intrinsic error Less than 0.2 percent of reading or 0.05 percent O2, whichever is greater
Zero drift Less than 0.05 percent O2 per week
Response time T90 Less than 4 seconds, base analyser
Noise, peak to peak Less than 0.04 percent O2, base analyser
Sample flow 50 to 250 ml/min for the base analyser
Ambient temperature coefficient 0.2 percent O2 plus or minus 0.5 percent of reading per 10 C
Operating temperature -10 C to +50 C
Analyser weight and size 26 kg; 448 x 235 x 229 mm

Where this transducer is used in Oman and GCC plants

Paramagnetic oxygen measurement is common wherever a safe-area oxygen reading is needed with a fast response. Typical homes for an Oxy 1800 carrying S1800989B are air separation and industrial gas filling plants, nitrogen generators on offshore and onshore utilities, heat-treatment furnaces in manufacturing, desalination and water-treatment ozone or oxygen feed lines, and laboratory gas supply panels. The analyser is not rated for hazardous areas or flammable samples; the manual points those duties to the 1900 series instead, so confirm the classification of the install point before ordering.

How S1800989B differs from other Oxy 1800 transducers

  • S1802989B: solvent resistant assembly, Chemraz seals, paired with Hastelloy C-276 tubes S1802966 for corrosive samples.
  • S1804989B: high-flow assembly with a response time under 5 seconds at 60 litres per hour and a flow effect of less than 0.2 percent O2 over 50 to 70 litres per hour.
  • S1806989B: combines the solvent resistant and high-flow features.
  • S1800989B: the standard cell, designed around the 50 to 250 ml/min bypass-free sample path and the lowest noise figure in the table.

If you are unsure which of the four applies, the transducer type is part of the analyser configuration, so the serial number or the initial order code is the quickest way to decide. A standard cell cannot simply be exchanged for a high-flow cell without changing the sample connections and flow device.

Sample conditions the replacement must be given

The sample table in the manual limits the base analyser to a non-condensing, particle-free gas below 3 micrometres, with a vent pressure of 11.5 to 18 psia that must never be restricted. With an automatic flow control device the inlet window is 1 to 5 psig at 1.2 to 3.5 litres per minute, and with a back pressure regulator the inlet is 17 to 22 psia at 1.0 to 2.0 litres per minute. A new S1800989B will be damaged just as quickly as the old one if these limits are ignored, so check the sample system before fitting it.

Fitting and commissioning notes for S1800989B

Replacement is a service task. Isolate sample and calibration gas, switch off the supply, and open the hinged left cover by removing the four M6 bolts with a 5 mm key. The service manual, part number 01800002B, is the reference for transducer exchange, and it is referred to in the quick start manual as intended for qualified personnel. After the new assembly is in place the manual calls for four hours of warm-up before sample gas is admitted, to prevent condensation in the cell. Then zero on high-purity nitrogen of at least 99.9 percent and span on clean instrument air at nominally 20.9 percent O2, or on a certified gas if the sample exceeds 21 percent.

Checks after installing S1800989B

Verify the isolated 4-20 mA output against the displayed value using the zero and span potentiometers in the terminal compartment, remembering that the current output never exceeds 22 mA and the load must stay under 600 ohms. Check the flow alarm relay and, if fitted, the filter seal. In oxygen purity applications atmospheric pressure influences the result by about 1 percent of reading per 1 percent pressure change, which the back pressure regulator option reduces to less than 0.13 percent.

Maintenance and spares planning around S1800989B

Routine care is light: replace the AFCD filter element every three months where the flow control device is fitted, wipe the housing with a slightly damp cloth, and never use solvents on it. Keep a fuse kit S1800987 and an enclosure seals kit S1800986 in stock; the transducer itself is a service-level spare held by sites that cannot tolerate long outages.

Documents and quotation

The Servomex SERVOTOUGH Oxy 1800 quick start manual, part number 01800003B revision 4, is the document we attach for Servomex S1800989B; it carries the spares table, the sample condition table and the performance figures quoted on this page. Please confirm in your request whether your analyser is the standard, solvent resistant or high flow version, because the three transducer part numbers are not interchangeable. To request a quotation, send the part number S1800989B, the quantity, the analyser serial number if you have it, and the delivery location in Oman or elsewhere in the GCC. We reply to each RFQ with availability and lead time.

Other Models in the Servomex Oxy1800 Range

Seven Star LLC supplies models from the Servomex Oxy1800 range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Servomex S1806989B Solvent Resistant High-Flow O2 TransducerGas Analyzer SystemsView details
Servomex S1802989B Solvent Resistant O2 Transducer Oxy 1800Gas Analyzer SystemsView details
Servomex S1800989B Standard O2 Transducer for Oxy 1800Gas Analyzer SystemsYou are viewing this model
Servomex S1804989B High-Flow O2 Transducer for Oxy 1800Gas Analyzer SystemsView details
Servomex S1800912A 4.5 Digit Display PCB for Oxy 1800Gas Analyzer SystemsView details
Servomex S1802966 Hastelloy C-276 Sample Tubes Kit Oxy 1800Gas Analyzer SystemsView details
Datasheets & Documents
  • PDF Datasheet Servomex S1800989B Standard O2 Transducer for Oxy 1800 Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Servomex is a British specialist in gas analysis, founded in 1952 by Richard Steel, John Crockett and Ronald Mudie; the name comes from servomechanism. It first made electronic products, then built its first paramagnetic oxygen analyser in 1961 after work with Charles Munday on a paramagnetic sensor. Servomex is based in Crowborough, East Sussex, and is part of Spectris.

Seven Star LLC lists Servomex process, combustion, emissions and portable gas analyzers, plus accessories, and sources them for refineries, petrochemical plants and industrial gas producers in Oman.

SERVOMEX company profile

Founded 1952, United Kingdom
Founders Richard Steel, John Crockett and Ronald Mudie
Headquarters Jarvis Brook, Crowborough, East Sussex
Ownership Spectris plc, since 1999
Key technologies Paramagnetic oxygen sensing (1961), infrared analysis (1969), zirconia oxygen sensors (1981)
Product families MultiExact and SERVOPRO MultiExact analysers, SERVOTOUGH OxyExact and Oxy 1800 oxygen analysers, SERVOTOUGH Laser 3 Plus TDL analysers, SERVOFLEX portable analysers, zirconia combustion analysers

Milestones

  • 1952: Servomex is founded.
  • 1961: The first Servomex paramagnetic oxygen analyser is introduced.
  • 1981: Servomex introduces zirconia oxygen sensors.
  • 1999: Spectris acquires Servomex.

Servomex analyzer families listed

Servomex groups its analyzers under family names that indicate where they can be used. SERVOTOUGH covers analyzers built for hazardous areas and field mounting, SERVOPRO covers units for general purpose areas and analyzer houses, and SERVOFLEX is the portable line. On this page:

  • SERVOTOUGH OxyExact 2200: paramagnetic oxygen analyzer for process control and safety
  • SERVOTOUGH SpectraExact 2500: photometric analyzer for process gas components
  • SERVOTOUGH Laser 3 Plus: compact tunable diode laser analyzer for combustion
  • SERVOPRO 4900 Multigas: multi-component analyzer for emissions monitoring
  • SERVOPRO NanoTrace DF-550E: ultra-trace oxygen analyzer for high purity gases
  • SERVOFLEX Micro i.s. 5100: intrinsically safe portable oxygen analyzer

A flow alarm relay box for the 2700 flue gas analyzer is also listed.

Where Servomex gas analyzers are used

Typical duties are oxygen measurement for inerting and blanketing, vapor recovery and flare header safety, combustion control on fired heaters and reformers, purity control in air separation and electronic gases, and continuous emissions monitoring. In the GCC the main users are refining, petrochemicals, fertilizer, LNG and industrial gas plants.

SERVOMEX products we list

Family Examples
Gas Analyzer Systems SERVOPRO 4900 Multigas Analyzer, SERVOPRO NanoTrace DF-550E Ultra-Trace O…, SERVOTOUGH Laser 3 Plus Combustion TDL G…, SERVOTOUGH OxyExact 2200 Oxygen Analyzer, ServoTough SpectraExact 2500 Gas Analyzer
Portable Test & Measurement Micro i.s. 5100 Portable Oxygen Analyzer
Gas Analyzer Accessories Tough Flue Gas 2700 Flow Alarm Relay Box

Ordering Servomex analyzers and spares: what we need

A Servomex gas analyzer is built to an order code that fixes the measured gas, range, sample wetted materials, outputs and hazardous area approval. For replacements and spare parts, send the model, the serial number and the order code from the data plate. For a new Servomex oxygen analyzer or multigas unit, send the full stream composition, the component and range to be measured, sample pressure and temperature, the area classification, power supply and communication needs. We check the configuration against Servomex documentation before quoting.

Facts checked against Servomex's own 70-year history and company profiles, October 2026.

FAQ

Questions about the Servomex S1800989B Standard O2 Transducer for Oxy

Answers from our engineering team on installation, compatibility and integration.

01What is Servomex S1800989B?
Servomex S1800989B is the standard oxygen transducer assembly listed under the standard sample option in the xendos 1800 Series spares table, intended for the SERVOTOUGH Oxy 1800 paramagnetic oxygen analyser.
02Which sample option does S1800989B belong to?
The quick start manual groups S1800989B with the standard sample option, alongside the S1800966 stainless steel tube kit, the S1800981 flow alarm, the S1800988 back pressure regulator and the two AFCD units.
03How does S1800989B differ from S1802989B?
S1802989B is the solvent resistant transducer for corrosive samples, with Chemraz instead of Viton seals. S1800989B is the standard transducer with Viton seals, suited to clean non-corrosive samples.
04What sample flow suits an analyser fitted with S1800989B?
The base analyser takes 50 to 250 ml/min at a measuring cell inlet of about 0.3 kPa gauge minimum, and the flow rate effect is less than 0.1 percent O2 over that range.
05Can Seven Star supply S1800989B with a datasheet?
Yes. We attach the Servomex quick start manual that lists the part, and we quote against the exact part number you send in your RFQ.