Brand:

Leybold 140 091 TRIVAC D 16 B-Ex Explosion Protected Vacuum Pump

Description

A TRIVAC for the places where gas can burn or explode

The Leybold 140 091 is the TRIVAC D 16 B-Ex IIB3 T4: an explosion-protected and pressure-burst-resistant two-stage rotary vane pump built as a complete system on a base frame, with an Ex e II T4 motor. Leybold specifies ATEX Category 1 inside the pump and Category 2 outside, marked II 1/2G Ex h IIB3 T4 Ga / Ex h IIC T4 Gb for an ambient range of 12 to 40 degrees C, under EC Type Examination Certificate IBExU03ATEX1017 X.

Category 1 inside the pump means the pump chamber itself is rated for an explosive atmosphere. The IIB3 T4 version is the subject of this page, while the sister code 140 092 (IIC T4, certificate IBExU03ATEX1016 X) is certified for group IIC gases with the exception of acetylene and carbon disulphide, as the footnote states. Check the group of your process gas before choosing between them.

Technical data of the D 16 B-Ex

Parameter Specification
Part number / type 140 091 / TRIVAC D 16 B-Ex IIB3 T4, two-stage, explosion protected and pressure burst resistant
Marking II 1/2G Ex h IIB3 T4 Ga / Ex h IIC T4 Gb (12 to 40 degrees C); EC Type Examination Certificate IBExU03ATEX1017 X
ATEX category Category 1 inside, Category 2 outside
Nominal pumping speed 18.9 m3/h (11.1 cfm)
Pumping speed (DIN 28 400) 16 m3/h (9.4 cfm) for 140 091; 15 m3/h (8.8 cfm) for 140 092
Ultimate partial pressure without gas ballast 1 x 10-4 mbar
Ultimate total pressure with gas ballast Below 5 x 10-3 mbar (below 3.8 x 10-3 Torr)
Water vapour tolerance 25 mbar (18.8 Torr)
Water vapour capacity 240 g/h (0.529 lbs/h)
Oil filling, min. / max. 0.55 l / 1.3 l (0.58 / 1.4 qt)
Motor 3~, 230 V / 400 V, 50 Hz, Ex e II T4
Type of protection IP 55
Maximum gas inlet temperature 60 degrees C (140 degrees F)
Highest permissible pressure in the oil box 1500 mbar (1125 Torr)
Ambient temperature 12 to 40 degrees C (46 to 104 degrees F)
Maximum surface temperature 135 degrees C (275 degrees F)
Maximum inlet pressure Atmospheric pressure
Weight, complete system 72 kg (159 lbs)
Connections Intake G 3/4 inch (inside thread); pressure side G 1 inch (inside thread)
Materials in contact with gas Steel, hardened steel, spring steel, stainless steel, zinc, aluminium and aluminium alloys, grey cast iron 25, FKM, felt, glass, silicone, polyamide

140 091 compared with 140 092

Item 140 091 140 092
Designation TRIVAC D 16 B-Ex IIB3 T4 TRIVAC D 16 B-Ex IIC T4
Marking as printed II 1/2G Ex h IIB3 T4 Ga / Ex h IIC T4 Gb II 1/2G Ex h IIC (no C2H2, CS2) T4 Ga / Ex h IIC T4 Gb
Certificate IBExU03ATEX1017 X IBExU03ATEX1016 X
Pumping speed 16 m3/h (9.4 cfm) 15 m3/h (8.8 cfm)

The catalogue asks that all enquiries and orders for category 1 and 2 ATEX products use the ATEX questionnaire printed at the end of the full-line catalogue.

Installation and safe operation points

  • The pump is delivered as a complete system including the Ex e motor, so connection must follow the explosion-protection documents supplied with it.
  • Keep the gas inlet temperature at or below 60 degrees C and the ambient between 12 and 40 degrees C; the maximum surface temperature is 135 degrees C, which sets the T4 temperature class.
  • The highest permissible pressure in the oil box is 1500 mbar, and the maximum inlet pressure is atmospheric pressure.
  • The catalogue drawing shows the connection layout: G 3/4 inch at the intake and G 1 inch at the pressure side.
  • The two pumping speed diagrams in the catalogue (IIB3 T4 and IIC T4) differ, so size the duty point against the curve of the part number you select.

Where it is used

The D 16 B-Ex is meant for chemical, pharmaceutical and petrochemical processes where the pumped gas or vapour can form an explosive mixture, such as solvent distillation, reactor evacuation, tank vapour handling and laboratory rigs inside classified areas. With 16 m3/h pumping speed and a water vapour tolerance of 25 mbar, it suits pilot units and small process vacuum lines, and the pressure-burst-resistant construction is the point of the product, not the capacity. As always the plant operator must assign zones and gas groups; the pump documentation does not replace a hazardous area assessment.

State the gas group and zone of your application when asking for 140 091 or 140 092, so the right variant is quoted. Wider choices are in the pumps category.

Selection and installation notes

A vacuum pump is selected from the pumping speed, the ultimate pressure and the gases or vapours it will handle. Use the oil grade the maker names, check the oil level before starting, connect the exhaust to a safe place and fit an oil mist filter where needed, use a trap or filter ahead of the pump for dust and vapour and warm the pump before pumping condensable vapour. Service the oil and filters on the schedule in the manual.

For a hazardous area, use only the certified version, keep the ATEX marking and the supply and gas conditions within the certificate and follow the local hazardous area rules.

Information needed for a quotation

Send the full model code, the quantity, the delivery location in Oman or the GCC, the intended use and any accessories needed, so that the complete package can be quoted together.

Documents and supply information

The manufacturer data sheet is attached to this page and should be read together with the project specification and the site risk assessment before a purchase is confirmed. Ratings, standards and dimensions are quoted as printed on that sheet. Lead time, stock position and certificates available with the shipment are confirmed in the quotation, and we supply to Oman and the wider GCC region for industrial, utility and building services projects.

Frequently asked questions

What is Leybold 140 091?

It is the TRIVAC D 16 B-Ex IIB3 T4, an explosion protected and pressure burst resistant two-stage rotary vane pump.

What marking does it carry?

II 1/2G Ex h IIB3 T4 Ga / Ex h IIC T4 Gb for 12 to 40 degrees C, with certificate IBExU03ATEX1017 X, as printed in the catalogue.

What motor is fitted?

A three-phase 230 V / 400 V, 50 Hz Ex e II T4 motor, protection class IP 55.

How does 140 092 differ?

It is the IIC T4 variant, certified for group IIC gases except acetylene and carbon disulphide, with 15 m3/h pumping speed.

Does Seven Star LLC decide which variant is allowed in a zone?

No. Seven Star LLC is an independent supplier in Muscat; the operator and the responsible engineer decide on the variant and zone.

Datasheets & Documents
  • PDF Datasheet Leybold Full Line Catalog 2022, oil sealed vacuum pumps: TRIVAC D 16 B-Ex extract (catalogue pages 34-35) Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Leybold is a German manufacturer of vacuum pumps and vacuum technology, headquartered in Cologne, with a history that goes back to 1850. Ernst Leybold's firm became known for vacuum research, including Wolfgang Gaede's work on the turbomolecular principle in 1911 and the diffusion pump in 1913. Leybold was part of Oerlikon until September 2016, when Atlas Copco completed its acquisition.

Seven Star LLC is an independent supplier in Muscat. We list selected Leybold pumps with the official catalogue pages for industrial and laboratory users in Oman and the GCC.

Leybold company profile

Origins 1850, Cologne, Germany
Headquarters Cologne, Germany
Ownership Atlas Copco, since September 2016
Previous owner Oerlikon
Product families SOGEVAC single-stage and TRIVAC two-stage rotary vane pumps (including ATEX versions), NEO D rotary vane pumps, VACUBE screw pumps, dry screw and Roots pumps, turbomolecular pumps, vacuum gauges

Milestones

  • 1850: The business is founded in Cologne.
  • 1911: The turbomolecular pump principle is developed with Wolfgang Gaede.
  • 2016: Atlas Copco acquires Leybold.
  • 2025: Leybold marks 175 years.

Leybold product series

SOGEVAC single-stage oil-sealed rotary vane pumps on this site include the SV 16 D, SV 25 B, SV 100 B and SV 220 B, and the SV 40 ATEX Category 2 pump for gas group IIB+H2. Two-stage rotary vane pumps include the NEO D 25 and the TRIVAC D 40 B, and the TRIVAC D 16 B-Ex is an explosion-protected version.

The VACUBE VQ 400 i and VQ 700 i are oil-sealed screw vacuum pumps with variable speed drive for larger volume flows. The catalogue numbers (for example 960 250) identify the exact pump variant and motor.

Where Leybold pumps are used

Vacuum packaging and food processing, drying and degassing, coating, metallurgy and heat treatment, transformer and power engineering work, chemical processes, laboratories and analytical instruments. ATEX versions serve areas where explosive gases may be pumped.

Leybold products we list

Family Examples
Pumps 10770400V01 VACUBE VQ 400 i Oil-Sealed S…, 112 86 TRIVAC D 40 B Two-Stage Rotary Va…, 140 091 TRIVAC D 16 B-Ex Explosion Prote…, 960 250 SOGEVAC SV 25 B Rotary Vane Vacu…, 960 345 SOGEVAC SV 40 ATEX Cat 2 Vacuum…, 970202V NEO D 25 Two-Stage Rotary Vane V…

Ordering Leybold pumps: what we need

Send the Leybold catalogue number from the nameplate, or the pumping speed, ultimate pressure, gas or vapour being pumped, motor voltage and frequency, and any ATEX requirement. For service, state the pump model and serial number so the right oil, filters and kits can be quoted.

Facts checked against Leybold's own 175-year release and Atlas Copco announcements, October 2026.

Frequently Asked Questions

What is Leybold 140 091?
It is the TRIVAC D 16 B-Ex IIB3 T4, an explosion protected and pressure burst resistant two-stage rotary vane pump.
What marking does it carry?
II 1/2G Ex h IIB3 T4 Ga / Ex h IIC T4 Gb for 12 to 40 degrees C, with certificate IBExU03ATEX1017 X, as printed in the catalogue.
What motor is fitted?
A three-phase 230 V / 400 V, 50 Hz Ex e II T4 motor, protection class IP 55.
How does 140 092 differ?
It is the IIC T4 variant, certified for group IIC gases except acetylene and carbon disulphide, with 15 m3/h pumping speed.
Does Seven Star LLC decide which variant is allowed in a zone?
No. Seven Star LLC is an independent supplier in Muscat; the operator and the responsible engineer decide on the variant and zone.